Overview: Setra's Model 204 pressure transducers are designed for precise pressure measurements in gas or liquid media compatible with 17-4 PH stainless steel. They are suitable for industrial, laboratory, and aerospace applications due to their high-level output signal, excellent stability, and fast dynamic response.
Specifications: The Model 204 offers a 0-5 VDC output, while the Model C204 provides a 4-20 mA output. Both models feature a one-piece stainless steel sensor and an insulated electrode forming a variable capacitor. The transducers are compensated for zero and sensitivity shifts due to environmental temperature variations.
Performance: The transducers boast an accuracy of ±0.11% FS, with non-linearity at ±0.07% FS and hysteresis at 0.08% FS. They are designed to operate under extreme temperature conditions with minimal hysteresis and stable performance.
Applications: These transducers are ideal for high accuracy general-purpose applications, R&D test and measurement, vacuum systems, dynamometers, and engine test cells.
Features: Key features include fast warm-up, low output noise, fast response time (less than 1 millisecond), and FM-approved intrinsically safe design. They meet CE conformance standards and are ISO 9001 certified.
Environmental Data: Operating temperature ranges from 0 to +175°F (-18 to +80°C), with storage temperatures from -65 to +250°F (-55 to +120°C). They can withstand vibrations up to 2g and shocks up to 50g.
Electrical Data: The Model 204 operates on a 4-wire circuit with an excitation of 22 to 30 VDC, while the Model C204 uses a 2-wire circuit with a 4 to 20 mA output. Both models have specific requirements for external load and power supply variations.
Options: Various electrical, performance, and mechanical options are available, including special outputs, intrinsic safe design, remote calibration, and explosion-proof housing.
Ordering Information: Customers can order as Model 204 for voltage output or Model C204 for current output, specifying pressure range and options.