PO2-24

PO2-24

Product catalog summary
Product Overview:
The PO2-24 Percent Oxygen Sensor is an electrochemical galvanic fuel cell sensor designed to measure oxygen levels from 1000 PPM to 25.0%. It is self-contained, requiring no additional electrolyte, and is suitable for use in CO2 and natural gas applications. The sensor offers excellent stability, performance, and linearity, making it ideal for both continuous and spot-checking analysis.
Specifications:
  • Sensor Technology: Micro Fuel Cell
  • Signal Output: 34 - 63 uA
  • Measuring Range: 0 - 100 Percent
  • Response Time (T90): 13 Seconds
  • Accuracy Full Scale: ± 1% of Signal
  • Repeatability Full Scale: ± 0.5%
  • Temperature Coefficient: 2.54 % / ºC
  • Operating Temperature: 0 to 50 ºC
  • Storage Temperature: 0 to 45 ºC
  • Recommended Flow: 0.5 - 5 SCFH
  • Humidity: 0 - 100% RH (Non-Condensing)
  • Expected Life: 25 Months
  • Storage Life: 3 Months
  • Warranty: 12 Months
  • PCB Connection: Center Foil Negative, Outer Foil Positive
Standards and Compliance:
The sensors are tested and validated to BSI Standard BS EN 50104:2010 and are designed, tested, and assembled in California, USA.
Additional Information:
Southland Sensing Ltd. provides a warranty for the sensors to be free from defects in materials and workmanship for the specified period. The company is not liable for damages due to customer neglect.
See more

Catalog excerpts

PO2-24-1

Product Specification Sheet Measures Oxygen From 1000 PPM to 25.0% Sensor Technology Micro Fuel Cell Measuring Range 0 - 100 Percent Electrochemical Fuel Cell Sensor Technology Self-Contained, No Need To Add Electrolyte Excellent Stability, Performance & Linearity Continuous And Spot Checking Analysis Response Time T90 Accuracy Full Scale1 Repeatability F.S. Temp Coefficient Operating Temp Storage Temp Recommended Flow Humidity Expected Life2 Storage Warranty3 PCB Connection 13 Seconds ± 1% of Signal ± 0.5% 2.54 % / ° C 0 to 50 ° C 0 to 45 ° C 0.5 - 5 SCFH 0 - 100% RH (Non - Condensing) 25 Months 3 Months 12 Months CenterFoil Negative Outer Foil Positive For Use In CO2 & Natural Gas Applications 1. Signal Output is measured in air at 25 ° C, Sea Level. Full Scale Accuracy is calculated with constant pressure, temperature, and proper calibration (80% O2 Value on Full Scale Range or a range higher). Drastic Temperature change can result in a maximum error of ± 10%. Expected life is calculated when O2 < 1000 ppm @ 25 ° C, Sea Level. Southland Sensing Ltd. warranties the sensors for the period noted above to be free from defects in materials and workmanship. SSO2 will not be held liable for sensors damaged due to customer neglect. 4. Southland Sensing Ltd. sensors are tested and validated to BSI Standard BS EN 50104:2010 Designed, Tested, and Assembled in California, USA 4045 E. Guasti Rd. #203 Ontario, CA 91761 USA : 1-949-398-2879 : [email protected] : www.sso2.com

 Open the catalog to page 1
*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.