FAQ Limit Lines on the SSA3000X Spectrum Analyzer Solution: The SIGLENT SSA3000X series of spectrum analyzers feature two limit lines that can be used to visually identify whether a signal or spectrograph meets or exceeds given parameters. Limit lines can be used for: - EMI Pre-compliance measurements using FCC/CISPR limit lines - Transmitter testing: Is the signal in the proper band? - AM NRSC measurements SSA3000X Limit Line Features: - Two limit lines can be activated at any time - Each limit line can be either an upper or lower limit. Upper limits define maximum values and lower limits define minimums. - Each limit line has two modes: Line or point. Lines are single lines that exist at a user defined amplitude. Point defined limits are build using line segments and can be used to build complex limits, like NRSC masks. In this FAQ, we are going to build an upper limit line based on the conducted emissions for a class B unintentional radiator per the FCC Subpart 15 regulations using the point method. The specifications for FCC Subpart 15, Class B unintentional radiators is given as: ^Decreases with the logarithm of the frequency.
Open the catalog to page 11) Configure the instrument to cover the frequency span of interest (150 kHz – 30 MHz, in this case) by pressing the button marked Frequency on the front panel and entering the start and stop frequencies. 2) The limits are given in amplitude units of dBuV. To set the SSA units, press Amplitude and set Units to dBuV. 3) Press the Limit button to open the Limit menu. Set Limit1 to ON to activate the Limit menus. 4) Press Limit1 Edit > Set Type to Upper 5) Set Mode to Point 6) Press the grey colored soft key next to the entry labeled X Axis and enter the frequency for the first data point, 150...
Open the catalog to page 27) Press the grey colored soft key next to the entry labeled Amplitude and enter the Average start amplitude, 56 dBuV. If using a Quasi-peak detector, use the Quasipeak values of the chart. 8) Press Add Point and enter X Axis and Amplitude values for the second point, 500 kHz and 46 dBuV. Now, we have two points (1 and 2) form the first line segment of the limit line. 9) Press Add Point to enter the third point, which corresponds to the start of the second line segment. Note that the end point for one line segment has the same value as the starting point of the next line segment. 10)...
Open the catalog to page 3America SIGLENT Technologies America, Inc 6557 Cochran Rd Solon, Ohio 44139 Tel: 440-398-5800 Toll Free:877-515-5551 Fax: 440-399-1211 info@siglent.com www.siglentamerica.com Headquarters SIGLENT TECHNOLOGIES CO., LTD. Blog No.4 & No.5, Antongda Industrial Zone, 3rd Liuxian Road, Bao’an District, Shenzhen, 518101, China. Tel:+ 86 755 3661 5186 Fax:+ 86 755 3359 1582 sales@siglent.com www.siglent.com/ens Europe SIGLENT TECHNOLOGIES EUROPE GmbH Liebigstrasse 2-20, Gebaeude 14, 22113 Hamburg Germany Tel: +49(0)40-819-95946 Fax: +49(0)40-819-95947 info-eu@siglent.com www.siglenteu.com
Open the catalog to page 5All Siglent Technologies Co. catalogs and technical brochures
-
SDS6000A Series
21 Pages
-
SDS 7000A Series
21 Pages
-
SDL1000X Series
10 Pages
-
SSM5000A Series
15 Pages
-
Spectrum Analyzer Bandwidth
6 Pages
-
Siglent_Product Catalog 2016
48 Pages
-
Siglent Selection Guide 2016
10 Pages
-
Benefits of deep memory
3 Pages
-
SIGLENT Probes Datasheet
7 Pages
-
Siglent SDS2000X User Manual
197 Pages
-
Siglent SDS2000X Service Manual
59 Pages
-
Probe Data Sheet + Siglent
7 Pages
-
USB-GPIB User Manual_Siglent
13 Pages
-
ISFE User Manual
12 Pages
-
Siglent TrueArb Technology
9 Pages
-
SDS2000 VS DS2000A_EN_Siglent
10 Pages
-
SIGLENT SDS1000CFL
8 Pages
-
SIGLENT SHS800 Data Sheet
10 Pages