1. Catalogs
  2. Specac Ltd
  3. Heatable Transmission Cells (Static and Flow)

Heatable Transmission Cells (Static and Flow)

Heatable Transmission Cells (Static and Flow)

Heatable Transmission Cells (Static and Flow)

Product catalog summary
Overview: This datasheet provides detailed specifications and ordering information for Specac's heatable solids and liquids cells, designed for temperature-controlled analysis of samples.
Heatable Solids Cells:
  • Constructed from stainless steel, suitable for temperatures ranging from 250°C to -190°C.
  • Sample diameters range from 12 to 30 mm, with a maximum thickness of 8 mm.
  • Optional thermocouple for temperature measurement.
Heatable Liquids Cells:
  • Stainless-steel construction, suitable for temperatures from 250°C to -70°C.
  • Available in static or flow configurations, with Luer fittings or Swagelok connections.
  • Pathlengths range from 6 to 1000 μm, with a variety of window materials covering UV to mid-IR spectral ranges.
  • Demountable windows for easy cleaning and replacement.
Heatable Transmission Cells:
  • Available for both solid and liquid samples, compatible with various cell holders.
  • Can be used with Specac's advanced transmission cell holders.
Compatible Cell Holders:
  • Variable Temperature (VT) Cell: For temperatures as low as -190°C and up to 250°C.
  • Electrical Heating Jacket: For ambient to 250°C.
  • Water Heating Jacket: For ambient to 90°C using circulating fluid.
  • Ambient Cell Holder: Basic mount for heatable cells.
Liquid Cell Volume Guide:
  • Provides approximate cell volumes for various spacer thicknesses, ranging from 1.60 μl to 270.0 μl.
Ordering Information:
  • Options for cell configuration, holder compatibility, window material, and pathlength.
  • Part numbers provided for different configurations and materials.
Spares and Accessories:
  • Replacement window sets and spacers available in various materials and sizes.
  • Essential spares kits for static and flow cells.
See more

Catalog excerpts

Heatable Transmission Cells (Static and Flow)-1

Product Datasheet Heatable Transmission Cells (Static and Flow) Heatable transmission cells for both solid and liquid sampling. The liquid cells can be specified for both static and flowing sample conditions. The cells may be used in conjunction with a 3”x2” slide mounted holder or fitted to any of Specac’s advanced transmission cell holders. Compatible Cell Holders Heatable Liquids Cells y Designed to be heated or cooled Stainless-steel cells suitable for heating/cooling from 250 to -70 °C. An additional measuring thermocouple (not included) can be fitted if required. Flow or Static Static cells are supplied with a Luer fitting for easy sample injection. Flow cells are fitted with 1/16” Swagelok connections for use with sample pumping systems. Extensive selection of window materials Our range of windows cover spectral ranges from the UV, to near- and mid-IR. Demountable windows The windows and spacers are removable for cleaning or replacement Variable Temperature (VT) Cell (GS21525) Vacuum-jacketed cooling chamber for analysis at temperatures as low as -190 °C, with electric heating for analysis up to 250 °C. Electrical Heating Jacket (GS20730) Heated cell holder for analysis from ambient temperature up to 250 °C. Water Heating Jacket (GS20710) Heated cell holder for analysis from ambient temperature up to 90 °C using a circulating fluid for heating. Ambient Cell Holder (GS20740) Basic 3”x2” inch cell mount for the heatable cell range. Heatable Solids Cells Liquid Cell Volume Guide y Designed to be heated or cooled Stainless-steel cells suitable for heating/cooling from 250 to -190 °C. An additional measuring thermocouple (not included) can be fitted if required. Variable aperture size Sample diameters from 12 to 30 mm, up to 8 mm thick. Electrical Heating Jacket: 12 to 28 mm diameter, up to 3 mm thick Approximate cell volume for given spacer: 0.006 mm: 1.60 μl

 Open the catalog to page 1
Heatable Transmission Cells (Static and Flow)-2

Product Datasheet Ordering Information Heatable Liquids Transmission Cell 1. Select Cell Configuration: Flow Condition Fittings Luer Adaptor 2. Check Holder Compatibility: 1/16 inch Pipe with 1/16 inch Swagelok Swagelok Union Fitting Electrical Heating Jacket (GS20730) Water Heating Jacket (GS20710) Ambient Cell Holder (GS20740) 3. Choose Window and Select Part Number: NaCl GS20510 Notes: * Requires flow tubes (GS20080) for use with the VT cell. Pathlength (mm) 0.006, 0.012, and 0.025 mm spacers are made from Mylar. All others are made from PTFE. Heatable Solid Transmission Cell Ambient Cell...

 Open the catalog to page 2

All Specac Ltd catalogs and technical brochures

  1. Benchmark

    1  Page

  2. PEarl

    2  Pages

  3. Liquid Pack

    3  Pages

  4. Quest ATR

    2  Pages

  5. Specacabinet

    2  Pages

  6. Specamill

    1  Page

  7. Pellet Dies

    3  Pages

  8. Specadie

    1  Page

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.