The EC91 is an intrinsically safe oxygen analyser designed for hazardous areas, certified under ATEX Directive 2014/34/EU. It is suitable for applications such as monitoring inert blanketing gas in oil and petrochemical industries, glove boxes, and metallurgy gas line monitoring.

Features & Benefits
  • Intrinsically safe design for Zone 0 areas, eliminating the need for work permits.
  • Maintenance-free sampling cell with a 5-year life span.
  • 36-month warranty covering workmanship and component failure.
  • Optional remote cell and safe area digital indicator and alarm.

Technical Specifications
  • Measurement ranges: 0-20 ppm to 0-30% oxygen, with a resolution of 0.05% of scale.
  • Accuracy: ±2% of reading at 20°C for >10ppm, ±5% over temperature range.
  • Response time: 90% of reading within 20 seconds.
  • Operating conditions: Sample pressure 0.1 to 1 Barg, flow rate 30ml/min to 5 ltr/min, temperature 0 to 40°C.
  • Power supply: 24Vdc, power consumption 10W.
  • Enclosure: Reinforced polyester, IP66/Nema 4X protection.

Installation & Operation
  • Sample pressure should be above 0.1 Bar; otherwise, a pump or aspirator is needed.
  • Connection via 1/8” tube fitting with a flow regulator on the inlet.
  • Optional dust or coalescing filters available.
  • Transmits a 4-20mA signal corresponding to the analyser range.

Principle of Operation
The sensor uses an electrochemical fuel cell with an anode, electrolyte, and air cathode. Oxygen is reduced to hydroxyl ions at the cathode, oxidizing the metal anode. The sensor has a guaranteed operational life of six months, extendable under low oxygen conditions.

Additional Options
  • Remote probe holder, control room display, aspirators, sample systems, and alarm outputs.
  • Local analogue display option.

Contact Information
Systech Illinois has offices in the U.S., UK, Thailand, and China, providing global support and expertise in gas analysis solutions.
See more

Catalog excerpts

EC91 Intrinsically safe oxygen analyser-1

EC91 IntrinsicaMy Safe Oxygen Analyser Intrinsically Safe ATEX, unsurpassed certified oxygen analyser for all hazardous areas Monitoring inert blanketing gas in oil and petrochemical applications Inert gases and Hydrogen Gas purity Glove boxes Oxygen deficiency monitoring Metallurgy Gas line monitoring Unsurpassed safety protection Air calibration facility Maintenance-free 5 year life sampling cell Certified for Zone 0 This instrument has a 36 month warranty which covers any faulty workmanship and normal component failure relating to electronic circuit cards C e Conforms to ATEX Directive 2014/34/EU, Marking ^ II 1G EX ia IIC T4 Ga

 Open the catalog to page 1
EC91 Intrinsically safe oxygen analyser-2

Intrinsically Safe Oxygen Analysis Systech Illinois have over 25 years of expertise in manufacturing gas analysers for the process industry and have an installed base of over a thousand instruments during this time. The EC91 Process Oxygen Transmitter will detect levels of oxygen as low as 1ppm, up to higher percent levels and can be used on most industrial gases and atmospheres. There is no need for routine maintenance of the fuel cell and the instrument can be easily calibrated, using ambient air or standard calibration samples. The enclosure is manufactured from moulded glass fibre reinforced...

 Open the catalog to page 2
EC91 Intrinsically safe oxygen analyser-3

lira Ranges Resolution Accuracy Response Time Calibration Range Measuring Cell Type Operating Conditions Sample inlet pressure Sample flow rate Sample temperature Ambient temperature Sample connections Unsuitable gases 6 selectable 0-20, 0-200, 0-2000 (ppm) 0-2%, 0-20%, 0-30% Other ranges available on request 0.05% of scale >10ppm ±2% of reading at 20°C ±5% of reading over temperature range <10ppm ±2% of reading + 0.4ppm at 20°C ±5% of reading + 0.4ppm + 0.15ppm/°C over temperature range 90% of reading within 20 seconds Ambient (20.9%) or certified gas Electrochemical fuel cell 0.1 to 1 Barg,...

 Open the catalog to page 3
*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.