Liquid Flow Meter SLG
7Pages

{{requestButtons}}

Catalog excerpts

Liquid Flow Meter SLG - 1

SLG Liquid Flow Meter Series Ultra-High Pressure Flow Meter for LC Applications ■ Ultra-low Flow Sensing in the nl/min Range ■ Operating Pressure up to 1200 bar (17500 psi) ■ 40 ms Response Time Interface Options Product Summary The SLG Liquid Flow Meter enables precise, bi-directional measurements of dynamic liquid flow rates in the nl/min to low pl/min range. This device has been designed for ultra-high pressure applications in Liquid Chromatography Systems (UHPLC). Excellent bio-compatibility and chemical resistance is ensured by the use of fused silica, stainless steel and PEEK for all wetted parts. Analog - Voltage Output (0-10 V) - Additional operation modes For more information on communication, please refer to page 2 of this document. Table 1: Model specific performance of SLG (all data for medium H2O, 23°C) aFlow rate at which the sensor output saturates. See Section 2 for performance between full scale and saturation point bLower/upper limit cMeasured value dNo dependence on the mounting orientation for SLG-0025 and SLG-0075. SLG-0150: maximum additional offset when mounted vertically. eNon-condensing, flow path empty

Open the catalog to page 1
Liquid Flow Meter SLG - 2

1.1 Precision/Extended Mode with the SLG-0075 The SLG-0075 is offering two different flow sensing modes. The Precision Mode allows for precise flow sensing starting at zero flow while the Extended Mode is made for a rough estimation of the current mass flow in higher flow ranges. The Precision Mode is fully bi-directional with continuous output at and around zero. The Extended Mode is uni-directional (positive flow direction only), and sensor output is restricted to the flow range between 2000 nl/min to 22000 nl/min. The calibration for each mode is stored on a separate calibration field...

Open the catalog to page 2
Liquid Flow Meter SLG - 3

2 Specifications Charts Accuracy (% of measured value) Accuracy (% of measured value) Accuracy (% of measured value) Accuracy (%of measured value) Flow rate (nl/min H20) SLG-0075 Extended Mode Absolute Accuracy with H20 Figure 1: Flow meter accuracy and repeatability across the flow range. Relative error in % of measured value (left column) and absolute error in nl/min (right column) for H2O www.sensirion.com

Open the catalog to page 3
Liquid Flow Meter SLG - 4

Communication with the Sensor Table 3: Electrical pinout 3.4 Digital Communication via RS485-Bus The SCC1-RS485 Sensor Cable for flow sensors allows the communication via RS485 interface for use in a demanding industrial automation environment. In addition to the standard commands available in the

Open the catalog to page 4
Liquid Flow Meter SLG - 5

4 Fluidic Connection Table 5: Fluidic Specifications ° ° lia—° 2x coned port 10-32 UNF, Upchurch and nanoViper™ compatible

Open the catalog to page 5
Liquid Flow Meter SLG - 6

For OEM applications, the flow meter can be purchased in larger quantities without any additional parts. For optimum performance, Sensirion recommends using the SLG flow meters in combination with the SCC1 interface cables. For laboratory use and technology evaluation, the Flow Meter Kit SLG-xxxx can be ordered. Ordering Information This laboratory-package contains: ■ Liquid Flow Meter SLG-Oxxx ■ Fitting Material for UHP fittings ■ PC Software (Viewer & Data Export Tool) ■ SCC1-USB Sensor Cable with USB connector for plug-and-play connection to a PC ■ SCC1-Analog Sensor Cable with...

Open the catalog to page 6
Liquid Flow Meter SLG - 7

Important Notices Warning, personal injury Do not use this product as safety or emergency stop devices or in any other application where failure of the product could result in personal injury (including death). Do not use this product for applications other than its intended and authorized use. Before installing, handling, using or servicing this product, please consult the datasheet and application notes. Failure to comply with these instructions could result in death or serious injury. If the Buyer shall purchase or use SENSIRION products for any unintended or unauthorized application,...