Application
The CS18 FF Calibration System is designed for secondary calibration of measuring microphones, sound level meters, and other acoustic sensors in compliance with IEC 61094-8 and IEC 61672 standards. It is also used for periodic single qualification according to IEC 61672-3 and various acoustical measurements such as directivity characteristic, diffuse-field sensitivity, and acoustic emission of small objects. It supports calibration of non-standard microphones in both axial and radial directions.
Range of Use
This system is suitable for certified calibration laboratories, verification of measuring instruments in research and industry sectors like civil engineering, aviation, and automotive engineering, and quality assurance in the manufacturing of microphones and sound level meters.
Features
The system includes reference standards traceable to the Physikalisch Technische Bundesanstalt (PTB) and offers true free-field calibration in an anechoic chamber. It can calibrate any measuring microphone type and supply a defined free-field sound pressure level for sound level meters. It is compatible with other SPEKTRA calibration systems such as CS18 SPL, CS18 SPL-E, and CS18 SPL-VLF.
Components
The system comprises a transportable anechoic chamber, vibration control system SRS-35, software CS18 FF, power amplifier PA 14-180, and various accessories. It includes reference and working standards like ½" condenser microphone cartridges and sound acoustic calibrators.
Specifications
The system operates under specific environmental conditions (temperature 23°C ± 2°C, relative humidity 30%-75%, and environmental noise LZeq < 60 dB). The anechoic chamber dimensions are specified, and the system requires a space of 2.00 m x 2.00 m x 6.00 m for operation. Calibration is performed using a comparison with a reference standard and substitution method. Recommended sound pressure levels and frequency ranges with expanded uncertainty are detailed for both measuring microphones and sound level meters.
Electrical Tests
The system provides electrical input signals for tests according to IEC 61672-3, ED1, with specified input signal and expanded uncertainty for sound level meters and measuring systems.
See more

Catalog excerpts

SQ-101-1

CS18 FF Calibration System Free Field  Secondary calibration of measuring microphones, sound level meters and other acoustic sensors according to IEC 61094-8 and IEC 61672  Periodic single qualification according to IEC 61672-3  Acoustical measurements like: - Measurement of directivity characteristic - Determination of diffuse-field sensitivity - Measurement of acoustic emission of small objects  Calibration of constructively mechanically nonstandard microphones, e.g. external microphone units, optionally in axial and radial direction of measurement  Reference standards traceable to Physikalisch Technische Bundesanstalt (PTB) Braunschweig by the SPEKTRA Calibration Laboratory D-K-15183-01-00 (DAkkS-calibration certificate)  True free-field calibration in acoustically dead (anechoic) chamber  Calibration of any measuring microphone (condenser, electrets, electro-dynamic etc.) with any conctruction with / without protection grid  Supply of a defined free-field sound pressure level for the calibration of sound level meters  Calibration of acoustic calibrators  Can be combined with the following SPEKTRA calibration systems, e.g.:  CS18 SPL (calibration system pressure chamber)  CS18 SPL-E (calibration system pressure chamber for dosimeter)  CS18 SPL-VLF (calibration system sound pressure level very low frequency) Range of Use  Certified calibration laboratories  Measuring instrument verification in research and industry, for example civil engineering, aviation and automotive engineering  Quality assurance in manufacturing of microphones and sound level meters

 Open the catalog to page 1
SQ-101-2

CS18 FF Calibration System Free Field Components:  Dedicated transportable acoustically dead (anechoic) chamber by SPEKTRA, completely lined with wedge-shaped absorbers, with loudspeaker, for alternately holding the reference standard and test object, with small window for reading off the indications of compact sound level meters without electrical output channel  Vibration control system SRS-35 by SPEKTRA  Software CS18 FF with corresponding operation modes  Power amplifier PA 14-180 by SPEKTRA  Boom (holder fixture for DUT’s), accessories, cables and standard-PC  Reference standards -...

 Open the catalog to page 2

All SPEKTRA Schwingungstechnik und Akustik GmbH Dresden catalogs and technical brochures

  1. DRE-01

    4  Pages

  2. SQ-4.2

    2  Pages

  3. SQ-4.1

    2  Pages

  4. S-TEST/16 Lab

    2  Pages

  5. S-TEST/16 Fab

    2  Pages

  6. S-TEST/2

    2  Pages

  7. FDG-15

    2  Pages

Archived catalogs

  1. SE-10

    3  Pages

  2. PA14-180

    2  Pages

  3. VCS 401

    2  Pages

  4. PA14-500

    2  Pages

  5. CS18 SPL-VLF

    2  Pages

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.