OMD-150-NG Hazardous Area Online Process Oxygen Analyzer
2Pages

{{requestButtons}}

Catalog excerpts

OMD-150-NG Hazardous Area Online Process Oxygen Analyzer - 1

MEASURE. ANALYZE. CONTROL. Product Specification Sheet Oxygen Analyzer Designed for the Natural Gas Industry Class 1, Div 2 Groups B,C,D Full Scale Range as low as 0 - 10 ppm Precision Fuel Cell Oxygen Sensor Technology Large Backlight Display w/ Auto-Range Measure Oxygen from 0.01ppm to 25% Intuitive User Friendly Interface Cost Effective and Low Maintenance MODBUS RS485 ASCII (Optional) 2 - Wire Loop Configuration (Optional) Specifications: Accuracy: Analyzer Range: Optional Range: Area Classification: Dimensions: Flow: Gas Connections: Output: Power: Pressure Response Time: Sample System: Sensor: Temperature: Warranty Sensor: Warranty Electronics: Weight: < +/-1% Full Scale Range12 0 - 10/100/1000/10000ppm/25% 0 - 1%/5%/10%/25%/100% Class 1, Div 2, Groups B, C, D 11.25" x 12.0" x 5.50" 0.5 - 5.0 SCFH 1/4" Compression Tube Isolated 4 - 20mA or 1 - 5 VDC 12 - 24VDC 0.1 - 50 PSIG Inlet, vent to atm T90in 10 Seconds Optional Precision Fuel Cell 0 to 50 deg C 12 Months 12 Months 12.5 lbs *Accuracy at constant conditions Optional Configurations: - Customizeable Measurement Ranges - Sample Conditioning Systems - Bi-Directional RS485 Modbus ASCII - Extreme Weather Packaging Applications: • Natural Gas Extraction & Pipelines • Natural Gas Processing • Acid (CO2) Gas Streams • Inert, Hydrocarbon, Hydrogen Processing And Many Other Industrial Applications "Inquiry for Application Expertise" Designed, Tested and Assembled in California, USA 4045 E. Guasti Rd. #203 Ontario, CA 91761 USA : 1-949-398-2879 : sales@sso2.com : www.sso2.com

Open the catalog to page 1
OMD-150-NG Hazardous Area Online Process Oxygen Analyzer - 2

OMD-15Q-NG Oxygen Analyzer Product Specifications Oxygen Sensor Technology: The model OMD-150-NG oxygen analyzer combines a rugged design with SSO2's precision oxygen sensors. The result is a highly reliable and cost effective compact design with easy-to-use user interface designed specifically for the Natural Gas Industry. The oxygen analyzer is designed to meet standards for Class 1, Div 2, Groups B,C,D installation. The oxygen sensor used in the OMD-150-NG is based on the galvanic electrochemical fuel cell principal. All oxygen sensors are manufactured in house by Southland Sensing Ltd....

Open the catalog to page 2

All Southland Sensing Ltd. catalogs and technical brochures

  1. EMD-482

    2 Pages

  2. OMD-501X

    2 Pages

  3. OMD-645

    1 Pages

  4. EMD-485M

    2 Pages

  5. TO2-1L

    1 Pages

  6. TO2-IT

    1 Pages

  7. TO2-IT-A

    1 Pages

  8. TO2-233

    1 Pages

  9. TO2-133

    1 Pages

  10. TO2-2x

    1 Pages

  11. TO2-1x

    1 Pages

  12. PO2-1x

    1 Pages

  13. PO2-13

    1 Pages

  14. PO2-2x

    1 Pages

  15. PO2-160

    1 Pages

  16. PO2-24

    1 Pages

  17. PO2-1120

    1 Pages

  18. TO2-1xL

    1 Pages

  19. TO2-19

    1 Pages

  20. TO2-233-T

    1 Pages

  21. TO2-238-T

    1 Pages

  22. OMD-625

    2 Pages

Archived catalogs

  1. PO2-132

    1 Pages

  2. PO2-133

    1 Pages

  3. PO2-15

    1 Pages

  4. PO2-154

    1 Pages

  5. PO2-253

    1 Pages