OSA100 On-Stream-Slurry Analysis
6Pages

{{requestButtons}}

Catalog excerpts

OSA100 On-Stream-Slurry Analysis - 1

Skyray Instrument 50 Biaintree Hill Park, Suite 201. Braintree, MA USA 02184 Website: www.skyrayinstrument.com Skyray Instrument XRF On-stream Slurry Analyzer (OSA100) Detection report R_-a«miiw*/U Zinc <onc*i>ttWK UiMSK RM*nrMor% Detection limes

Open the catalog to page 1
OSA100 On-Stream-Slurry Analysis - 2

Evident economic value OSA100 XRF On-stream Slurry Analyzer Instead manual sampling analysis, therefore save analysis expense The reasonable use of floating medicine reduces cost OSA100 can carry on the real-time on-line monitoring for important operation in production process; in addition, it can provide the latest information and change tendency for operators or managers, so as to point out the change fluctuation, and thus make quick reply to rapid change, and succeed in instructing production. OSA100 is the key way for auto mineral processing, stable production process, reducing...

Open the catalog to page 2
OSA100 On-Stream-Slurry Analysis - 3

Product Configuration Working Principle The primary sampler can take part samples and send them to multiplexer, and then a few samples separated from the mdtiptexer wil be sent to sample editor analyzing. Adopting EDX Energy Dtspersrve X-Ray Spectrometer, Characteristics X-Ray of dement wil be emitted from the sample irradiated by primary X-Ray form X-Ray tube. The characteristics X-fley separated by detector wil be recorded intensity by electric device, and the dement content can be calculated based onthe intensity. Excellent Characteristics G Automation, High Inteligent Degree The system...

Open the catalog to page 3
OSA100 On-Stream-Slurry Analysis - 4

On-line analysis and manual analysis comparison(24 channles) Pressure Sampler Safe and Radiation-free The system of modulus design, high integration, and small volume is easy to operate and maintain. And using pneumatic and electric control working model, it runs stably, reliably and maintains easily without any pollution. Moreover, the system realizes computer automation, intelligent control and management. Maintenance is more or less needed. Easy Operation and Maintenance The system of modulus design, high integration, and small volume is easy to operate and maintain. And using pneumatic...

Open the catalog to page 4
OSA100 On-Stream-Slurry Analysis - 5

Secondary Sampling Secondary sampling is to send 20L/min primary sampling separated from the multiplexer box for analyzing. The multiplexer is a multi-function and automatic device combined channel flush and composition sampling. Designed with 6 channels, the expansion capacity can be up to 24 channels. The integrated composition sampler can realize sampling all the day. Multi-Channel Switch Converter Sampler can work efficiently for a long time under the stable and reliable pneumatic and electric control circuit. The integrated chamber structure exempts the maintenance. The unique mesh...

Open the catalog to page 5
OSA100 On-Stream-Slurry Analysis - 6

Requirement for The typical installation method of the system is of 2 layers layout, and the position H is not more than 1000mm. Primary sampling: 70120L/min Secondary sampling: About 0L/min Composition Sampling and Period: Typical routine sampling is 50ml, and the period is 10 minutes, adjustable. Measurement Time of strength: Set different measurement times for every channel. Typical value is 30 seconds. Measurement PeriodCalculate on the basis of measurement time of every channel. Taking 30 seconds as the basis, it takes 12-13 minutes to measure 12 samples current. Short-time...

Open the catalog to page 6

All Skyray Instrument catalogs and technical brochures

  1. HM-3000P

    2 Pages

  2. OES1000VMI

    4 Pages

  3. GC5400

    4 Pages

  4. OSA100

    6 Pages

  5. GC-MS 6800

    2 Pages

  6. LC-MS 1000

    2 Pages

  7. ICP-MS2000

    2 Pages

  8. genius 3000

    2 Pages

  9. genius 7000

    2 Pages

  10. genius 9000

    2 Pages

  11. genius 5000