1. Catalogs
  2. Siglent Technologies Co., Ltd
  3. Probe Data Sheet + Siglent

Probe Data Sheet + Siglent

Probe Data Sheet + Siglent

Probe Data Sheet + Siglent

Product catalog summary
Overview: This document is a datasheet for SIGLENT series probes, detailing specifications for various models including passive, active, current, high voltage differential, and near field probes. It provides technical parameters essential for selecting the appropriate probe for specific applications.
Passive Probes: The PB470, PP510, PP215, and PP430 models feature an attenuation rate of 1X/10X, bandwidths ranging from DC-6MHz to DC-300MHz, and input impedance of 1MΩ/10MΩ. They support input voltages up to 600Vpp and operate in temperatures from -15℃ to 75℃.
Active Probes: Models ZS1500 and ZS1000 offer bandwidths of 1.5GHz and 1GHz respectively, with an input impedance of 1MΩ and capacitance of 0.9pF. They support a dynamic state of ±8V and have a non-destructive voltage range of 20V.
Current Probes: The CP5030 series supports bandwidths from DC-5MHz to DC-100MHz, with rise times as low as ≤3.5nS. They handle AC currents up to 500 Arms and peak values up to 750 A. The CP4020 series offers bandwidths up to DC-1MHz and supports AC currents up to 70 Arms.
High Voltage Differential Probes: Models like DPB5150 and DPB5700 provide bandwidths up to DC-100MHz, with maximum differential test voltages reaching 7000V. They feature high input impedance and low capacitance, ensuring accurate measurements in high voltage environments.
Near Field Probes: The SRF5030 series is designed for detecting HF magnetic fields in devices, with frequency ranges from 30MHz to 3GHz. They are suitable for localized interference detection and field distribution analysis.
Company Information: SIGLENT Technologies is a leading manufacturer of test and measurement instruments, with a strong presence in R&D, sales, and production. The company has a global footprint with headquarters in China and offices in the USA and Europe.
See more

Catalog excerpts

Probe Data Sheet + Siglent-1

SIGLENT Series SIGLENT TECHNOLOGIES CO.,LTD

 Open the catalog to page 1
Probe Data Sheet + Siglent-2

Passive Probe Parameter Attenuation Rate Input Impedance Input Capacitance Compensation Range Input Voltage Operation Temp Cable Length Attenuation Rate Input Impedance Input Capacitance Compensation Range Input Voltage Operation Temp Cable Length Active Probe Parameter Input Impedance Input Capacitance Input Dynamic State Offset Range Non-destructive Voltage Range Cable Length

 Open the catalog to page 2
Probe Data Sheet + Siglent-3

Current Probe Parameter Rise time Peak-Peak Value Range Switch Power Supply Max. rated voltage to earth Conductor Size Cable Length Rise time Peak-Peak Value Range Overload Value Current Transfer Ratio Measurement Resolution DC Accuracy Power Supply Max. rated voltage to earth Cable Length

 Open the catalog to page 3
Probe Data Sheet + Siglent-4

High Voltage Differential Probe Model Rise time Attenuation Ratio Max Differential Test Voltage (DC + Peak AC) Max input common Mode voltage (voltage-to-earth Vrms) Single-ended to Input ground Impedance Two inputs Input Capacitance CMRR Noise (Vrms) Propagation Delay Bandwidth limit Differential overvoltage Detection level Null Null Overload indicator(red light) Overload Alarm Automatic Save Offset Setting function Terminate Load Power Supply Probe body dimensions Probe body weight High Voltage Probe Model Rise time Single / Noise Attenuation Ratio Input Impedance Input Capacitance Compensation...

 Open the catalog to page 4
Probe Data Sheet + Siglent-5

Logic Probe Model Input Impedance Input Impedance Working Voltage Non-destructive Voltage Sample Rate User defined threshold range User defined threshold range Threshold Selections Threshold Selections TTL (1.4V), 5V COMS (+2.5V), ECL(-1.3V), User defined Threshold Accuracy Threshold Accuracy Delay Window Input Dynamic Range Min. Input Voltage Swing Min. Input Voltage Swing Input level Limit Threshold Groupings Cable length Line length 18CM±1CM (data line) , 12CM±1CM (ground lead) Record length Min. Detectable Pulse Width Channel Skew Input Impedance Working Voltage Non-destructive Voltage User...

 Open the catalog to page 5
Probe Data Sheet + Siglent-6

Near Field Probe Parameter Frequency Range It can be used at a distance of up to 10 cm from the units. The probe detects the spatial distribution of HF magnetic fields in devices and assemblies and allows the user to draw conclusions with regard to disturbance emissions. Frequency range:30MHz to 3GHz It is suitable for measurements u p t o 3 c m . In t e r fe re n c e sources can be localized by detecting the distribution and orientation of the field, therefore enabling a more exact use of higher resolution probes. Frequency range: 30MHz to 3GHz It is suitable for measurements u p t o 3 c m ....

 Open the catalog to page 6
Probe Data Sheet + Siglent-7

DataSheet SIGLENT Series Probe About SIGLENT SIGLENT is an international high-tech company, concentrating on R&D, sales, production and services of test &measurement Instruments. SIGLENT began to research and develop the Digital Oscilloscope independently in 2002. After a decade of development products have included digital oscilloscopes, isolated handheld oscilloscopes, function/arbitrary waveform generators, digital multimeters, DC power supplies, spectrum analyzers, and other general purpose test instrumentation. Since SIGLENTs first oscilloscope, the ADS 7000 series produced in 2005, SIGLENT...

 Open the catalog to page 7

All Siglent Technologies Co., Ltd catalogs and technical brochures

  1. SDS6000A Series

    21  Pages

  2. SDL1000X Series

    10  Pages

  3. SSM5000A Series

    15  Pages

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.