Universal temperature transmitter for wall installation. The T2N offers excellent measurement accuracy and stability wherever a wallmounted sensor is equired ● Dimensions 127 x 65 mm ● Long term stability 0,1% over 12 month ● Quick installation ● 2 output signals en option TECHNICAL SPECIFICATIONS Output signal: 4-20 mA, 0 to 10 Vdc Sensor type: PT100 acc. IEC 751 Linealitzation: analog optional by µP. Measurement range: -20/+80º and 0 to 50ºC (customer specific ranges under request as option) Accuracy: 0,3ºC (from -10 to m50ºC) Supply voltage mA output: 10 to 30 Vdc (polarity protected) Supply voltage Vdc output: 18 to 30 Vdc Load resistance (output mA): R>(U-12)/0,02 Load resistance (output Vdc): >3 Kohm Supply voltage mA output (-40ºC version): 8 to 40 Vdc (polarity protected) Response time: 10 seg Load influence: Negligible Standard resolution: 0,3 µA Input cable: PG7 Wiring connection: by screw terminal Supply: 10 to 30 Vdc Working temperature: Standard version -20/65ºC (optional -40 to 85 ºC) Storage temperature: -40 to 85ºC Humidity: 10 a 95% non-condensed Shock and vibration resistance: 4 g/2 to 150Hz as per IEC 60 068-26 Electromagnetic Compatibility (EMC): According GB/T17626.2 . IEC 61000-4-3 Degree of protection: IP 65 Dimensions and weight: 127 x 65 mm, 130 gr. Housing material: ABS Measuring stem: carbon steel nickel plated Fixing to wall: by screws Other output ranges: From -40 to +80ºC Extended temperature range: Environmental temperature -40 to 80ºC Separate sensor: Separate stem with PVC cable Weather protection cap for outdoor use: Multiplate natural ventilation radiant screen for wall installation Programmable range. Unit programable by PC Hart protocol output: Ou