SERVOTOUGH Oxy 1900 Functional Safety Manual 01910006B_3

SERVOTOUGH Oxy 1900 Functional Safety Manual 01910006B_3
1 / 27 PagesView full catalog

SERVOTOUGH Oxy 1900 Functional Safety Manual 01910006B_3

Product catalog summary
Introduction
The SERVOTOUGH Oxy 1900 is an oxygen concentration measurement device using a paramagnetic transducer, suitable for hazardous environments. It offers digital and mA signal outputs, with optional features like flow alarm and pressure compensation.

Safety Information
Certified for safety instrumented systems, the device operates in Low Demand mode, supporting gas concentration and sample flow loss indications. The safety system includes the transducer and signal processing, excluding remote validation/calibration.

Safety Instructions
Installation must adhere to the Operator Manual. A continuous flow of clean sample gas is essential, with external devices needed if the internal flow alarm is absent.

Proof Tests
Routine and additional checks for configurations with auto-validation, internal flow sensor, or pressure compensation ensure safety and functionality.

Failure Rate Data
Failure rates for different configurations support safety functions, with PFDAVG calculated based on a 24-hour MTTR and a 1-year proof test interval.

Product Configurations
The manual details various configurations, from basic to complete packs, including auto-validation hardware.

Additional Information
Definitions of terms like PFDAVG, dangerous failure, and safety integrity level clarify safety claims and data.
Specifications and Procedures
Maintaining sample gas composition and correct gas presentation to analyzers is crucial. Monitoring functions in auto-validation systems are emphasized, and mA outputs must treat out-of-range currents as fault indications.

Settings and Configuration
Post-installation, specific settings must support safety functions, including mA output and flow alarm settings. Passwords should be changed from defaults to prevent unauthorized changes.

Safety Checks and Maintenance
Regular safety checks are required per IEC 61508, with a recommended proof test interval of one year. Fault conditions must be resolved before tests. Routine maintenance includes calibration of oxygen measurement and flow sensors.

Proof Tests
These tests detect undiagnosed dangerous faults, involving mA output verification, service-in-progress relay, and low/high calibrations. Additional checks are needed for auto-validation, internal flow sensor, or pressure compensation options.

Definitions and Descriptions
Definitions for terms such as PFDAVG, Dangerous Failure, FMEDA, and SIL are provided. The document explains hardware fault tolerance, safe failure fraction, and SIL, distinguishing between Type A and Type B subsystems based on complexity and failure data.
See more

Catalog excerpts

SERVOTOUGH Oxy 1900 Functional Safety Manual 01910006B_3-1

FUNCTIONAL SAFETY MANUAL OPERATOR MANUAL Part Number: Revision: Language:

 Open the catalog to page 1
SERVOTOUGH Oxy 1900 Functional Safety Manual 01910006B_3-3

IMPORTANT INFORMATION Continued safe and reliable operation of this equipment is conditional on all installation, operation and maintenance procedures being carried out in accordance with the appropriate manuals, by personnel having appropriate qualifications, experience and training. Failure to observe the requirements of the manual may result in the user being held responsible for the consequences and may invalidate any warranty. Servomex will accept no liability for unauthorized modifications to Servomex supplied equipment. This manual should be read in conjunction with the other product manuals...

 Open the catalog to page 3
SERVOTOUGH Oxy 1900 Functional Safety Manual 01910006B_3-4

© Servomex Group Limited 2019

 Open the catalog to page 4
SERVOTOUGH Oxy 1900 Functional Safety Manual 01910006B_3-5

© Servomex Group Limited 2019

 Open the catalog to page 5
SERVOTOUGH Oxy 1900 Functional Safety Manual 01910006B_3-6

© Servomex Group Limited 2019

 Open the catalog to page 6
SERVOTOUGH Oxy 1900 Functional Safety Manual 01910006B_3-7

About this manual Safety information Sections of this manual contain supplementary installation and configuration instructions. These should always be read in conjunction with the other manuals listed below which provide more detailed instructions and contain important safety information. Scope of the manual This manual provides information and instructions that will be needed in order to use the SERVOTOUGH Oxy in safety instrumented systems. It is aimed at those responsible for planning, designing, installing, commissioning and maintaining safety instrumented systems using the SERVOTOUGH Oxy...

 Open the catalog to page 7
SERVOTOUGH Oxy 1900 Functional Safety Manual 01910006B_3-8

Introduction Product description The SERVOTOUGH Oxy is designed to provide a stable and accurate measurement of oxygen concentration using a paramagnetic transducer. It is suitable for use in hazardous areas. The sample measurement is shown on the analyzer display and is also available via digital communications and as a mA output. The analyzer also provides fault, alarm, maintenance required, service in progress and mA range change signal outputs. The analyzer can also be equipped with an optional flow alarm, pressure compensation, sample heating and auto-validation/auto-calibration functionality....

 Open the catalog to page 8
SERVOTOUGH Oxy 1900 Functional Safety Manual 01910006B_3-9

Safety information Demand Mode The safety integrity level claimed for this product is based on usage in a Low Demand mode of operation in a 1oo1 (simplex) architecture, as defined in IEC 61508 Edition 2.0 2010-01. Safety functions The SERVOTOUGH Oxy is certified suitable for use in safety instrumented systems in respect of hardware safety integrity ONLY, according to IEC 61508 Edition 2.0 201004. The usable safety functions are defined as: Safety Function 1: The indication of gas concentration by means of a 4-20mA output to the level of performance stated in the products literature. A diagnosed...

 Open the catalog to page 9
SERVOTOUGH Oxy 1900 Functional Safety Manual 01910006B_3-10

Boundary of safety system The following aspects of the are considered to be within the boundary of the safety system covered in this manual: • • • • • • • • Paramagnetic oxygen transducer Transducer heating All signal processing and support functions Power supply, distribution and intrinsic safety protection mA output Internal pressure compensation (optional) Internal flow measurement (optional) Internal sample pipe work heating (optional) Remote validation/calibration initiation and control option is explicitly excluded from the SIL 2 assessment. Data is provided to support an independent implementation...

 Open the catalog to page 10
SERVOTOUGH Oxy 1900 Functional Safety Manual 01910006B_3-11

Product configurations This manual explicitly covers the following analyzer configurations based on standard “packs” and options. Analyzer Configuration Options Entry Pack Paramagnetic oxygen transducer Transducer heating All signal processing, power and support functions mA output Relays Entry pack plus internal sample pipe work heating Pressure and Flow Pack Hot pack plus internal pressure compensation and flow sensor Complete Pack All options fitted (Excluding Auto-validation). In addition, by considering the combination of options that produces the lowest Safe Failure Fraction the manual...

 Open the catalog to page 11
SERVOTOUGH Oxy 1900 Functional Safety Manual 01910006B_3-12

Failure rate data The following section(s) contain the detailed supporting information for the various product configurations. All failures rates were predicted from an FMEDA (Failure Modes, Effects, and Diagnostics Analysis) and are stated in “failures per million hours” (fpmh). These figures may be multiplied by 1000 to obtain the failure rates in units of FIT. The PFDAVG is based on a Mean Time To Repair (MTTR) of 24 hours and a Proof Test Interval (T1) of 8760 hours (1 year). Failure data. a) Safety function 1 - This table contains the functional safety data that applies when the safety function...

 Open the catalog to page 12
SERVOTOUGH Oxy 1900 Functional Safety Manual 01910006B_3-13

b) Safety function 2 - This table contains the functional safety data that applies to the flow fault monitor via the 4-20mA output. Safety Related Characteristic Device type Hardware fault tolerance Safety integrity level Dangerous detected failure rate Dangerous undetected failure rate Safe failure rate Diagnostic coverage % Safe failure fraction Total failure rate Average probability of failure PFDAVG on demand c) Auto-validation – This table contains functional safety data for the additional hardware required for remote validation/calibration function, as it affects safety function 1. Safety...

 Open the catalog to page 13
SERVOTOUGH Oxy 1900 Functional Safety Manual 01910006B_3-14

Safety instructions General observations The must be installed and operated as detailed in the Operator Manual and product literature. Sample Gas A continuous controlled flow of clean sample gas is required. Suitable sample conditioning must be used to prevent contamination of the transducers and associated pipe work. Where the internal flow alarm is not fitted it is essential that the safety instrumented system incorporates an external device to detect loss of flow. Where the internal flow sensor is fitted, it is assumed that the composition of the sample gas is broadly the same as when it was...

 Open the catalog to page 14
SERVOTOUGH Oxy 1900 Functional Safety Manual 01910006B_3-15

Refer to the Operator manual for detail on how the auto-validation function works. Settings After installation and commissioning according to the instructions in the product manuals, the following settings MUST be set to support the safety function. Please refer to the Operator Manual for step by step instructions on these settings. Note also that a password will typically be required in this process. mA output settings Any mA output used in a safety function MUST be set to indicate an out-of-range current under fault conditions. This is achieved as follows. Set the mA output range to the 4-20mA...