Thermo-FID ES

Thermo-FID ES

Thermo-FID ES

Product catalog summary
Product Overview: The Thermo-FID ES is a flame ionization detector designed for total hydrocarbon analysis, housed in a 19" rack. It is used across various industries for environmental protection, research, and development, including emission control and process optimization.
Technical Specifications:
  • Weight and Dimensions: Injectors weigh 9 kg and the membrane pump 10 kg. Dimensions for injectors are 483 mm (W) x 135 mm (H) x 310 mm (D), and for the membrane pump, 445 mm (W) x 154 mm (H) x 320 mm (D).
  • Measurement Display: Options include ppm, mg/m³, g/m³, Vol%, and %UEG.
  • Linearity: Less than 4% above the measurement range.
  • Output: 0(4)mA to 20mA with a burden of 600Ω, not galvanically isolated.
  • Operating Conditions: Ambient temperature ranges from -5°C to 40°C with injectors and +5°C to 40°C with the membrane pump. Air humidity should be less than 90% at 20°C and less than 50% at 40°C. Suitable for altitudes up to 1500 m above sea level.
  • Protection Class: IP20 according to DIN40050.
Applications: The Thermo-FID ES is suitable for LEL-control, emission and immission control, analytical exhaust control in the chemical industry, engine development, and analytical control of TLV and TRC values.
Technical Design: Features an integrated CPU for menu-driven handling, automatic self-control, and failure analysis. The system includes a process-controlled dynamic amplifier and digitized signal transmission to prevent distortion.
Accreditations: Includes 17. BImSchV / TA Air, QAL 1 (DIN EN 14181 and DIN EN ISO 14956), and MCerts (Sira MC 050062/00).
Performance:
  • Measurement Range: 0.5 mg org.C/m³ to 100000 mg org.C/m³ for injectors, and 2 mg org.C/m³ to 100000 mg org.C/m³ for the membrane pump.
  • Detection Limit: Less than 0.01 mg org.C/m³ for injectors and less than 0.05 mg org.C/m³ for the membrane pump.
  • Response Time: Less than 0.5 seconds for values above 20 mg org.C/m³ and less than 5.0 seconds for values below 20 mg org.C/m³.
Order Options: Includes status and alarm boards, methane-free modules, internal and external membrane pumps, dilution probes, and more.
Contact Information: SK Elektronik, Benzstr. 23-25, D-51381 Leverkusen, Tel. +49 2171 3955-0, Fax. +49 3955-49, www.sk-elektronik.de, [email protected]
See more

Catalog excerpts

Thermo-FID ES-1

Thermo-FID ES Flameionisationdetector - Rack Technical data Weight Injectors Weight Membrane pump Dimension Injectors Dimension Membrane pump Measured value display The Thermo-FID ES is a Totalhydrocarbon-analysor in a 19" rack. Burnerair and zerogas are standardised over catalyst. The sample gas conveyance takes place with injectors over the external instrument air. (Leakage control; Exhaust Emission control; Plant monitoring; Methanefree Measurements) General application < 4% above measurement range Measured value output 0(4)mA to 20mA; burden 600Ω not galvanically isolated The Thermo-FID is applied in a variety of applications for all kind of industries, environmental protection and as well for research and development. The implementations reach from an LEL-control, over emission and immission control to analytical exhaust control for the chemical industry and in the field of engine-development. Furthermore there is process optimisation and the FID is also used in the field of analytical control of TLV- and TRC- values. Ambient temperature with injectors Ambient temperature with membrane pump Technical design The electronic system of the FID allows several extra functions. The integrated CPU allows a menu-driven handling as well as a full automatic self-control and failure analysis of the system. On the alphanumeric display are shown several operation and service instructions in clear text which give an easy maintenance process. The process-controlled and dynamic amplifier analysis always within the optimum range. The amplified signals are digitised directly at the detector and get transmitted as digital data to the CPU for the analysis. Therefore there is no distortion of the little currents because of badly shielded or too long signal ways. < 50% rel. humidity, + 40°C Geographical altitude

 Open the catalog to page 1
Thermo-FID ES-2

17. BImSchV / TA Air (936/806016) QAL 1 (DIN EN 14181 and DIN EN ISO 14956) MCerts (Sira MC 050062/00) Vacuum system Membrane pump Measurement range Detection limit Response time (T90) Samplegas (self-drawed) Supplementary heating system (PT100) < 0.5 s if measured value > 20 mg org.C/m³ < 5.0 s if measured value < 20 mg org.C/m³ ca. 25 l/h with 1013 hPa alternative ca. 2 l/h; ca. 5 l/h; ca. 60 l/h 3.0 Bar to 3.9 Bar / < 2 Nm³/h Quality to ISO8573-1 minimum 1.2.1 Ambient air above Membrane pump Hydrogen 0.7 Bar to 1.0 Bar / < 80 ml/min Quality 5.0 Over internal catalyst or optional synthetic...

 Open the catalog to page 2
Thermo-FID ES-3

FID ES with external pump Back side Order data Thermo-FID 'ES' Options 4x 0/4-20mA galv. separation/ 4 potentialfree alarmcontacts inclusive Status-/Alarmboard Internal Membrane pump Set membrane pump internal sample-gas pump External Membrane pumpe Set membrane pump external sample-gas pump Dilution Probe heated without enclosure Conversion kit Zero gas external to measure two measuring streams Status- and alarmboard Module Methanefree-Methane Dilution probe 1:10/ 1:20 External Zerogas Heated 2/3 Direction Control Valve 180°C max. For E S, TG and FE MSU 4/8x HT In Line Flame Trap FS30 Eex d...

 Open the catalog to page 3

All SK ELEKTRONIK GMBH catalogs and technical brochures

  1. Thermo-FID TG

    4  Pages

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.