SUP-PH6.3 pH meter
1 / 10Pages

Catalog excerpts

SUP-PH6.3 pH meter - 1

E-mail: sales@supmea.com Online pH/ORP meter Committed to process automation solutions Add: 5th floor, Building 4, Singapore-Hangzhou Science & Technology Park, Hangzhou,China

Open the catalog to page 1
SUP-PH6.3 pH meter - 2

E-mail: sales@supmea.com SUP-pH6.3 pH meter is a true multivariable analyzer used for measuring/controlling the pH with temperature and ORP (The function is switchable on the device itself). Depending on the measured variable, combination electrodes (e. g. pH sensors) or split versions (glass electrodes with a separate reference electrode) can be readily connected. Truly unique is the pH meter in the Human Machine Interface. The high resolution digital display and LCD screen make the measurements are clearly legible and the keyboard operation make it easier for the user to configure the...

Open the catalog to page 2
SUP-PH6.3 pH meter - 3

E-mail: sales@supmea.com Features ■ Easy operation ■ NEMA enclosure for field mounting and panel mounting ■ Automatically Temperature Compensation ■ Directly switchable to PH or ORP ■ Large LCD display with background lighting ■ PH or ORP sensors can be connected thanks to the sensor supply integrated in the output ■ Using the setup program: user-friendly programming ■ 4-20mA analog output ■ RS485 communication ■ Relay output Add: 5th floor, Building 4, Singapore-Hangzhou Science & Technology Par

Open the catalog to page 3
SUP-PH6.3 pH meter - 4

E-mail: sales@supmea.com Description The device is designed for use on site. A rugged housing protects the electronics and the electrical connections from corrosive environmental conditions . As an alternative, the device can also be installed in a control panel. Temperature:Compensation temperature Analyst output:Analyst output Measured value:Real-time measurements value High alarm:High alarm Low alarm:Low alarm ESC:Check related warning status on the “monitoring page” Return to previous level page in the up& down level page linked to “menu page” 7. Menu: Enter the MENU on the “monitoring...

Open the catalog to page 4
SUP-PH6.3 pH meter - 5

E-mail: sales@supmea.com Wiring diagram Identification of terminal ·INPUT: Measuring terminal of the electrode ·REF: Reference terminal of the electrode ·A: Temperature compensation terminal A,NTC10K and PT1000 connect here ·B: Temperature compensation terminal B,NTC10K and PT1000 connect here ·C: Temperature compensation terminal C, PT1000 three-wire temperature grounding, PT1000 two-wire need to be short-connected to TEMPB, not NTC10K. ·NC: Unidentified ·RS485 (A +): RS485 communication interface A + ·RS485 (B -): RS485 communication interface B·4-20mA (+): 4-20mA output end+ ·4-20mA (-):...

Open the catalog to page 5
SUP-PH6.3 pH meter - 6

E-mail: sales@supmea.com pH calibration A pH calibration is the procedure of adjusting the pH meter by measuring solutions of known pH values. Why you need to calibrate: The characteristic of a pH electrode will change with time due to electrode coating and aging. And even a pH electrode would be stable over time, pH electrodes cannot be produced with identical characteristics. In practice the response of a real pH sensor does not exactly follow the Nernst equation. This difference between the theoretical and actual behavior of a pH electrode must be compensated for. A calibration is...

Open the catalog to page 6
SUP-PH6.3 pH meter - 7

E-mail: sales@supmea.com How to calibrate: Main Menu- Online Calibration- PH Calibration PH Calibration: Enter the PH calibration screen, the first PH electrode into the 4.00PH standard solution, standing for a moment, to be stable after the show, press the 【ENT】, then the PH electrode into the 6.86PH standard solution Set the PH electrode into the 9.18PH standard solution, put it aside for a moment, after the test is stable, press 【ENT】 , the display shows the success of the calibration, the whole process of PH calibration completes. PH Modification: The measured PH can be modified between...

Open the catalog to page 7
SUP-PH6.3 pH meter - 8

E-mail: sales@supmea.com Multi-point calibration To achieve the best possible accuracy, the calibration should cover the range of the desired measurement values. If the readings go beyond the calibrated range, the pH meter assumes linearity and simply extrapolates the value to be displayed. The true value may be slightly different. More advanced pH meters will let the user calibrate at three, four or five and even higher numbers of pH values. A multi-point calibration mean, in comparison to a two-point calibration, that you can calibrate your pH tester on both sides of the zero point (pH...

Open the catalog to page 8
SUP-PH6.3 pH meter - 9

E-mail: sales@supmea.com Ordering Code pH Probe C:Plastic pH sensor with NTC 10K (0-60℃) D: Glass pH sensor with NTC 10K (0-80℃) E:High Tmeperature pH sensor with NTC 10K (0-130℃) F: Made in Germany pH sensor E+H (0-80℃) Medium P: Normal water Q: Waste water R: Nutrient solution S: Galvanic plating T: Cooling water U: Drinking water V: Others: G:Made in Germany pH sensor Jumo (0-80℃) H: Plastic ORP sensor with NTC 10K (0-60℃) I: Glass ORP sensor with NTC 10K (0-80℃) Cable Length J:5 meters K: 10 meters J:15 meters Note: 4-20mA, RS485 and relay output are standard functions Make inquiries by...

Open the catalog to page 9
SUP-PH6.3 pH meter - 10

E-mail: sales@supmea.com Supmea China Headquarters Supmea Singapore Branch Supmea German Branch Supmea Malaysia Branch Address: 5th floor, Building 4, Address: No 3, Jalan Emas Singapore-Hangzhou Science & Jaya 1, Taman Industries Technology Park, Hangzhou, Exchange Singapore Emas jaya Tongkang Pecah , Batu Pahat Add: 5th floor, Building 4, Singapore-Hangzhou Science & Technology Park, Hangzhou,China

Open the catalog to page 10