ATM.ECO/IS Analog Transmitter with Temperature compensation
8Pages

{{requestButtons}}

Catalog excerpts

ATM.ECO/IS Analog Transmitter with Temperature compensation - 1

Pressure Transmitter - FM / FM-C certified ATM.ECO/IS - Analog Transmitter with Temperature compensation CUSTOMER BENEFITS • Certificate: FM, FM-C & IECEx & ATEX • Entry level series into precision pressure measurement • High accuracy is ensured by sophisticated digital temperature compensation algorithms • Fast customization thanks to configurable product design • Reverse polarity and short circuit protected, with integrated overvoltage protection

Open the catalog to page 1
ATM.ECO/IS Analog Transmitter with Temperature compensation - 2

Long term stability, (7) < 0.5% FS / < 4 mbar < 0.2% FS / < 4 mbar < 0.1% FS / < 0.2% FS Burst pressure, (4) > 850 / < 1500 bar > 1500 bar Response time, (typ.) < 1ms / 10 ... 90% FS < 1ms / 10 ... 90% FS Long term stability, (7) < 0.1% FS / < 0.2% FS < 0.1% FS / < 0.2% FS (2) Overpressure and burst pressure 1500 bar (stainless steel) optional (3) Process connection frontal and flush diaphragm availables 600 bar (5) Zero based accuracy according to DIN-16086, incl. hysteresis and repeatability at ambient temperature (6) Total error including accuracy and...

Open the catalog to page 2
ATM.ECO/IS Analog Transmitter with Temperature compensation - 3

Power supply Supply influence Circuit diagram Load resistance Load influence < 0.05% FS Reverse polarity protection Pout to +Vin Certificate, (1) SEV 09 ATEX 0108 X IECEx SEV 10.0003 X IECEx MSC 14.0002 X Dust II 1D Ex ia IIIC IP6x T140°C EN 61241-0 / -11 intrinsically safe circuit (1) For detailed Ex specifications see certificate and operating and safety instructions (2) Without any information about temperature class the transmitter will be delivered for T4

Open the catalog to page 3
ATM.ECO/IS Analog Transmitter with Temperature compensation - 4

Description Level Typical interferences mmpp) EN 60068-2-27 Shock 100 G (impulse duration 6 ms) EN 55022 Emission, class B < 30 dBpV/m (0.03 ... 1 EN 61000-4-3 Irradiated RF 10V/m (0.08 ... 2.7 GHz, 3s) Radio sets, wireless phones EN 61000-4-4 Transients (burst) 4 kV Motors, valves Line-Earth: 1 kV/42 Q EN 61000-4-6 Conducted RF 3 V (0.15 ... 80 MHz, 3 s) Frequency converters Transducer Stainless steel (316L / 1.4404) Seals Viton (Standard), EPDM,

Open the catalog to page 4
ATM.ECO/IS Analog Transmitter with Temperature compensation - 5

Cable Socket Connector Article number

Open the catalog to page 5
ATM.ECO/IS Analog Transmitter with Temperature compensation - 7

(3) Process connection available < 600 bar (4) Cable socket connector not included (5) Please specify the required cable length and medium (6) For operating temperature > 50°C, FEP cable must be used (7) Only with pressure connection Fig. 2, Fig. 3, Fig. 6 and Fig. 7 (16) Connector side not to be used in Zone 0 or Ex ia IIC, explosion risk

Open the catalog to page 7
ATM.ECO/IS Analog Transmitter with Temperature compensation - 8

Pressure connections Electrical Connections Version for medium temperature up to 125°C Pin 2-wire L - 25 mm for connector DIN 43650 E EP (only Ex) Version for medium temperature >125°C up to 150°C socket connector L ^ 52 mm for connector DIN 43650 Pin 2-wire UiTTF 5 EP (only Ex) socket connector Pin 2-wire white +Vin yellow Pout socket connector 3 Pout 4 +Vin Specifications may change without notice. STS Headquarters, Switzerland: STS Sensor Technik Sirnach AG Rutihofstrasse 8 | 8370 Sirnach | Switzerland sales@stssensors.com | www.stssensors.com STS Germany: STS Sensoren Transmitter...

Open the catalog to page 8

All STS Sensor Technik Sirnach AG catalogs and technical brochures

  1. Accessories