SUP-PH2.2T pH controller
10Pages

{{requestButtons}}

Catalog excerpts

SUP-PH2.2T pH controller - 1

E-mail: info@supmea.com Online pH/ORP controller Committed to process automation solutions Add : 5th floor, Building 4, Singapore Hangzhou Science Technology Park, Hangzhou Economic Development Area, 310018, Hangzhou, China

Open the catalog to page 1
SUP-PH2.2T pH controller - 2

E-mail: info@supmea.com SUP-PH 2.2T meter is an instrument independently developed by our company for online monitoring of pH/ORP value and through the current analog output to the monitoring room for record preservation. pH /ORP controller is widely used in thermal power, chemical fertilizer, metallurgy, environmental protection, Pharmaceutical, biochemical, food and water industries and for solution pH or ORP Values and temperatures are continuously monitored.The continuous monitoring data can be recorded by remote transmission through substation and output connection. Characteristics ■...

Open the catalog to page 2
SUP-PH2.2T pH controller - 3

E-mail: info@supmea.com www.supmea.com temperature measurement range, maximum load 750 Ω, output accuracy + / - 0.2% FS. Alarm function Relative humidity Operating temperature Power supply Storage condition Relative humidity: 5~85%RH (no condensation) Altitude: <2000m Add : 5th floor, Building 4, Singapore Hangzhou Science Technology Park, Hangzhou Economic Development Area, 31001

Open the catalog to page 3
SUP-PH2.2T pH controller - 4

E-mail: info@supmea.com Installation dimension Fixed installation Add : 5th floor, Building 4, Singapore Hangzhou Science Technology Park, Hangzhou Economic Development Area, 310018, Hangzhou, China

Open the catalog to page 4
SUP-PH2.2T pH controller - 5

E-mail: info@supmea.com pH 2.2T pH controller 1. Measuring parameter pH 2. Measuring parameter ORP 3. High setting sign 4. Low setting sign 5. 4~20mA put up signs 6. Online calibration sign 7. Configuration settings sign 8. Buzzer sign 9. Current unit 10. ORP unit 11. pH unit 12. Temperature display 13. Temperature unit 14. Main display 15. Manual temperature compensation sign 16. Automatic temperature compensation sign Add : 5th floor, Building 4, Singapore Hangzhou Science Technology Park, Hangzhou Economic Development Area, 310018, Hangzhou, China

Open the catalog to page 5
SUP-PH2.2T pH controller - 6

E-mail: info@supmea.com Identification of terminal ■ REF: Reference terminal of the electrode ■ INPUT: Measuring terminal of the electrode ■ TEMP1: Temperature compensating end 1 ■ TEMP2: Temperature compensating end 2 ■ RS485A+:RS485 communication interface A+ ■ RS485B-:RS485 communication interface B■ NC: Unidentified ■ 4-20mA (+): 4-20mA output end+ ■ 4-20mA (-): 4-20mA output end■ AC220V (L): AC220V Fire Wire ■ AC220V (N): AC220V zero line ■ LO: low alarm normally open relay ■ HI(AH): High alarm normally open relay Add : 5th floor, Building 4, Singapore Hangzhou Science Technology Park,...

Open the catalog to page 6
SUP-PH2.2T pH controller - 7

E-mail: info@supmea.com pH calibration A pH calibration is the procedure of adjusting the pH meter by measuring solutions of known pH values. Why you need to calibrate: The characteristic of a pH electrode will change with time due to electrode coating and aging. And even a pH electrode would be stable over time, pH electrodes cannot be produced with identical characteristics. In practice the response of a real pH sensor does not exactly follow the Nernst equation. This difference between the theoretical and actual behavior of a pH electrode must be compensated for. A calibration is...

Open the catalog to page 7
SUP-PH2.2T pH controller - 8

E-mail: info@supmea.com Multi-point calibration To achieve the best possible accuracy, the calibration should cover the range of the desired measurement values. If the readings go beyond the calibrated range, the pH meter assumes linearity and simply extrapolates the value to be displayed. The true value may be slightly different. More advanced pH meters will let the user calibrate at three, four or five and even higher numbers of pH values. A multi-point calibration mean, in comparison to a two-point calibration, that you can calibrate your pH tester on both sides of the zero point (pH...

Open the catalog to page 8
SUP-PH2.2T pH controller - 9

E-mail: info@supmea.com Ordering Code Basic Type pH2.2T Plastic pH/ORP sensor with NTC 10K (0-60℃) Glass pH/ORP sensor with NTC 10K (0-80℃) High Temperature pH sensor with NTC 10K (0-130℃) pH/ORP Probe Made in Germany pH sensor Made in Germany pH sensor Jumo (0-80℃) Plastic pH/ORP sensor with NTC 10K (0-60℃) Glass pH/ORP sensor with NTC 10K (0-80℃) 5 meters Cable Length Calibration Points Power Supply 24VDC Normal water Waste water Nutrient solution Galvanic plating Cooling water Drinking water Others: Note: 4-20mA, RS485 and relay output are standard functions, make inquiries by placing...

Open the catalog to page 9
SUP-PH2.2T pH controller - 10

E-mail: info@supmea.com

Open the catalog to page 10

All Supmea Automation catalogs and technical brochures