Overview: The FS752 GNSS Time and Frequency Reference by Stanford Research Systems is a precision instrument designed to provide calibrated time and frequency distribution using GNSS constellations such as GPS, GLONASS, BEIDOU, and GALILEO. It features a double-oven OCXO timebase with low phase noise and offers multiple output options for laboratory use.
Specifications: The FS752 includes a double-oven OCXO timebase with a phase noise of less than -125 dBc/Hz at 10 Hz offset. It provides five 10 MHz outputs and two 1 pps outputs, with options for additional outputs. The GNSS receiver can track all visible satellites and optimize timing solutions, with a time wander of less than 20 ns rms under clear sky conditions.
GNSS Receiver and Antennas: The FS752 supports remote active GNSS antennas with a net gain requirement of +20 dBi to +32 dBi. It offers two antenna solutions with LNAs and a 50 Ω characteristic impedance. The receiver includes a configurable alarm system for monitoring GNSS signal lock and timebase health.
Communication and Control: The device can be controlled via a USB port using a high-level command set. A free GNSSDO application is available for easy configuration and status monitoring.
Performance and Accuracy: The FS752 offers a 1 pps output with a phase accuracy of less than 2 ns relative to the internal reference and a jitter of less than 50 ps rms. The 10 MHz output has an amplitude of 13 dBm with an accuracy of ±1 dB.
Power and Compliance: The unit operates on 30 W AC power with a range of 90 to 264 VAC and complies with FCC Part 15 and CISPR-22 Class B standards.
Ordering Information: The base price for the FS752 is $2250, with options for additional outputs and antennas available at extra cost. The device comes with a one-year warranty covering parts and labor for defects in materials and workmanship.
See more

Catalog excerpts

FS752 — GNSS Time and Frequency Reference-1

Frequency Standards FS752 — GNSS Time and Frequency Reference FS752 GPS Time & Frequency Reference · GPS/GNSS disciplined 10 MHz · Double-oven OCXO timebase · Low phase noise · Five 10 MHz outputs · Two 1 pps outputs · Up to eight additional 10 MHz & 1 pps outputs (opt.) The FS752 GNSS Disciplined Time and Frequency Reference provides calibrated time and frequency distribution to your laboratory. A built-in receiver tracks any of the four major GNSS constellations: GPS, GLONASS, BEIDOU, or GALILEO. The instrument comes with a double-oven OCXO timebase that has phase noise of less than -125 dBc/Hz at 10 Hz offset. 10 MHz and 1 pps Distribution The FS752 provides five buffered 10 MHz outputs and two buffered 1 PPS outputs. The 10 MHz outputs generate 1 Vrms into Stanford Research Systems

 Open the catalog to page 1
FS752 — GNSS Time and Frequency Reference-2

FS752 Time and Frequency Reference 50 Ω and may be used as frequency references for laboratory equipment. The 1 p p s o u tp u ts g e n era te 1 0 µs pulses, with 5 V CMOS logic and rising edges aligned to UTC. Up to eight additional 10 MHz and 1 pps outputs are available as rear-panel options. for the receiver’s sawtooth error, then phase locks the timebase to the GNSS 1 pps. The TDEV between two instruments is a few nanoseconds. If the GNSS signal is lost, the timebase is left at the last locked frequency value. The timebase will age or drift in frequency by less than ±0.05 ppm/year. GNSS Antennas...

 Open the catalog to page 2
FS752 — GNSS Time and Frequency Reference-3

OCXO Timebase Oscillator type Temp. Stability Aging Phase noise (SSB) Stability Holdover GPS Receiver Model Satelite acq. time Almanac acq. time Optimized for static applications Accuracy of UTC Time wander Antenna delay correction range 1 pps Output Period Width Phase accuracy Jitter Level Transition time Source impedance Reverse protection Double oven controlled, 3rd OT, SC-cut crystal <2 x 10-9 (20 to 30 °C) <0.2 ppm/year (undisciplined to GPS) <-125 dBc/Hz (typical) See graphs next page <40 ps / 24 hr. u-blox, NEO-M8T Less than 1 minute (typ.) Approximately 15 minutes when continuously tracking...

 Open the catalog to page 3

All Stanford Research Systems catalogs and technical brochures

  1. SR556

    1  Page

  2. QMSc

    3  Pages

  3. FS730/1

    4  Pages

  4. SC10

    2  Pages

  5. PRS10

    4  Pages

  6. SR715/720

    3  Pages

  7. CG635

    5  Pages

  8. PERF10

    2  Pages

  9. SR620

    4  Pages

  10. LCR Meters

    3  Pages

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.