AX-212223-43000
3Pages

{{requestButtons}}

Catalog excerpts

AX-212223-43000 - 1

ANAREX CEMS Analyzer AX-212223-43000 Multigas analyzer Easy operation via touch screen Easy sensor calibration Best measurement performance For installation in a 19" rack Analog or digital output Photoacoustic sensors NDIR sensors Paramagnetic sensors Technology Electro chemical sensors The ANAREX CEMS analyzer of smartGAS is a high-performance solution for CEMS and process control applications. It convinces with its highly accurate and stable measurement performance as well as its customerfriendly interfacing via touch screen along with a simple sensor calibration. The ANAREX is designed as a multi-gas analyzer and is suitable for installation in a 19" rack. The smartGAS analyzer can also use different technologies to fulfil all kinds of application requirement. The smartGAS Analyzer is equipped with a 5.6" touch screen, that displays the gas concentration as real-time reading. The internal software is designed to control zero, span and third point calibration to ensure a perfect accuracy of the measurement values. Application examples CEMS Process control Measuring principles Infrared NDIR Peripherals Gas cooler Pumps Particle filter Tubes Mounting equipment smartGAS Mikrosensorik GmbH I Huenderstr. 1 I 74080 Heilbronn I Germany T +49 7131 797553-0 I F +49 7131 797553-10 I sales@smartgas.eu I www.smartgas.eu Customizing options Measuring gases Detection ranges Design and Software Gas pre treatment

Open the catalog to page 1
AX-212223-43000 - 2

General features Measurement principle: Measurement range: Detection Limits: CO2 … NDIR CO … NDIR NO … NDIR SO2 … NDIR 0 … 20 Vol.-% (CO2) 0 … 1000 ppm (CO) 0 … 2000 ppm (NO) 0 … 1000 ppm (SO2) ≤ 0.23 Vol.-% (CO2) ≤ 16 ppm (CO) ≤ 30 ppm (NO) ≤ 12 ppm (SO2) ≤ ± 0.06 Vol.-% (CO2) ≤ ± 6 ppm (CO) ≤ ± 10ppm (NO) ≤ ± 4 ppm (SO2) ≤ ± 0.2 Vol.-% over 1000 h period (CO2) ≤ ± 20 ppm over 1000 h period (CO) ≤ ± 50 ppm over 1000 h period (NO) ≤ ± 20 ppm over 1000 h period (SO2) Basic information Display: Analogue output: 4-20 mA (in operation) 2 mA (Warm-up or alarm) Load resistance: Alarm output:...

Open the catalog to page 2
AX-212223-43000 - 3

Gas flow & Environment information Gas Inlet flow: (0.4 ~ 0.8) L/min (flow fluctuation ≤ 0.02 L/min) non-condensing Inlet dew point: 5°C ± 0.1°C 99.999 % N2 (NDIR) Clean air, free of sample gas (Photoacoustic) 75 % ~ 110 % of span point Clean air (0 ~ 25 % O2 measurement) Operation temperature: Typical values related to 1013 hPa, Ta = 22 °C, flow = 0.7 l / min for dry (not condensing) and clean sample gas. Stated values exclude calibration gas tolerance. All rights reserved. Any logos and/or product names are trademarks of smartGAS. The reproduction, transfer, distribution or storage of...

Open the catalog to page 3

All SmartGAS Mikrosensorik GmbH catalogs and technical brochures

  1. PAS

    4 Pages

  2. SILAREX

    4 Pages

  3. FLOW EVO

    2 Pages

  4. ANAREX

    4 Pages