Product catalog summary
Utility Requirements:
  • Compressed Air: 5.5 barg (80 psig)
  • Cooling Water: 19-30 Lpm (5-8 gpm)
  • Inert Gas for Backfilling and Vacuum Release: 0.21-0.34 barg (3-5 psig)
  • Inert Gas Consumption for Full Vacuum Release: 860 L @ 1 ATM
  • Steam for Sterilization In Place (SIP): 40.82 kg/hour @ 1.38-1.72 barg
  • Clean In Place (CIP) Water/Media: 150 Lpm @ 4.14 barg
  • Cooling Water Heat Removal: 41.03 kW/hr
  • Heat Output (Room), Peak: 2.93 kW
Key Features:
  • High process accuracy and reliability
  • Line of Sight™ approach for cycle development and scale-up
  • Conforms to good instrumentation practices
  • Optional LyoFLux* TDLAS instrument and ControLyo® Nucleation Technology
Specifications:
  • Lowest Shelf Temperature: -60 °C
  • Shelf Temperature Control Range: -55 °C to 65 °C
  • Shelf Temperature Control Accuracy: ± 0.5 °C
  • Shelf Pull-Down Time: ≤ 25 minutes
  • Lowest Condenser Temperature: -65 °C
  • Condenser Surface Area: 1.85 m²
  • Maximum Sublimation Rate: 2,050 g/hour
  • Compressor Horsepower: 10 hp
  • System Refrigerant: R410A
  • Vacuum Time to 100 mTorr: ≤ 20 minutes
Electrical Requirements:
  • Voltage: 200–250 VAC, 360–440 VAC, 420–500 VAC
  • Hertz: 50 Hz / 60 Hz
  • Phase: 3 Φ
  • Breaker Amperage: 40-70 A
ControLyo® Requirements (Optional):
  • Volume to Vent Percentage Ratio (Vv/Vr): < 6.5%
Dimensional Data:
  • Width: 147 cm
  • Length: 277 cm
  • Height: 199 cm
  • Maximum Weight: 3175.1 kg
Materials of Construction:
  • Product Chamber, Shelves, and Condenser Chamber: AISI 316L stainless steel
Options / Other Considerations:
  • ControLyo® Nucleation Technology
  • LyoFlux* 200 TDLAS Mass Flow Sensor
  • Independent CIP Skid
  • Tempris* Wireless Temperature Sensors
  • Virtual Servers & Thin Clients
See more

Catalog excerpts

LyoConstellation™ S20-1

LyoConstellation™ S20 Pilot Lyophilizer Key Features • Unmatched process accuracy and reliability • Deliver Line of Sight™ approach in cycle development & scale up using same technologies from R&D to production • Designed to conform to good instrumentation practices • Optional LyoFLux* TDLAS instrument with integrated communication with SP Hull LyoS™ 2.0 controls system software • Optional ControLyo® Nucleation Technology for improved product homogeneity and process efficiency Specifications Lowest Shelf Temperature (50 Hz / 60 Hz) Shelf Temperature Control Range Shelf Temperature Control Accuracy Shelf Pull-Down From 20 °C To -40 °C Shelf Temperature Uniformity Lowest Condenser Temperature Condenser Surface Area Condenser Pull-Down From 20 °C To -45 °C SP Hull LyoConstellation S20 Pilot Lyophilizer Utility Requirements Compressed Air Compressor Horsepower System Refrigerant Inert Gas For VLC/ Vacuum Release Inert Gas Consumption (Full Vacuum Release) Steam For Sterilization In Place (SIP) 40.82 kg/hour (90 lbs/hour) @ 1.38 -1.72 barg (20 -25 psig) Clean In Place (CIP) Water / Media Cooling Water Heat Removal Heat Output (Room), Peak Note: Performance specifications are based on SP test data from clean, dry and empty (CDE) units operating at an ambient room temperature of approximately 20° C. SP recommends an optimum operating range of 15-25 °C (59-77 °F) with an RH of ≤ 80 % at sea level. Electrical Requirements Voltage Breaker Amperage ControLyo® Requirements (Optional) Volume To Vent Percentage Ratio (Vv/Vr) Note: Vent Volume (Vv) is the product chamber volume multiplied by 10. The volume of a standard SP Hull LyoConstellation S20 chamber is 0.51 m3 (18 ft3). Vr is the volume of the room in which the lyop

 Open the catalog to page 1
LyoConstellation™ S20-2

LyoConstellation™ S20 Pilot Lyophilizer 78 FRAME HEIGHT 103 OVERALL HEIGHT Dimensional Data Shelf Configuration Shelf Area Shelf Clearance 5 Usable Shelves (1 Radiant) Height Height of Stoppering Ram Maximum Weight Shelf Dimensions (W × L) Shelf Loading Height Note: LyoFLux* TDLAS installation increases overall length by 61.0 cm (24 in). SP recommends a 60.7 cm (24 in) clearance around all sides of the unit for serviceability. Refrigeration Information Product Chamber Shelf Isolation Valve Disc AISI 316L stainless steel AISI 316L stainless steel AISI 316L stainless steel Condenser Chamber AISI...

 Open the catalog to page 2
*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.