PO2-1x

PO2-1x

Product catalog summary
Product Overview
The PO2-1x Percent Oxygen Sensor is an electrochemical galvanic fuel cell sensor designed to measure oxygen levels from 1000 PPM to 100%. It features self-contained technology, eliminating the need for additional electrolytes, and offers excellent stability, performance, and linearity.

Specifications
  • Sensor Technology: Micro Fuel Cell
  • Signal Output: 245 - 430 uA
  • Measuring Range: 0 - 100 Percent
  • Response Time (T90): 7 Seconds
  • Accuracy Full Scale: ± 1% of Signal
  • Repeatability Full Scale: ± 0.5%
  • Temperature Coefficient: 2.54 % / ºC
  • Operating Temperature: 0 to 50 ºC
  • Storage Temperature: 0 to 45 ºC
  • Recommended Flow: 0.5 - 5 SCFH
  • Humidity: 0 - 100% RH (Non-Condensing)
  • Expected Life: 12 - 15 Months
  • Storage Life: 6 Months
  • Warranty: 9 Months
  • PCB Connection: Center Foil Negative, Outer Foil Positive

Performance and Compliance
The sensor is designed, tested, and assembled in California, USA, and complies with BSI Standard BS EN 50104:2010. It is warranted by Southland Sensing Ltd. to be free from defects in materials and workmanship for the specified period, with limitations on liability for customer neglect.

Contact Information
Southland Sensing Ltd., 4045 E. Guasti Rd. #203 Ontario, CA 91761 USA. Phone: 1-949-398-2879. Email: [email protected]. Website: www.sso2.com.

Revision Information
Rev 1.03, November 12th, 2020_DF
See more

Catalog excerpts

PO2-1x-1

Product Specification Sheet Measures Oxygen From 1000 PPM to 100% Sensor Technology Micro Fuel Cell Measuring Range 0 - 100 Percent Electrochemical Fuel Cell Sensor Technology Self-Contained, No Need To Add Electrolyte Excellent Stability, Performance, & Linearity High Output Percent Oxygen Sensor Response Time T90 Accuracy Full Scale1 Repeatability F.S. Temp Coefficient Operating Temp Storage Temp Recommended Flow Humidity Expected Life2 Storage Warranty3 PCB Connection 7 Seconds ± 1% of Signal ± 0.5% 2.54 % / ° C 0 to 50 ° C 0 to 45 ° C 0.5 - 5 SCFH 0 - 100% RH (Non - Condensing) 12 - 15 Months 6 Months 9 Months CenterFoil Negative Outer Foil Positive 1. Signal Output is measured in air at 25 ° C, Sea Level. Full Scale Accuracy is calculated with constant pressure, temperature, and proper calibration (80% O2 Value on Full Scale Range or a range higher). Drastic Temperature change can result in a maximum error of ± 10%. Expected life is calculated when O2 < 1000 ppm @ 25 ° C, Sea Level. Southland Sensing Ltd. warranties the sensors for the period noted above to be free from defects in materials and workmanship. SSO2 will not be held liable for sensors damaged due to customer neglect. 4. Southland Sensing Ltd. sensors are tested and validated to BSI Standard BS EN 50104:2010 Designed, Tested, and Assembled in California, USA 4045 E. Guasti Rd. #203 Ontario, CA 91761 USA : 1-949-398-2879 : [email protected] : www.sso2.com