TOC-L

TOC-L
1 / 16 PagesView full catalog

TOC-L

Product catalog summary
Overview: The document details the Shimadzu TOC-L series, designed for Total Organic Carbon (TOC) and Total Nitrogen (TN) analysis using catalytic oxidation and non-dispersive infrared detection. It includes standalone and PC-controlled models.
Specifications: The TOC-L series features a multifunction sample pretreatment and injection system, a TC combustion tube, and a non-dispersive infrared detector (NDIR). It supports USB and optional LAN connectivity, compatible with Windows XP and 7, requiring specific hardware specifications.
Procedures: The system allows sample addition during continuous measurement with an autosampler, supports schedule files for routine measurements, and includes accuracy control to ensure reliable results. It can exclude anomalous values and recalculate results automatically.
Features: User-friendly software with a color LCD screen and keyboard, automatic sleep/restart functions, optimal measurement settings, and automatic calibration curve selection. The system is space-saving and energy-efficient, with reduced width and energy consumption.
Options and Accessories: Configurable with options like the ASI-L autosampler, OCT-L 8-port sampler, TNM-L unit for simultaneous TOC and TN measurements, and SSM-5000A solid sample combustion unit.
Recommendations: Use general-purpose USB printers for data output. Compatible with Shimadzu CLASS-Agent software for database consolidation and compliant with 21CFR Part11 for user authentication and operational history logging.
TN Measurement Specifications: Chemiluminescence-based TN measurement system with a range of 0 to 10,000 mg/L, detection limits of 5 µg/L for CPH/CSH and 20 µg/L for CPN/CSN, and a reproducibility CV of 3% max. Measuring time is approximately 4 minutes.
Accessories: Lists accessories for various vial sizes and part numbers, options for autosamplers, TN units, solid sample combustion units, and kits for specific measurement needs.
Compatibility and Options: Details compatibility of options and accessories with different models, and specific kits for measurement enhancements like high-salt sample combustion and volatile organic carbon measurement.
Related Products: Mentions related products like the TOC-VW Wet Oxidation/NDIR Method and the ON-LINE TOC-VCSH Combustion Catalytic Oxidation/NDIR Method Online Model.
Additional Information: Includes external dimension diagrams, product availability notes, and disclaimers regarding trademarks and information accuracy.
See more

Catalog excerpts

TOC-L-1

Total Organic Carbon Analyzer

 Open the catalog to page 1
TOC-L-2

Global Standard for TOC Analyzers Combustion Catalytic Oxidation/NDIR Detection Method TOC Analyzers with a User-Friendly Design 4 TOC-V series, the world's top seller, has evolved. Easy-to-operate keyboard and easy-to-read TFT color LCD screen (standalone model) • Output of measurement data to USB memory sticks or conventional PC printer (standalone model) • A wealth of options, including sea water sample measurement and compatibility with small samples volumes • Space-saving and energy-saving design A Full Range of Models and Options Provide Total Solutions to Suit Your Application 4 • Select...

 Open the catalog to page 2
TOC-L-4

Process Control Effluent treatment process control Processes (Plating, etching, washing, water-based cutting) Ultrapure water recycling and re-purification processes Example of TOC Measurement of Nickel Plating Solution Analysis instrument: TOC-LCPH Measurement method: TOC measurement of Nickel Plating Solution, thousand fold dilution with pure water (TOC measurement (NPOC measurement) with sample acidification and sparging) Measurement results: TOC = 12.80 mg/L (C.V. = 0.22 %) (before dilution TOC = 1.280 %) Quality Control Drinking water Aluminum foil Electronic components Water supply equipment...

 Open the catalog to page 4
TOC-L-5

Water Quality Control Tap water (Drinking water, raw Ultrapure water (Ultrapure water used in semiconductor manufacturing, Liquid Crystal manufacturing, pharmaceutical manufacturing, and nuclear power generation, as well as used ultrapure water) Effluent (Industrial effluent, water treatment effluent, etc.) Pool water, spa water, bath water, boiler water, water from industrial processes Example of TOC Measurement of Tap Water Analysis instrument: TOC-LCPH Measurement method: TOC measurement (NPOC measurement) with sample acidification and sparging Measurement results: TOC = 932 µg/L (C.V. = 0.72...

 Open the catalog to page 5
TOC-L-6

Shimadzu 680 °C Combustion Catalytic Oxidation/ NDIR Detection Method Measurement System Combining Experience and Reliability Features The most important feature of a TOC analyzer is its ability to efficiently oxidize not only easily decomposed, low molecular weight organic compounds, but also hard-to-decompose insoluble and macromolecular organic compounds. The 680 °C combustion catalytic oxidation method, developed by Shimadzu and now used worldwide, can efficiently analyze all organic compounds. ■ Extremely wide measurement range from 4 µg/L to 30,000 mg/L, applicable to everything from ultrapure...

 Open the catalog to page 6
TOC-L-7

Flow Line Diagrams TOC-LCPH High-Sensitivity Model Printer Acid Dilution water ASI-L autosampler Electronic gas flowrate controller TC sample injection port, slide type Carrier gas Multifunction sample pretreatment/ injection system Ultrapure water trap for blank check Non-dispersive infrared detector (NDIR) IC sample injection port, slide type Personal computer (Not included in the TOC analyzer components.) Dehumidifier/gas pretreatment unit IC reaction reagent IC reaction vessel TC furnace TOC-LCSN Standard Model Acid Multifunction sample pretreatment/ injection system Keyboard Electronic gas...

 Open the catalog to page 7
TOC-L-8

Software Features Intuitive Operability and a Wealth of Functions T O C - L CPH/CPN P C - C o n t r o l l e d M o d e l Enabling Simple, Intuitive Operation Icons and function names shown on large buttons User-friendly display of the name, ID, and measurement results for selected samples, all in specific columns ・List of files used by type ・Can be sorted by file name and date created ・File details are displayed with tooltip Operations are accessed by right-clicking on the table Insert samples by dragging and dropping measurement conditions files Drag the mouse over cells to batch-enter identical...

 Open the catalog to page 8
TOC-L-9

■ Addition of samples during continuous measurement Samples can be inserted during continuous measurement by an autosampler. ■ Creation of schedule files The measurement of multiple samples, configured by measurement condition and vial number, can be registered as a schedule file. This provides support for routine measurements. ■ Input/output of text files Measurement results can be output as text files, which can be loaded by Excel and other applications. In addition, text files can be loaded as measurement schedules. ■ Accuracy control function For accuracy control, samples can be inserted...

 Open the catalog to page 9
TOC-L-10

Standalone Model with a TFT Color LCD Screen Providing Outstanding Visibility T O C - L CSH/CSN Standalone Model ■ Color screen and keyboard Easy-to-read TFT (Thin Film Transistor) color LCD screen and keyboard designed with simplicity and user-friendliness in mind. Color Screen ■ Use a general-purpose USB printer To use a PC printer or portable thermal printer, simply connect it to the USB terminal on the back of the instrument. (Contact your Shimadzu representative for a list of suitable printers.) ■ Data output to USB memory stick Measurement results can be output in CSV format to a USB memory...

 Open the catalog to page 10
TOC-L-11

Space-Saving Design |- The width of the instrument is 20 % less in comparison with conventional Shimadzu models. This enables more effective use of laboratory space. The instrument width is unchanged even when the TN unit is added. This product conforms to Shimadzu's Eco-labeled designation. Energy consumption has been reduced by 36 % (100 V) and 43 % (200 V) in comparison with conventional Shimadzu models. (Assuming 8 hours operation/day x 5 days/week) eco Common to PC-Controlled Model and Standalone Model ■ Automatic setting of optimal measurement conditions When the standard solution concentration...

 Open the catalog to page 11
TOC-L-12

Options for Configuring an Automatic Measurement System ASI-L Autosampler Even More Functionality and Convenience, Enabling Samples to Be Added During Continuous Measurement Features ■ Select from three vial types with different capacities to suit your application. Combination of vial capacity and number of vials ・ 9 mL vials × 93 ・ 24 mL vials × 93 ・ 40 mL vials × 68 Two types of ASI-L units are available, one for 24 mL vials and the other for 9 mL and 40 mL vials. ■ Optional magnetic stirrers agitate the sample in the vials to prevent the settling of suspended solids. Magnetic stirrers are...

 Open the catalog to page 12

All Shimadzu France catalogs and technical brochures

  1. LCMS-2050

    16  Pages

  2. MCT Series

    12  Pages

  3. LC/MS/MS

    2  Pages

  4. SALD-7500nano

    16  Pages

  5. LC/MS/MS MRM

    2  Pages

  6. LC/MS/MS

    2  Pages

  7. TRViewX

    8  Pages

  8. UW/UX Series

    4  Pages

  9. LCMS-8060

    20  Pages

  10. QC Reporter

    6  Pages

  11. OneSight

    4  Pages

  12. TD-30 Series

    12  Pages

  13. Nexis GC-2030

    20  Pages

  14. MOC63u

    6  Pages

  15. UHPLC Nexera X2

    20  Pages

  16. RF-6000

    24  Pages

  17. iMScope Trio

    28  Pages

  18. UV-3600 Plus

    26  Pages

  19. EDX-7000/8000

    24  Pages

  20. LCMS-IT-TOF

    16  Pages

  21. ELB SERIES

    2  Pages

  22. TW TWC SERIES

    8  Pages

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.