SERVOMEX OxyDetect Certification Manual - Hazardous Area - 27 Pages

Add to MyDirectIndustry favorites
SERVOMEX OxyDetect Certification Manual - Hazardous Area

Catalog excerpts

SERVOMEX OxyDetect Part Number: Revision: Language: SERVOMEX \ A MEASUREABLE ADVANTAGE

 Open the catalog to page 1

1 INTRODUCTION 2    EU DECLARATION OF CONFORMITY 3    ATEX CERTIFICATION (SGS BASEEFA) 4 IECEx INTERNATIONAL CERTIFICATION (SGS BASEEFA) 5 NORTH AMERICAN HAZLOC CERTIFICATION (SGS) 6 NORTH AMERICAN GP CERTIFICATION (SGS)

 Open the catalog to page 3

SECTION 1 INTRODUCTION The 05311B1 OxyDetect has been approved by various authorities for installation in hazardous and non-hazardous areas. The certification offered depends upon the analyser specification and country of delivery. See the certification labels attached to the analyser for exact approval details. The following information summarises the approvals that have been obtained. This certification manual covers: 1. 05311B1 OxyDetect The installer and/or the user of the analyser must be satisfied that the installation complies with any “Special Conditions for Safe Use” or “Schedule...

 Open the catalog to page 5

SECTION 2 EU DECLARATION OF CONFORMITY

 Open the catalog to page 7

SERVOMEX \ EU DECLARATION OF CONFORMITY This declaration of conformity is issued under the sole responsibility of the manufacturer. Equipment Type:    OxyDetect (0531 IB) series oxygen deficiency gas detector Directive    ATEX 2014/34/EU (Applicable to items located in a hazardous area) Type Examination Certificate & Notified Body:    Baseefal5ATEX0165X, SGS Baseefa; 1180 Harmonised standards used: EN 60079-0:2012 + A11:2013 EN 60079-11:2012 EN 60079-15:2010 Directive    Low Voltage (LVD) 2014/35/EU (Applicable to items located in a non-hazardous area) Harmonised standards used: EN...

 Open the catalog to page 9

SECTION 3 ATEX CERTIFICATION Certificate No.: 05311B OxyDetect has been certified for: Rated Supply Voltage 10 - 28Vdc II 3G Special Conditions of Safe Use: 1. Possible electrostatic risk - clean only with a damp cloth. 2. The equipment is not capable of passing the dielectric strength test prescribed by the standards, and so this must be taken into account during installation. 3. The equipment shall be used in an area not exceeding pollution degree 2 as defined in IEC 60664

 Open the catalog to page 11

Issued 2 November 2015 Page 1 of 2 1    TYPE EXAMINATION CERTIFICATE 2    Equipment Intended for use in Potentially Explosive Atmospheres 5    Manufacturer:    Servomex Group Limited 6    Address:    Jarvis Brook, Crowborough, East Sussex, TN6 3FB 7 This equipment and any acceptable variation thereto is specified in the schedule to this certificate and the documents therein referred to. 8 Baseefa certifies that this equipment has been found to comply with the Essential Health and Safety Requirements relating to the design and construction of equipment of Category 3 intended for use in...

 Open the catalog to page 13

Certificate Number Baseefal 5ATEX0165X Issued 2 November 2015 Page 2 of 2 13    Schedule 14    Certificate Number Baseefal5ATEX0165X 15 Description of Equipment The OxyDetect is designed to monitor ambient oxygen levels, providing a local reading on a display; analogue output; Modbus Ethernet output and relay alarm volt-free contacts. The equipment comprises several PCB assemblies, including a para-magnetic oxygen sensor, all contained within a metal enclosure measuring approximately 160mm x 200mm x 70mm (H x W x D). All electrical connections apart from the Modbus Ethernet connections are...

 Open the catalog to page 14

SECTION 4 IECEx INTERNATIONAL CERTIFICATION Certificate No.: 05311B OxyDetect has been certified for: Rated Supply Voltage 10 - 28Vdc Ex ic nA nC IIC T4 Gc (-5°C ≤ Ta ≤ +45°C) Special Conditions of Safe Use: 1. Possible electrostatic risk - clean only with a damp cloth. 2. The equipment is not capable of passing the dielectric strength test prescribed by the standards, and so this must be taken into account during installation. 3. The equipment shall be used in an area not exceeding pollution degree 2 as defined in IEC 60664

 Open the catalog to page 15

lECEx Certificate of ConformityINTERNATIONAL ELECTROTECHNICAL COMMISSION IEC Certification Scheme for Explosive Atmospheres for rules and details of the lECEx Scheme visit www.iecex.com Certificate No.: Status: Date of Issue: Applicant: Electrical Apparatus: Optional accessory: Certificate history: 2015-11-02 Servomex Group Limited Jarvis Brook, Crowborough, East Sussex, TN63FB United Kingdom Approved for issue on behalf of the lECEx Certification Body: Position: OxyDetect Intrinsic Safety Ex ic nA nC IIC T4 Gc (-5°C < Ta < +45°C) R S Sinclair Technical Manager Signature: (for printed...

 Open the catalog to page 17

lECEx Certificate of Conformity Certificate No.:    lECEx BAS 15.0112X Jarvis Brook, Crowborough, East Sussex, TN63FB United Kingdom Additional Manufacturing location (s): This certificate is issued as verification that a sample(s), representative of production, was assessed and tested and found to comply with the IEC Standard list below and that the manufacturer's quality system, relating to the Ex products covered by this certificate, was assessed and found to comply with the lECEx Quality system requirements. This certificate is granted subject to the conditions as set out in lECEx...

 Open the catalog to page 18

lECEx Certificate of Conformity Certificate No.: Date of Issue: Schedule EQUIPMENT: Equipment and systems covered by this certificate are as follows: The OxyDetect is designed to monitor ambient oxygen levels, providing a local reading on a display; analogue output; Modbus Ethernet output and relay alarm volt-free contacts. The equipment comprises of several PCB assemblies, including a para-magnetic oxygen sensor, all contained within a metal enclosure measuring approximately 160mm x 200mm x 70mm (H x W x D). All electrical connections apart from the Modbus Ethernet connections are made to...

 Open the catalog to page 19

All SERVOMEX catalogs and technical brochures

  1. 4900 Multigas

    4 Pages