Calistair R4000

Calistair R4000

Calistair R4000

Product catalog summary
Overview: The R4000 Mobile Air Decontamination Unit is designed for easy setup and operates silently with a flow rate of up to 4000 m3/h. It is available in two versions: Premium and COV, with optional over-pressure and negative pressure kits. Manufactured in France, it serves various sectors including clean rooms, industrial, childcare, retirement homes, food industry, and healthcare.
Specifications: The unit complies with standards such as EN 1886, NF B44-200, NF S 90-351, ISO 14644, and ISO 14698. It features a particulate cleanliness class of ISO 7 and ISO 8, with air change rates of 15 vol/h and 10 vol/h respectively. The differential pressure is maintained at 15 Pa ± 5 Pa.
Components: Key components include a user interface, supply grill with cooling fin, terminal filter (H14 or U15), management controller, adsorbent filter for VOCs, pressure probes, and a motor turbine EC. The unit also includes a non-thermal catalysis system for contaminant destruction.
Technical Specifications: The unit supports air flow from 800 to 4000 m3/h with a pressure availability of 450 Pa. It can treat zones from 50 to 400 m3 and surfaces from 20 to 150 m2. The unit is equipped with double skin panelling, noise insulation, and motorised impellers with electronic commutation.
Maintenance and Operation: Maintenance is straightforward, taking approximately 10 minutes, with easy bio-cleaning and incinerable filters. The unit is mobile, with connection possibilities to building management systems, and features alarms for over-pressure and under-pressure conditions.
Applications: Suitable for various environments such as surgery rooms, laboratories, pharmacies, patient rooms, and treatment areas. It effectively treats chemical gas pollutants including organic acids, aldehydes, alcohols, and aromatic substances.
Dimensions and Power: The unit measures 700 x 850 x 1940 mm and weighs 280 kg with filters. It operates on a power supply of 210 – 240 V, 50 Hz, single-phase.
See more

Catalog excerpts

Calistair R4000-1

HEALTHCARE AND HOSPITALS: Operation blocks Particulate cleanliness class: ISO 7 Particles elimination kinetics: CP10 Microbiological cleanliness class: M10 Air change rate: 15 vol/h Differential pressure: 15 Pa + or – 5 Pa Unidirectional or no unidrectional air flow Mobile Air decontamination unit Particulate cleanliness class: ISO 8 Particles elimination kinetics: CP20 Microbiological cleanliness class: M100 Air change rate: 10 vol/h Differential pressure: : 15 Pa + or – 5 Pa No unirectional air flow OTHER SECTORS: Clean room Industrial sector Very silent Retirement homes Food industry Living working spaces Options: - Over-pressure kit - Negative pressure kit Livestock farming Chemical risks 1, rue Montchavant 77250 Moret Loing et Orvanne France Phone: +33 (0)1 64 32 60 64 [email protected] Anatomopathology Endoscopes sterilisation Anaesthetic gas Other chemical risks Formaldehyde, Acetaldehyde, … Peracetic acid, Acetic acid, … Xylenes, Toluene, Benzene,… Isopropanol, Ethanol, Methanol,…

 Open the catalog to page 1
Calistair R4000-2

Destruction of contaminants via Non-Thermal Catalysis Supply grill with cooling fin Compliance zone 3 and zone 2 (ISO7 and ISO 8) Management controller Treatment of surface from 20 to 150 m2 User interface Easy bio-cleaning Incinerable filters Compression locking system Very silent Clamp with internal cam Technical specifications Pressure available: 450 Pa Volume of the zone to be treated: from 50 to 400 m3 Insulating panel Panelling double skin thickness 40 mm Stainless steel painted RAL 9010 Connection possibility Alarms Over-pressure / Under-pressure (option) Non-thermal Catalysis Destruction...