1. Catalogs
  2. Specac Ltd
  3. Short Pathlength Gas Cell

Short Pathlength Gas Cell

Short Pathlength Gas Cell

Short Pathlength Gas Cell

Product catalog summary
Overview: The Storm™ 10 series gas cells are designed for measuring infrared spectra of concentrated gases, gas mixtures, and vapors at ambient temperatures and pressures. They are compatible with modern FT-IR spectrometers and feature a 10cm fixed pathlength design that is user demountable for easy maintenance.
Specifications:
  • Pathlength: 10cm
  • Volume: 132cm3 (Pyrex), 114cm3 (Stainless Steel)
  • Pressure Range: Vacuum to ambient
  • Temperature Range: Ambient
  • Body Material: Pyrex or Stainless Steel
  • Window Seals: Neoprene
  • Clear Aperture: 38mm
  • Mount: Standard 3"x 2"
Applications: Suitable for analysis of concentrated gases and vapors, residual solvent vapor analysis, and liquid headspace analysis.
Product Highlights:
  • Demountable design for ease of maintenance
  • Choice of cell body and window materials
  • Twin taps for evacuation, static, or flow-through operation
Ordering Information: Includes options for Pyrex and Stainless Steel gas cell bodies, cell mounts, and various window materials such as NaCl, KBr, CaF2, BaF2, and ZnSe.
Storm™ 10H Heated Gas Cell: This variant allows for heated operation up to 250°C, suitable for decomposition studies of solids and liquids. It features a stainless steel chamber, low voltage heaters, and a high stability temperature controller.
Standard Features of Storm™ 10H:
  • Heated operation up to 250ºC
  • Temperature control to +/- 1ºC
  • Demountable design for ease of cleaning
  • CE compliant
Specifications for Storm™ 10H:
  • Volume: 114cm3
  • Temperature Range: Ambient to 250°C
  • Body Material: Stainless Steel
  • Seals: Silicon
  • Heaters: 36 Volts / 120 Watts
  • Thermocouple: K-type NiCr/NiAl
  • Temperature Controller: Specac Series 4000
Ordering Information for Storm™ 10H: Includes options for heated gas cell body with controller and various window materials.
Contact Information: Specac Ltd. in the UK and Specac Inc. in the USA provide contact details for sales and support.
See more

Catalog excerpts

Short Pathlength Gas Cell-1

brilliant spectroscopy Storm™ 10 - Short Pathlength Gas Cell The Storm™ 10 series gas cells are the ideal choice for measuring Applications flexibility and chemicals compatibility is ensured through a concentrated gases, gas mixtures range of body materials (Pyrex or and vapors (>1% levels) at ambient Stainless Steel) and a wide choice of temperatures and pressures. Cells window materials. are suitable for operation in all modern FT-IR spectrometers using Gases can be analysed in static or flow Product Highlights • 10cm fixed pathlength single pass design • Demountable design for ease of maintenance The 10cm fixed pathlength design is modes, and stopcock valves are pro- • Choice of cell body materials vided for evacuation, sample introduc- • Choice of window materials user demountable allowing for easy • Twin taps for evacuation, static or flow-through operation Applications • Analysis of concentrated gases & vapors • Residual solvent vapor analysis • Liquid head space analysis Ordering Information 10cm Pyrex Gas Cell body 10cm Stainless Steel Gas Cell body Cell Mount for 10cm Gas Cells For a complete Storm™ 10 Gas Cell system, please order gas cell body, cell mount and a pair of windows of your choice. Spares and Consumables GS5040 Seal Kit for Pyrex Gas Cell GS5050 Stopcock for Pyrex Gas Cell GS5803 Vacuum Valve for Stainless Steel Gas Cell GS5804 Seal Kit for Stainless Steel Gas Cell Window Options GS5020 Pair of NaCl windows GS5021 Pair of KBr windows GS5022 Pair of CaF2 windows GS5023 Pair of BaF2 windows GS5096 Pair of ZnSe windows Options GS28001 GS28001 GS28002 RS232 Connection kit USB Connection kit RS485 Connection kit Pressure range: Temperature range: Ambient Body material: Pyrex or Stainless Steel Window seals: Clear aperture:

 Open the catalog to page 1
Short Pathlength Gas Cell-2

Storm™ 10H - Short Pathlength Heated Gas Cell The Storm™ 10H Heated Gas Cell is a single pass, 10cm fixed pathlength design, for the measurement of the infrared spectra of concentrated gases (> 1% levels), gas mixtures and vapors at temperatures up to 250°C. Solid and liquid decomposition studies can also be performed and the cell can be operated in static or flow-through modes. The cell consists of a stainless steel evacuable chamber with an injection septum sample introduction port, thermocouple port and vacuum valve. Windows are secured using threaded end caps and chemically resistant seals....

 Open the catalog to page 2

All Specac Ltd catalogs and technical brochures

  1. Benchmark

    1  Page

  2. PEarl

    2  Pages

  3. Liquid Pack

    3  Pages

  4. Quest ATR

    2  Pages

  5. Specacabinet

    2  Pages

  6. Specamill

    1  Page

  7. Pellet Dies

    3  Pages

  8. Specadie

    1  Page

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.