Overview: The document provides technical specifications for the SP-NG Series Laser Interferometers by SIOS Meßtechnik GmbH, detailing two models: SP 5000 NG and SP 15000 NG.
Measurement Specifications:- Measurement Range: SP 5000 NG ranges from 0 m to ≥8 m with a standard reflector and up to ≤80 m with a long-range reflector. SP 15000 NG ranges from 0 m to ≥20 m.
- Resolution: Both models offer a resolution of 5 pm, with some devices limited to 0.3 nm for legal reasons.
- Maximum Tilting Angle: SP 5000 NG allows ±12.5° with a center of rotation in the reflector and ±430 µrad with a plane mirror. SP 15000 NG allows ±22.5°.
- Measuring Uncertainty: Length measurement uncertainty is 0.15 µm/m under stable conditions.
Laser Specifications:- Wavelength: 632.8 nm.
- Frequency Stability: 2·10-8 after warm-up.
- Warm-up Time: 10 to 20 minutes.
Operating Conditions:- Temperature Range: 15 to 30°C.
- Maximum Displacement Speed: 3 m/s.
Geometric Data:- Dimensions: Sensor head dimensions vary between models, with the SP 5000 NG being more compact.
- Mass: Sensor head and reflector weights differ, with the SP 15000 NG being heavier.
Electrical Data:- Interfaces: Standard RS232C and USB, with other options available on request.
- Cable Length: Standard 6 m, customizable on request.
- Power Supply: 100 to 240 VAC / 47 to 63 Hz.
Safety and Compliance:- Laser Safety Class: Classified as 2M according to EN 60825-1:2014 and ANSI Z136.1 (CDRH).
Note: Specifications are subject to change by October 2026.