1. Catalogs
  2. SCHMIDT + HAENSCH
  3. S+H Refractometer ATR-BR

S+H Refractometer ATR-BR

S+H Refractometer ATR-BR

S+H Refractometer ATR-BR

Product catalog summary
Specifications
  • Measuring Scales: Refractive Index (RI), Sucrose (%Brix), others on request.
  • Measuring Range: 1.32000 - 1.54000 RI / 100% Brix.
  • Resolution: 0.00001 RI / 0.01% Brix.
  • Precision: ± 0.00002 RI / ± 0.01% Brix for 1.32 - 1.44 RI; ± 0.00003 RI / ± 0.01% Brix for 1.44 - 1.54 RI.
  • Reproducibility: ± 0.00001 RI / ± 0.01% Brix.
  • Temperature: Ambient +10° to +40°C; Automatic compensation +5° to +50°C.
  • Temperature Measurement: NTC sensor inside the prism.
  • Measurement Mode: Single measurement.
  • Prism: Sapphire.
  • Light Source/Wavelength: LED, interference filter 589 nm.
  • Display: Back-lit LCD, 16 x 16 characters.
  • Operation: 20 key membrane including function keys.
  • Interfaces: 1 x RS232 C serial and parallel.
  • Conformity: International Pharmacopoea, ASTM, AOAC, DIN, FDA, ICUMSA, and others.
Highlights
  • Stand-alone device with high quality and precision.
  • Very stable measurements and user-friendly operation.
  • Excellent price-performance ratio.
  • Very flat and small sample room for easy cleaning.
Applications
  • Determination of refractive index, dry substance, and mass percent.
  • Brix measurement and qualitative analysis for sample identification.
  • Quantitative analysis of dissolved solids in water or other solvents.
  • Typical industries: Sugar, Beverage, Food, and Chemical.
See more

Catalog excerpts

S+H Refractometer ATR-BR-1

Q LABORATORY Q PROCESS Q SOFTWARE Q AUTOMATION fTT?! SCHMIDT jH haensch V_J innovators by tradition since 1864 ATR-BR Refractometer The generalist with excel lent price performance ratio

 Open the catalog to page 1
S+H Refractometer ATR-BR-2

SPECIFICATIONS ATR-BR Refractive Index (RI), Sucrose (%Brix), others on request Measuring range 1.32000 - 1.54000 RI / 100% Brix Precision Measuring range 1.32 - 1.44: ± 0.00002 RI / ± 0.01% Brix * Measuring range 1.44 - 1.54: ± 0.00003 RI / ± 0.01% Brix * Automatic temperature compensation + 5° to + 50°C Temperature measurement NTC sensor for measurement of sample temperature placed inside the prism Measurement mode Single measurement Light source / wavelength LED, interference filter 589 nm Display Back-lit LCD, 16 x 16 characters Operation 20 key membrane including function keys Interfaces...

 Open the catalog to page 2

All SCHMIDT + HAENSCH catalogs and technical brochures

  1. iPR B3