1. Catalogs
  2. smartGAS Mikrosensorik GmbH
  3. Sulfur Dioxide SO2 2000 ppm
video corpo

Sulfur Dioxide SO2 2000 ppm

Sulfur Dioxide SO2 2000 ppm

Sulfur Dioxide SO2 2000 ppm

Product catalog summary
Product Overview
The FLOWEVO Infrared Gas Sensor for Sulfur Dioxide (SO2) is designed for applications such as emission monitoring, process control, lab analysis, and environmental analysis. It features a compact design, pre-calibration, and supports Modbus ASCII or RTU communication. The sensor operates with a supply voltage of 3.3–6 V DC and indicates status via an LED.
Specifications
  • Measurement Principle: Non-Dispersive Infra-Red (NDIR), dual wavelength
  • Measurement Range: 0 to 2000 ppm Full Scale (FS)
  • Flow Rate: 0.1 to 1.0 l/min
  • Mounting Dimensions: 156 mm x 30 mm x 50 mm
  • Warm-up Time: Less than 2 minutes for start-up, less than 30 minutes for full specification
Measuring Response
  • Digital Resolution: 1 ppm
  • Response Time: t90 ≤ 11.6 s, ton ≤ 15.9 s at 0.7 l/min
  • Detection Limit: ≤ 12 ppm
  • Repeatability: ≤ ± 15 ppm
  • Linearity Error: ≤ ± 20 ppm
  • Long Term Stability: Zero ≤ ± 90 ppm, Span ≤ ± 110 ppm over 1000 hours
Environmental and Electrical Parameters
  • Operating Temperature: 0 to +50 °C
  • Storage Temperature: -20 to +60 °C
  • Air Pressure: 800 to 1150 hPa
  • Humidity: 0 to 95% relative humidity (non-condensing)
  • Supply Voltage: 3.3 V to 6.0 VDC
  • Power Consumption: < 800 mW
Additional Information
The sensor requires a stable, particle-free, and clean sample gas with a dew point below +5°C. Calibration is recommended at first initiation and periodically thereafter. The product is made in Germany and all specifications are subject to change without notice.
See more

Catalog excerpts

Sulfur Dioxide SO2 2000 ppm-1

FLOWEVO Infrared gas Sensor Sulfur Dioxide SO2 2000 ppm smartGAS item number: F3-432205-05000 Pre calibrated Compact Design 3/5 mm gas line connector 3.3–6 V DC supply voltage Modbus ASCII or RTU Status indicated by LED Low drift Application examples Emission monitoring Process control Lab analysis Environmental analysis Available equipment Gas cooler Particle filter Gas pump Calibration Software Mounting equipment smartGAS Mikrosensorik GmbH I Huenderstr. 1 I 74080 Heilbronn I Germany T +49 7131 797553-0 I F +49 7131 797553-10 I [email protected] I www.smartgas.eu Available design in support Mechanical Installation Data communication Gas pre-treatment

 Open the catalog to page 1
Sulfur Dioxide SO2 2000 ppm-2

FLOWEVO I Sulfur Dioxide SO2 I F3-432205-05000 General features Measurement principle: Non-Dispersive Infra-Red (NDIR), dual wavelength Measurement range: by flow (nearly atmospheric pressure) Flow rate: Mounting dimensions: Warm-up time: < 2 minutes (start-up time) < 30 minutes (full specification) Measuring response* Digital resolution: Linearity error (straight line deviation): Long term stability (zero): Long term stability (span): Influence of T, P, flow rate, other* Temp. dependence (zero): Temp. dependence (span): Pressure dependence: Flow rate dependence: Cross sensitivity (zero) other...