SCION 456-GC
1 /7Pages

SCION 456-GC

SCION 456-GC
1 /7Pages

Catalog excerpts

SCION 456-GC-1

• SCION 456-GCSpecification Sheet Dimensions and Weights Height: 57 cm (22.5 in.), Width: 66 cm (26.0 in.), Depth: 56 cm (22.0 in.) Weight*: 43 kg (59lb) * Typical values Environmental Conditions Operating temperatures: 10°C to 40°C. Operating humidity (relative): 5% to 95% Line voltage requirements: 120 V, 230 V (±10% nominal) Column Oven Dimensions: 23 cm (w) x 11 cm (d) x 28 cm (h) Temperature range: • Ambient: +4°C to 450°C • Liquid CO2:-60°C to 450°C Temperature program ramps/holds: 24/25 Maximum temperature ramp rate: • 150°C/min for all voltages and 180°C/min with insert accessory Cool down rate: 400 °C to 50 °C in 4.5 minutes Temperature set-point resolution: 0.1 °C Ambient temperature reject <0.01°C change in oven for 1°C change in ambient temp. Retention Time Repeatability <0.008% or < 0.0008 min, based on n-Pentadecane under temperature programming conditions. Area repeatability < 1% RSD General Specifications Up to 9 EFC modules total, injector, detector and auxiliary Optional backflush GC Control: External events (digital output): 8 standard 8 optional, total 16 Max number of timed events: 30 # Heated zones: Standard 9 Two power outlets 24V (1A max. each) SCION Instruments 1 Bain Square, Kirkton Campus Livingston, UK EH54 7DQ

 Open the catalog to page 1
SCION 456-GC-2

Methods: • Maximum stored internal methods: 50 (max. 30 alphanumeric characters) Logging: • Run log file (stored with the chromatogram when using CompassCDS) Error log file Local Display: • TFT full color screen • WVGA resolution (800 x 480) Size 23 cm (9”) Local Control: • Touch screen • Hard Keys Languages: • English, German, French, Spanish, Italian, Portuguese, Cyrillic, Kanji, Chinese (standard and traditional), Thai, Korean and Dutch Local automation: • Method lines: 25 • Modes: Infinite looping Dual and duplicate injection Communication Ethernet: Protocol: TCP/IP Data rate: 100 Mbps Control: • GC...

 Open the catalog to page 2
SCION 456-GC-3

Injector Options Maximum injectors: two, operating concurrently. Pneumatics: Electronic Flow Control (EFC): • S/SL Split/Splitless injector* • PTV Programmable Temperature • Vaporizing* COC Cold On-Column injector* • Flash injector • PWC Packed/ Wide bore On-Column injector * including septum purge S/SL Split/Splitless Injector Pressure range: 0-150 psi Total flow: • 500 mL/min for N2/Ar Maximum temperature: 450°C Split range: 1-10,000 (column dependent) Suited for columns: • Wide bore: (0.53 mm) COC Cold On-Column Injector Pressure range: 0-150 psi Total Flow: 50 mL/min...

 Open the catalog to page 3
SCION 456-GC-4

Temperature range: • Ambient + 10 °C to 450°C using air cooling • -160 °C to 450 °C using liquid N2 cooling • -60 °C to 450 °C using liquid C02 cooling Maximum temperature ramp rate: 200°C/min Temperature ramps/holds: 24/25 Split range: 1-10,000 (column dependent) Operational capabilities: • Large volume injection • Temperature ramped splitless • Cold on-column Suited for columns: • Wide bore (0.53 mm) • Narrow bore (0.05 to 0.32 mm) Maximum injection volume: 250 qL (LVI mode) Flash Injector Pressure range: 0-150 psi • Total flow: 50 mL/min (Type 23 EFC) Maximum temperature:...

 Open the catalog to page 4
SCION 456-GC-5

Available traps: Two lengths A wide range of standard packing and custom packings Detector Options Maximum detectors: two, operating concurrently (one of which is a Single or Triple Quad MS) Pneumatics: Electronic Flow Control (DEFC) Detector types: FID Flame Ionization Detector TCD Thermal Conductivity Detector ECD Electron Capture Detector NPD NPD Nitrogen-Phosphorus Detector PFPD Pulsed Flame Photometric Detector PDHID Pulsed Discharge Helium Ionization Detector MS Mass Spectrometry (see GC/MS brochure and datasheet) Note: Data Acquisition Rate: 600Hz for all detectors, (exception pfpd) FID...

 Open the catalog to page 5
SCION 456-GC-6

PFPD Pulsed Flame Photometric Detector Photomultiplier tube: • s/p Maximum temperature: 450 °C Detectivity: • S: 1 pg S/sec (S/p tube) • P: 100 fg P/sec (S/p tube) • N: 20 pg N/sec (S/p/n tube) Linear dynamic range: • S: 103 • Up to 23 elements can be detected PDHID Pulsed Discharge Helium Ionization Detector Detectivity: 50 ppb (Methane) Linear dynamic range: 104 (Methane) Operational quality: • Gold plated connections • Welded column connections Detectors (DEFC) Module types: 6 detector-specific modules Accuracy: ± 7 % set point flow Resolution: 0.1 or 1 mL/min Automation Options 8410...

 Open the catalog to page 6
SCION 456-GC-7

8400 Autosampler Sample capacity: 100 x 2 mL vials Large solvent wash vial: 2 x 120 mL* Dual and duplicate mode Internal standard addition Modes of operation: • Liquid • Ambient headspace* • spme* • Sample heating and cooling* • Pre-programmed modes of injection Syringes: • 1|mL, 2 mL,5 mL,10 mL,100 mL,250 mL for liquid injection • spme * Optional Optionally a CTC-PAL autosampler, including specific options may be installed. SCION Instruments 1 Bain Square, Kirkton Campus Livingston, UK EH54 7DQ

 Open the catalog to page 7

All SCION INSTRUMENTS catalogs and technical brochures

  1. SCION GC Series

    12  Pages

  2. compassCDS

    8  Pages

  3. SCION 436-GC

    7  Pages

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.