SDHmini-Logging Hand Held Hygrometer Instruction Manual

SDHmini-Logging Hand Held Hygrometer Instruction Manual
1 / 36 PagesView full catalog

SDHmini-Logging Hand Held Hygrometer Instruction Manual

Product catalog summary
Introduction
The SDHmini-L is a micro-controller based, battery-operated dewpoint hygrometer designed for measuring moisture content in gases. It features USB and Bluetooth interfaces, a moisture sensor, and a graphical user interface with multiple language options. The package includes the instrument, carry case, and various accessories.
Important Information
The device complies with Bluetooth certifications and is powered by a non-user-replaceable lithium-ion cell. It should be charged at room temperature. USB communication may vary, and a USB hub is recommended if issues arise. The measuring head can withstand a maximum pressure of 0.3 bar (4 psi).
Installation
Avoid exposure to gases like ammonia and chlorine to prevent sensor damage. Minimize the sample system's volume and maintain flow rates between 2 and 5 liters/min. A piping installation schematic is provided for setup.
Normal Operations
Includes powering on/off, graphical viewing, display options, and unit selection.
Calibration
Features Auto Calibration (AutoCal) and Calibration Setpoint (CalSet) with error message guidance.
Sensor Related Settings
Settings include sensor type, gas types, molecular weights, and pressure adjustments.
Non-Measurement Related Configuration Functions
Includes changing date, time, international settings, and hotkeys.
Bluetooth Printer Configuration
Details on configuring a Bluetooth printer for use with the device.
Data Logging
Features Quicklog, Snapshot, Advanced Data Logging, and data review/storage options.
Additional Information
Includes temperature, response time, guarantee, basic definitions, and a problem guide.
Specifications and Diagrams
Technical specifications and a general assembly diagram are provided.
Specifications and Setup
The SDHmini-L measures moisture content in gases with a flow rate between 1 and 10 litres/min. It includes a desiccant head to keep the sensor dry when not in use. Sample connections include Fir Tree and Swagelok fittings, with recommended materials being PTFE and stainless steel.
Operation Procedures
To take a moisture reading, purge the system of residual moisture, connect the SDHmini-L, and allow the gas to purge for two minutes. The desiccant head should be extended during sampling and retracted when not in use.
Desiccant Dry-Down Technology
This technology keeps the sensor dry between tests, allowing rapid spot checks by isolating the sensor from ambient air.
Pressure and Flow Control
Reduce gas pressure to atmospheric levels using a needle valve or pressure regulator. Stainless steel regulators are recommended to avoid moisture ingress.
Contamination Prevention
Use inline particle filters and coalescing or membrane filters to prevent contamination, which can delay sensor response and affect accuracy.
Calibration
Field calibrations should be performed every two to three months using AutoCal or CalSet functions.
Sensor Settings
Configure sensor type, gas types, molecular weights, and pressure settings for accurate readings.
Non-Measurement Configuration
Includes changing the date and other settings to ensure correct operation.
Time and International Settings Configuration
Adjust time and international settings, including language and date format, through the Configuration menu.
Hotkeys and Bluetooth Printer Configuration
Access hotkeys for quick functions and configure Bluetooth printers through the Configuration menu.
Data Logging
Initiate data logging with Quicklog and Snapshot functions. Advanced settings allow customization of sample rates and durations.
Data Review and Storage
Manage logged data through the Data File Management function, which supports various file types and storage capacities.
Additional Information
Includes temperature and response time considerations, and a problem guide for troubleshooting common issues.
Specifications
Specifications include accuracy, dimensions, weight, display features, power supply, operating conditions, and sensor ranges.
See more

Catalog excerpts

SDHmini-Logging Hand Held Hygrometer Instruction Manual-1

Model SDHmini-L Dewpoint Meter Instruction Manual 3005 SDHmini-L User Manual - Iss 1.3

 Open the catalog to page 1
SDHmini-Logging Hand Held Hygrometer Instruction Manual-5

1 Introduction 1.1 Unpacking Your SHAW Moisture Meters SDHmini-L Please examine the SDHmini-L package for any damage or mishandling. If any damage is evident please notify the carrier and the SHAW Moisture Meters representative from where this unit was purchased. You should have received (if ordered): ● 1 model SDHmini-L instrument ● 1 carry case ● 1 Allen key ● 1 two metre length of PTFE tubing ● 1 USB charging cable ● 1 Fir Tree fitting plus 1 x fittings as ordered ● 1 AC/DC adaptor ● 1 pressure dewpoint circular calculator ● 1 USB flash drive with SHAW SDHmini User Tool software ● 1 instruction...

 Open the catalog to page 5
SDHmini-Logging Hand Held Hygrometer Instruction Manual-6

The displayed units can be changed on both the primary and secondary displays for the selectable moisture unit e.g. °C, °F, g/m3, ppm(v), ppm(w) and lbs/MMscf. The SDHmini-L settings can be adjusted by a user-friendly software interface incorporating multiple menus, hot key combinations, information displays and control functions. Use the supplied USB cable, attach between the PC and the instrument, an automatic connection over USB is established. The USB port allows: ● Charging of the battery (Li-ion) when connected to PC or dedicated power supply ● Downloading of new firmware to the unit ●...

 Open the catalog to page 6
SDHmini-Logging Hand Held Hygrometer Instruction Manual-7

2 Important Information 2.1 Bluetooth Certifications This equipment complies with FCC radiation exposure limits set forth for an uncontrolled environment - Contains FCC ID:QOQWT41/ QOQWT41u. This device complies with Industry Canada licence-exempt RSS standard(s) Contains IC:5123A-BGTWT41/ 5123A-BGTWT41u. WT41 meets the requirements of the standards below and hence fulfils the requirements of EMC Directive 89/336/EEC as amended by Directives 92/31/EEC and 93/68/EEC within CE marking requirement. WT41 has modular certification with certification number R 209-J00047/J00231. 2.2 Symbols Table 2.3...

 Open the catalog to page 7
SDHmini-Logging Hand Held Hygrometer Instruction Manual-8

2.4 Pressure Exposure The maximum pressure to which the telescopic measuring head is exposed should not be more than 0.3 bar, 4 psi Exposing the measuring head to higher pressures may damage the instrument and result in injury to the operator or other personnel in the area. 2.5 USB Communication Performance Note: The use of the USB communication channel and charging power port within the instrument may result in not all connections to all USB ports on all PCs or laptops being entirely successful. Many USBs ports even within the same PC or laptop differ from each other in electrical hardware or...

 Open the catalog to page 8
SDHmini-Logging Hand Held Hygrometer Instruction Manual-9

3 Installation 3.1 Gases to Avoid The SHAW moisture sensors are suitable for many different industrial and research applications. Most gases can be checked for their moisture content with no need for the calibration to be altered when changing between different gases, as the sensor operates only with reference to the water vapour content. There are however, some gases that must be avoided as they are not compatible with the material of construction of the sensor. Ammonia (NH3), Ozone (O3) and Chlorine (Cl2) must be avoided at all times, even in small quantities. Hydrogen Chloride (HCl) also attacks...

 Open the catalog to page 9
SDHmini-Logging Hand Held Hygrometer Instruction Manual-10

the SDHmini-L response time to an unacceptable level. In general, a flow rate of between 2 and 5 litres/min at normal temperature and pressure will give the right balance. The SDHmini-L is designed to operate at atmospheric pressure with a maximum of 0.3 bar. 3.3 Piping Installation Schematic MAIN PROCESS LINE Notes a. The sample point should be on the upper surface of the horizontal pipe or from a vertical section of pipe, wherever possible. b. The sample tube should run upwards from the sample point. If this is not possible, then an inspection port or drain tap should be installed at the lowest...

 Open the catalog to page 10
SDHmini-Logging Hand Held Hygrometer Instruction Manual-11

3.4 Piping Schematic Component Index 1. Sample Isolation Valve - This is a recommended item as it allows access to the sample system without interrupting the main process line. 2. Filter Unit – A filter unit is recommended when the samples are likely to contain particulate matter. If the air/gas sample contains heavy hydrocarbon condensate, the filter must be of the coalescing type with a drain. The filter unit should be positioned as close to the sample point as practical. 3. Sample Tube – This should be stainless steel for dry air or gas applications but copper or carbon steel can be used where...

 Open the catalog to page 11
SDHmini-Logging Hand Held Hygrometer Instruction Manual-12

Continue this for two minutes to remove any residual moisture from the sample pipe work and components. Check that no liquid or particulate contamination, which could damage the sensor, is passing through the sample pipe. Directing the sample onto a piece of white paper (such as a filter paper) will make it easier to see any dust or oil mist in the sample. Connect the SDHmini-L. The SDHmini-L is not flow sensitive, however, the sample flow needs to be enough to fill the sensor head with gas effectively, so anything less than 1 litre/min would mean that the instrument would take too long to settle...

 Open the catalog to page 12
SDHmini-Logging Hand Held Hygrometer Instruction Manual-13

Note: It is important that the head assembly is either fully up or fully down to keep the desiccant isolated from the ambient air so avoiding wetting it. When the instrument is not in use it should be stored with the head in the down position. 3.7 Desiccant Dry-Down Technology The unique design allows purging of the gas in the “DryDown Position” which will bring all surfaces and voids to equilibrium. Sensor surrounded by desiccant ensuring it is dry, ready for the next measurement. The desiccant and sensor are always isolated from atmospheric air to ensure an accurate reading. Sensor exposed...

 Open the catalog to page 13
SDHmini-Logging Hand Held Hygrometer Instruction Manual-14

3.8 Sample Connections The SDHmini-L is supplied with two connections: a Fir Tree on one side and a Swagelok compression fitting for ¼”, ⅛”, 6 mm OD pipe or another Fir Tree fitting (as specified on your order) on the other. Either can be used as the inlet connection (6). PTFE and stainless steel are recommended materials for sample pipes. Some specialist flexible tubing may be suitable as sample pipe but common flexible tubing such as PVC, nylon, polythene or rubber, should not be used, as they are permeable and water vapour from the atmosphere will diffuse through them into the sample gas....