TO2-1L

TO2-1L

Product catalog summary
Product Overview:
The TO2-1L PPM Oxygen Sensor is an electrochemical galvanic fuel cell sensor designed for trace oxygen measurement. It operates within a range of 0.01 PPM to 25.0% oxygen concentration, offering excellent stability, performance, and linearity. The sensor is self-contained, requiring no additional electrolyte, and is suitable for both continuous and spot-checking analysis.
Specifications:
  • Sensor Technology: Micro Fuel Cell
  • Signal Output: 285 - 590 uA
  • Measuring Range: 0 - 10 PPM (Min), 0 - 10000 PPM (Max)
  • Response Time (T90): 13 Seconds
  • Accuracy Full Scale: ± 1% of Signal
  • Repeatability: ± 0.5% F.S.
  • Temperature Coefficient: 2.54 % / ºC
  • Operating Temperature: 0 to 50 ºC
  • Storage Temperature: 0 to 45 ºC
  • Recommended Flow: 0.5 - 5 SCFH
  • Humidity: 0 - 100% RH (Non-Condensing)
  • Expected Life: 24 Months
  • Storage Life: 12 Months
  • Warranty: 12 Months
  • PCB Connection: Center Foil Negative, Outer Foil Positive
Standards and Compliance:
The sensors are tested and validated to BSI Standard BS EN 50104:2010. They are designed, tested, and assembled in California, USA.
Additional Information:
Southland Sensing Ltd. provides a warranty for the sensors to be free from defects in materials and workmanship for the specified period. The company is not liable for damages due to customer neglect.
See more

Catalog excerpts

TO2-1L-1

Product Specification Sheet Electrochemical Galvanic Fuel Cell Trace Oxygen Sensor Measures Oxygen From 0.01 PPM to 25.0% Electrochemical Fuel Cell Sensor Technology Self-Contained, No Need To Add Electrolyte Excellent Stability, Performance & Linearity Continuous And Spot Checking Analysis Specifications: Sensor Technology Signal Output1 Measuring Range Response Time T90 Accuracy Full Scale2 Repeatability F.S. Temp Coefficient Operating Temp Storage Temp Recommended Flow Humidity Expected Life3 Storage Warranty4 PCB Connection Micro Fuel Cell 285 - 590 uA 0 - 10 PPM (Min) 0 - 10000 PPM (Max) 13 Seconds ± 1% of Signal ± 0.5% 2.54 % / º C 0 to 50 º C 0 to 45 º C 0.5 - 5 SCFH 0 - 100% RH (Non - Condensing) 24 Months 12 Months 12 Months CenterFoilNegative Outer Foil Positive Sensing Surface 1. Signal Output is measured in air at 25 º C, Sea Level. 2. Full Scale Accuracy is calculated with constant pressure, temperature, and proper calibration (80% O2 Value on Full Scale Range or a range higher). Drastic Temperature change can result in a maximum error of ± 10%. 3. Expected life is calculated when O2 < 1000 ppm @ 25 º C, Sea Level. 4. Southland Sensing Ltd. warranties the sensors for the period noted above to be free from defects in materials and workmanship. SSO2 will not be held liable for sensors damaged due to customer neglect. 4. Southland Sensing Ltd. sensors are tested and validated to BSI Standard BS EN 50104:2010 Rev 1.03 November 12th, 2020_DF Designed, Tested, and Assembled in California, USA 4045 E. Guasti Rd. #203 Ontario, CA 91761 USA : 1-949-398-2879 : [email protected] : www.sso