Product catalog summary
Applications: The EMD-485M Oxygen Transmitter is suitable for various applications including inert glove box systems, nitrogen and O2 PSA generators, laboratories, universities, beverage grade CO2 monitoring, and air separation plants.
Specifications: The oxygen sensor offers an accuracy of less than 1% full scale range under constant conditions. Calibration is recommended every 2-3 months. It is classified as general purpose with dimensions of 3.25 x 2.00 inches for the PCB. The sensor operates within a temperature range of 0 - 50°C and includes integral temperature compensation. Output options include 4-20mA, 0-1VDC, and 0-10VDC. Sensor housing options are H6 or H3, with flow sensitivity between 0.5 - 5.0 SCFH. The sensor type is a precision fuel cell, with a 12-month warranty for both the sensor and electronics.
Configurations: The transmitter supports user-selectable ranges from 0-10ppm to 100%, with digital push buttons for local span calibration. It features a graphical display with backlight and various sensor housing options. Additional customized options are available.
Product Features: The EMD-485M is a configurable OEM oxygen transmitter with a 6-wire MODBUS RS485 ASCII interface. It is designed, tested, and assembled in California, USA. The transmitter is well-suited for OEM installations and uses precision fuel cell oxygen sensor technology. It can measure oxygen from 0.1ppm to 100% and offers outputs of 4-20mA, 0-1V DC, and 0-10V DC. The design includes an updated microprocessor and bi-directional RD485 MODBUS ASCII communication.
Ordering Information: The transmitter can be ordered with various options including trace or percent analysis sensors, optional local display, and different sensor housings. Input power requirements are 12-24 V DC with a maximum current draw of 40 mA. For assistance with ordering, contact the inside sales specialist.
Additional Notes: The oxygen sensors are manufactured by Southland Sensing Ltd. and are unaffected by background gases such as H2, He, or hydrocarbons. Acidic cells are suitable for environments with acid gases like CO2 or natural gas. The sensors require minimal maintenance and offer excellent performance, accuracy, and stability.
See more

Catalog excerpts

EMD-485M-1

EMD-485M Oxygen Transmitter Product Specification Sheet Configurable OEM Oxygen Transmitter, 6-wire, MODBUS RS485 ASCII Well Suited for OEM Installations Precision Fuel Cell Oxygen Sensor Technology Trace or Percent Configuration Measure Oxygen from 0.1ppm to 100% 4 - 20mA, 0 - 1V DC, 0 - 10V DC Outputs Bi-Directional RD485 MODBUS ASCII Updated Microprocessor Based Design Cost Effective, Compact & Low Maintenance Configurations: H3 Flow Through Sensor Housing Specifications: Oxygen Sensor User Selectable Ranges (Pre-Configured): 0 - 10ppm/100/1000ppm/1%/5%/10%/25%/100% Digital Push Buttons to Perform Local Span Cal Sensor Housings: H6 KF-40 or H3 Flow Through Graphical Display with Backlight Many additional customized options available Integral Span Pot to Adjust 4mA Offset Inert Glove Box Systems Nitrogen and O2 PSA Generators Laboratories & Universities Beverage Grade CO2 Monitoring Air Separation Plants & Many Others Accuracy: < 1% Full Scale Range* Calibration: 2 - 3 months or as needed Classification: General Purpose Dimensions (PCB): 3.25 x 2.00 inch Display: Optional Temperature: 0 - 50 deg C Temperature Compensation: Integral Output: 4 - 20mA, 0 - 1VDC, 0 - 10VDC Sensor Housing: Optional, H6 or H3 Flow Sensitivity: 0.5 - 5.0 SCFH Sensor Type: Precision Fuel Cell Warranty: 12 Months Sensor Warranty: 12 Months Electronics *Accuracy at constant conditions “Inquiry for Application Expertise” Designed, Tested, and Assembled in California, USA

 Open the catalog to page 1
EMD-485M-2

Product Specifications Oxygen Transmitter: Oxygen Sensor Technology: The model EMD-485M oxygen transmitter combines a rugged electronic design with SSO2’s precision oxygen sensors. The result is a highly reliable and cost effective compact design with an easy-to-use user interface. The oxygen sensor used in the EMD-485M is based on the galvanic electrochemical fuel cell principal. All oxygen sensors are manufactured in house by Southland Sensing Ltd. under a strict quality program. The transmitter comes with a variety of options, which makes it configurable for a large array of applications....

 Open the catalog to page 2