1. Catalogs
  2. SK ELEKTRONIK GMBH
  3. Thermo-FID MK and MK IP65

Thermo-FID MK and MK IP65

Thermo-FID MK and MK IP65

Thermo-FID MK and MK IP65

Product catalog summary
Product Overview: The document provides technical specifications for the Thermo-FID MK and MK IP65 flame ionization detectors, designed for various industrial applications including environmental protection and research. These devices are used for emission control and process optimization.

Specifications:
  • Models: MK and MK IP65
  • Weight: MK - 44 kg, MK IP65 - 49 kg
  • Dimensions: MK - 470x520x465 mm, MK IP65 - 500x520x465 mm
  • Display Options: ppm, mg/m³, g/m³, Vol%, %LEL
  • Linearity: < 4% above measurement range
  • Output: 0(4)mA to 20mA, 600Ω burden
  • Operating Conditions: -5°C to 40°C, < 90% relative humidity
  • Protection Class: MK - IP54, MK IP65 - IP65

Technical Design: The FID features an integrated CPU for menu-driven handling, automatic self-control, and failure analysis. It includes a process-controlled amplifier for optimal signal processing.

Accreditations: Complies with 17. BImSchV / TA Air, QAL 1, and MCERTS standards.

Measurement Capabilities:
  • Range: 0.5 mg org.C/m³ to 100000 mg org.C/m³
  • Detection Limit: < 0.01 mg org.C/m³
  • Response Time: < 0.5 s for values > 20 mg org.C/m³

Operational Requirements:
  • Sample Gas: 25 l/h at 1013 hPa
  • Detector Temperature: T2: 110°C to 200°C
  • Instrument Air: 3.0 to 3.9 Bar
  • Fuel Gas: Hydrogen, 0.7 to 1.0 Bar

Order Data and Options: Includes options for status and alarm boards, dilution probes, and external zero gas conversion kits.

Contact Information: SK Elektronik, Benzstr. 23-25, D-51381 Leverkusen, Tel. +49 2171 3955-0, Fax. +49 3955-49, www.sk-elektronik.de, [email protected]
See more

Catalog excerpts

Thermo-FID MK and MK IP65-1

Thermo-FID MK and MK IP65 Flameionisationdetector – Measuring head Technical Data Models Measured value display Completely mounted analyser in a fieldhousing for the rough applications in the field. With a measuring head attached directly without any filter or sample gas conditioning to a DIN gauge connection. Over a status signal or by the remotecontrol function it is possible to control and see the analyser function in real-time. The housing is available in protection class IP 54 or in protection class IP65. (Exhaust Emission control; Plant monitoring) General application Vol% %LEL Linearity < 4% above measurement range Measured value output 0(4)mA to 20mA; burden 600Ω not galvanically isolated Ambient temperature Geographical altitude Probe tube

 Open the catalog to page 1
Thermo-FID MK and MK IP65-2

Vacuum system Measurement range 17. BImSchV / TA Air (936/806016) QAL 1 (DIN EN 14181 and DIN EN ISO 14956) MCerts (Sira MC 050062/00) Injectors 0.5 mg org.C/m³ to 100000 mg org.C/m³ Detection limit Dilution probe Response time (T90) Samplegas (self-drawed) < 0.5 s if measured value > 20 mg org.C/m³ < 5.0 s if measured value < 20 mg org.C/m³ ca. 25 l/h with 1013 hPa alternative ca. 2l/h; ca. 5l/h; ca.60l/h Supplementary heating system (PT100) at the measuring head for the probe Catalyst-temperature 3.0 Bar to 3.9 Bar /< 2 Nm³/h Quality to ISO8573-1 minimum 1.2.1 Hydrogen 0.7 Bar to 1.0 Bar /...

 Open the catalog to page 2
Thermo-FID MK and MK IP65-3

Order data Thermo-FID Measuring head ‚MK’ With measuring head IP 54 4 x 0/4-20mA galv. seperation/ 4 potentialfree alarmcontacts Conversion kit zerogas external Conversion kit zerogas external and synthetic air as burnerair Back Purge Sample Filter Thermo-FIDMeasuring head ‚MK IP65’ Thermo-FID ‚MK IP65’ with vortexcooler Options Status- and alarmboard Dilution probe 1:10/ 1:20 External Zerogas

 Open the catalog to page 3

All SK ELEKTRONIK GMBH catalogs and technical brochures

  1. Thermo-FID TG

    4  Pages

  2. Thermo-FID ES

    4  Pages

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.