This is your Dry Bath System user manual.
12Pages

{{requestButtons}}

Catalog excerpts

This is your Dry Bath System user manual. - 1

This is your Dry Bath System user manual. USER MANUAL · BEDIENUNGSANLEITUNG MODE D‘EMPLOI ISTRUZIONI PER L‘USO · INSTRUCCIONES DE MANEJO

Open the catalog to page 1
This is your Dry Bath System user manual. - 2

EC Declaration of Conformity EG-Konformitätserklärung Wir / We / Nous: Starlab International GmbH erklären in alleiniger Verantwortung, dass folgendes Produkt / declare under our sole responsibility that the following product / déclarons sous notre propre responsabilité que le produit ci-dessous Starlab Dry Bath System, Single Block, Digital. N2400-4001 Starlab Dry Bath System, Dual Block, Digital. N2400-4002 auf das sich diese Erklärung bezieht, mit den folgenden Normen oder normativen Dokumenten übereinstimmt: / to which this declaration relates, is in conformity with the following...

Open the catalog to page 2
This is your Dry Bath System user manual. - 4

This equipment has been tested and found to comply with the limits for the CE regulation. These limits are designed to provide reasonable protection against harmful interference when the equipment is operated in a commercial environment. This equipment generates, uses, and can radiate radio frequency energy, and if not installed and used in accordance with the This equipment has been tested and found to comply with the limits for the CE regulation. instruction manual, may cause harmful interference to radio communications. Operation of These limits are designed to provide reasonable...

Open the catalog to page 4
This is your Dry Bath System user manual. - 5

1. Safety Information As with any electrical device, safety precautions need to be taken. Before connecting with the electrical supply, check the voltage is within the range stated at the rating label and that the device is earthed. Place the unit in a safe and dry location. The unit MUST NOT touch anything in the surrounding area. Always follow the safety precautions for chemicals/ dangerous materials, and Hot Surface. Environmental Conditions Ensure the instrument is installed and operated strictly in the following conditions: < = 95 % RH, 75 KPa-106 Kpa. Altitude must not exceed 2000...

Open the catalog to page 5
This is your Dry Bath System user manual. - 6

Equipment Operation Follow the guidelines below to ensure safe operation of the unit: Features Equipment Operation 1. To avoid significant discrepancies between set temperature and temperature reached, Follow theplease monitor the displayed temperature regularly eg. by using a calibrated laboratory guidelines below to ensure safe operation of the unit: thermometer. See page 9 “Temperature Calibration” for detailed information. 1. To avoid significant discrepancies between set temperature and temperature reached, 2. please NEVER access displayed temperature regularly monitor the any...

Open the catalog to page 6
This is your Dry Bath System user manual. - 7

Digital Microprocessor Controller Operating Temperature Temperature Increase Temperature Calibration Temperature Uniformity Temperature Accuracy Safety Device Leak-proof heating chamber Heating Chamber Material PTFE-coated aluminium Block Material Block Dimensions Water Bath Volume (approx.) Rated Voltages Unit Dimensions 3. Product Description The single block (N2400-4001) and dual block (N2400-4002) units use interchangeable heating block modules for a variety of applications, including restriction digests, denaturing DNA, BUN, melting agar, coagulation studies, in-situ hybridization and...

Open the catalog to page 7
This is your Dry Bath System user manual. - 8

4. Product Overview 4. Product Overview STARLAB Dry Bath Heating Systems are designed to cover a wide variety of 4. Product Heating Systems are designed to cover a wide variety of applications. Excellent temperature applications. Excellent temperature control delivers accurate and reliable experimental Starlab Dry BathOverview control delivers accurate and reliable experimental results, time after time. Starlab Dry Bath Heating Systems results, time after time. STARLAB Dry Bath Heating Systems are space-saving, compact are space-saving, compact instruments, offering great value and...

Open the catalog to page 8
This is your Dry Bath System user manual. - 9

6. Installation Instructions The Starlab Dry Bath Heating Systems are ready to ‘Plug and Play’. The unit should only be used on a sturdy, level surface in a safe, dry place. Before operation, ensure one or two heating aluminium block(s) is / are inserted into the bath, or if using as a water bath, that water has been added. Place the unit on a sturdy, level surface in a safe, dry place, away from laboratory traffic. Ensure that the AC power switch is OFF, then plug the power cord into a grounded AC outlet of appropriate voltage (115 V or 220 V as indicated on the rating sticker near the AC...

Open the catalog to page 9
This is your Dry Bath System user manual. - 10

8. Temperature Calibration The Starlab Dry Heating System with (optional) metal heating blocks has been calibrated as a set. However, the different kinds of block do not generate the same results. Therefore, for optimum temperature control accuracy, and when changing to different kinds of block, the unit should be calibrated using the procedure outlined below. 1. Insert a calibrated, 45 mm laboratory thermometer into the Thermometer holding port on the block. SIMULTANEOUSLY Press the ON/OFF power switch on the back of the unit and the START Key on the front. During this, the unit will omit...

Open the catalog to page 10
This is your Dry Bath System user manual. - 11

10. Maintenance The Starlab Dry Bath Heating System may be cleaned with a moist cloth containing a mild soap solution. The blocks are aluminum and may be cleaned with any of the commercial aluminum cleaners on the market. Starlab Dry Bath Heating System, Single Block*, Digital Starlab Dry Bath Heating System, Dual Block*, Digital Metal Block for 64 x 0.2 ml PCR Tubes Metal Block for 20 x 1.5/2.0 ml Microcentrifuge Tubes Metal Block for 12 x 15 ml Centrifuge Tubes Metal Block for 4 x 50 ml Centrifuge Tubes Metal Block for 20 x 1.5/2.0 ml Microcentrifuge Tubes (V-Bottoms) Metal Block for 20 x...

Open the catalog to page 11
This is your Dry Bath System user manual. - 12

Starlab GmbH Starlab France SARL Neuer Höltigbaum 38 22143 Hamburg Germany 5 Tanners Drive Milton Keynes MK14 5BU United Kingdom 30 Rue Jean Rostand 91400 Orsay France Via Pinturicchio 1 20133 Milano Italy

Open the catalog to page 12