The Thermoset-KV is an online analyzer designed for continuous measurement of kinematic viscosity at a reference temperature. It utilizes vibration at resonance frequency technology, with a vibrating rod that adjusts its amplitude and frequency based on dynamic viscosity and density, respectively. The analyzer calculates real-time kinematic viscosity using a flow-cell where the product flows along the sensor.

Specifications
  • Measuring Range:
    • Viscosity: 0-500 cSt
    • Density: 0.6 -1.6 g/cc
  • Precision: 1% of measurement at operating point (50%-100% of Full Scale)
  • Operating Conditions:
    • Max Temperature: 200°C/390°F (higher available on request)
    • Reference Temperature: 40°C to 135°C/100°F to 275°F
    • Max Pressure: 16 bars/230 psi
  • Utilities:
    • Power Supply: 110 or 240 VAC, single phase, 50-60Hz
    • Electrical Consumption: below 100W
    • Compressed Air: 7 bar/100 psi, 0.5 m3/h / 0.3 SCFM
    • Chilled Water: Must be supplied by customer, temperature application dependent
  • Certifications: NEC Class 1 Div 2 and ATEX II 2 G Ex IIB or II 3 G Ex IIB

Results on Selected Product
The analyzer was tested on a viscosity standard rated at 182.6 cSt at 40°C. The measured value was 181.55 cSt with a standard deviation of 0.6%.
See more

Catalog excerpts

Thermoset-KV (kinematic viscosity) analyser at reference temperature-1

THERMOSET- KV PRELIMINARY BROCHURE DESCRIPTION The Thermoset-KV (Kinematic Viscosity) is an online analyzer designed to continuously measure kinematic viscosity at reference temperature. The measurement principle is based on the vibration at resonance frequency technology developed by Sofraser in 1981. The active part of the measurement is a vibrating rod driven by constant electrical power. The amplitude of the vibration varies according to dynamic viscosity and the frequency varies according to density. The analyzer uses a flow-cell in which the product flows along the sensor described above. Using the combined dynamic viscosity and density measurement, the electronic panel calculates real time kinematic viscosity of the product By adjusting coolant temperature of the flow-cell, the end-user can select the reference temperature. Schematic diagram of the Thermoset-KV

 Open the catalog to page 1
Thermoset-KV (kinematic viscosity) analyser at reference temperature-2

Measuring Range o Viscosity : 0-500 cSt o Density : 0.6 -1.6 g/cc Precision : 1% of measurement at operating point (between 50% and 100% of Full Scale) Operating conditions: o Maximum Temperature : 200°C/390°F (higher temperature available on request) o Reference Temperature : Between 40°C and 135°C/100°F and 275°F o Maximum Pressure : 16 bars/230 psi Utilities o Power Supply : 110 or 240 VAC, single phase, 50-60Hz o Electrical consumption below 100W o Compressed Air : 7 bar/100 psi, 0.5 m3/h / 0.3 SCFM o Chilled Water (must be supplied by customer, temperature application dependent) RESULTS...

 Open the catalog to page 2