The document provides a detailed overview of the SERVOPRO NanoTrace DF-745, a high-purity gas moisture analyzer using Tunable Diode Laser Absorption Spectroscopy (TDLAS) technology. It is designed for ultra-trace moisture measurements in ultra-high purity electronic gases, crucial for industries like semiconductor manufacturing.

Key Applications
- ASU bulk gas production quality control
- Leak detection for UHP electronic gases in semiconductor fabs

Unrivalled Performance
- Utilizes high-stability TDLAS technology
- Features a new solid-state hard drive and CPU
- Offers a lower detection limit of 1ppb
- Manufactured by Servomex with 70 years of experience

Flexibility
- Broad detection range from 0ppb to 20ppm
- Operable via front panel or digital communication
- Resistant to gas cell contamination

Ease of Use
- Improved uptime with TDLAS first principle physics
- Laser lock system ensures accurate moisture spectra peak location
- High reliability with repeatable baseline measurements

Low Cost of Ownership
- Herriott Cell sensor design used in NASA’s Mars rovers
- No zero drift reduces calibration needs
- Modular design allows field component replacement
- No consumables required

Compliance
- Complies with IEC 61010-1, EU EMC Directive, and EU Low Voltage Directive
- Classified as a Class 1 laser product

Technical Specifications
- Measures H2O in N2, O2, H2, He, Ar, and CO2
- Measurement range: 0-20ppm
- Lower detection limit: 1ppb
- Response time: <3 minutes at 1l/min
- Operating temperature: +10°C to +40°C
- Sample flow range: 0.5 to 2 l/min
- Pressure range: 30 to 150psi

Utilities
- Supply voltage: 110V ac @ 5A or 230V ac 50/60 Hz @ 2.5A
- Optional zero gas recommended near LDL
- Complies with EU directives and electrical safety standards

Physical Characteristics
- Size: 483mm x 266mm x 631mm
- Weight: 33.2kg
- Mounting: 19” rack mount NEMA 1 enclosure

Additional Information
- The document emphasizes that the analyzer is not intended for medical use.
- Servomex reserves the right to change specifications without notice.
See more

Catalog excerpts

SERVOPRO NanoTrace DF-745 Product Brochure-1

Product overview High Purity Tunable diode laser absorption spectroscopy (TDLAS) trace/ultra-trace moisture measurements for ultra high purity electronic gases Unrivalled performance SENSING TECHNOLOGY Laser Moisture Uses industry-leading, high stability Tunable Diode Laser (TDLAS) sensing technology New Solid State Hard Drive and CPU 1ppb Lower Detection Limit (LDL) Manufactured by Servomex 70 years’ experience innovating and pioneering gas analysis and thousands of units used in the field every year Flexible Broad detection range: 0ppb – 20ppm NanoTrace SERIES Operable via front panel or digital communication options Analysis resistant to gas cell contamination: DF-745 operates to specification with up to 90% signal loss Easy to use Improved uptime with TDLAS first principle physics methodology Laser lock system guarantees location of the moisture spectra peak High reliability - repeatable baseline measurements are not affected by a loss in mirror reflectivity Herriott Cell sensor design, the same as used in NASA’s Mars rovers Absence of zero drift reduces calibration requirements Modular design allows individual component replacement in the field No consumables required Key applications ASU bulk gas production quality control prior to final purification to UHP specifications Pollution Degree 2 EU EMC Directive EU Low Voltage Directive Class 1 laser product Leak detection checks for UHP electronic gases used in semiconductor fabs For more information visit servomex.com/cont

 Open the catalog to page 1
SERVOPRO NanoTrace DF-745 Product Brochure-2

NanoTrace DF-745 Flexible TDLAS trace and ultra-trace measurements Modern LCD and LED manufacturing processes require an ultra-trace quality measurement for moisture contaminants in high purity electronics grade gases. In such a demanding application, users need analysis capable of delivering high-accuracy and ultra-low detection limits in multiple background gases. No matter how demanding the application requirement, you’ll want a device that reduces preventative maintenance costs, maximizes uptime and has a long-life in the market place. We don’t believe you should have to compromise. A no...

 Open the catalog to page 2
SERVOPRO NanoTrace DF-745 Product Brochure-3

Technical data sheet Tunable Diode Laser Absorption Spectroscopy (TDLAS) Performance Measurement range Lower detection limit Intrinsic error (accuracy) FS ±3% of reading / ±0.2ppb (whichever is greater) Zero drift/month Span drift/month Upset recovery time <5 minutes to return to within 10ppb of previous stable reading Signal outputs/inputs Analog output Analog output range Scalable to any range between 0-2ppb to 0-20ppm Visual alarms 4 moisture levels, temperature, system error, pressure range and hydrogen safety system (if applicable) Dual scale range 2 user selectable analog output ranges...

 Open the catalog to page 3
SERVOPRO NanoTrace DF-745 Product Brochure-4

Physical Size 483mm (19”) Wide x 266mm (10.5”) High x 631mm (24.9”) Deep (see drawing below) Standard aspirator connection 1/4” compression inlet and outlet fittings 19” rack mount NEMA 1 enclosure, IP10 Utilities Supply voltage Optional - recommended if operating near LDL Standard aspirator gas supply (gauge) N2 or CDA at 80psig (±3psig) 15l/min with a backpressure on outlet stream of <2psig N2 or CDA 60 to 100psi, 4.14 to 6.89 Bar, 413.7 to 689.5 KPa (Isolation panel option) This product complies with the EU EMC Directive, EU Low Voltage Directive, Pollution Degree 2. This is a class 1 laser...

 Open the catalog to page 4
SERVOPRO NanoTrace DF-745 Product Brochure-5

Dimensional drawings Dimensions shown in millimetres [inches]

 Open the catalog to page 5
SERVOPRO NanoTrace DF-745 Product Brochure-6

We’re ready to help Whatever your gas analysis requirements, wherever you are. These analyzers are not intended for any form of use on humans and are not medical devices as described in the Medical Devices legislation or regulation. Please note: Whilst every effort has been made to ensure accuracy, no responsibility can be accepted for errors and omissions. Data may change, as well as legislation, and you are strongly advised to obtain copies of the most recently issued regulations, standards and guidelines. This document is not intended to form the basis of a contract. Servomex has a policy...

 Open the catalog to page 6

All SERVOMEX catalogs and technical brochures

  1. ES 29 Clean Air

    11  Pages

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.