  • Optical NDIR sensor technology
  • Sensitivity range: 0-5000 ppm CH4
  • Minimum detectable level: 100 ppm
  • Output: Modbus RTU 9600 bps
  • No periodic calibration required
  • Annual bump test and zero recommended
  • Life expectancy: > 10 years
Description

The document describes the Model SA-5100-CH4, a highly sensitive CH4 sensor utilizing non-dispersive infrared (NDIR) technology for methane detection. It is designed for environments requiring reliable gas monitoring without frequent calibration.

Specifications
  • Electrical Specifications: Input Voltage: 10-30 VDC, Nominal 24VDC; Optional 110/220 VAC Module; Power Consumption: 400 milliwatts (gas measurement module), ~2.5 watts (with color local display); Serial Output: Modbus RTU 9600 bps; Optional Local Alarm Relays: Alarm level 1 & 2 + Fault, Form “C” 2A contacts.
  • Environmental: Temperature Range: –40°F to +158°F (-40°C to +70°C); Humidity Range: 15-95% non-condensing.
  • Mechanical: Enclosure Material: Electropolished 316 stainless steel; Dimensions: 3.4” OD, 7.35” L; 12” W, 5” D, 4.75” H.
  • System: Detector Type: Non-dispersive infrared (NDIR) HC absorption; Range of Sensitivity: 0-5000 ppm CH4; Minimum Detectable Level: 100 ppm ± 25 ppm; Detection Method: Diffusion captive open path optical chamber; Warranty: 2 Years.
Area Classification

Flame Proof: Class I; Division 1; Groups BCD; Class I; Zone I; Group IIB +H2.

Optional Components and Accessories
  • Various junction boxes and modules including LoRaWAN and MODBUS transmitter modules.
User Interface Options

Includes optional digital display and a portable user interface kit with a magnetic wand.

OpCheck Performance Diagnostics
  • Zero stability, response time, span drift, signal stability, and signal strength are key performance metrics.
Response Factor Data

Provides response factors for various gases, with methane having a response factor of 1.0.

See more

Catalog excerpts

IR-5100-CH4-1

Highly Sensitive CH4 Sensors Features – Functions - Benefits • • Minimum detectable level 100 ppm No periodic calibration required Annual bump test and zero recommended Optical NDIR sensor technology

 Open the catalog to page 1
IR-5100-CH4-2

System Detector Type Range of Sensitivity Minimum Detectable Level Detection Method Warranty Enclosure Material Non dispersive infrared (NDIR) HC absorption 0-5000 ppm CH4 100 ppm ± 25 ppm Diffusion captive open path optical chamber 2 Years Dimensions (with IP 66 cover) Optional Digital Display OpCheck Performance Diagnostics Input Voltage Magnetic Wand and Safeguard Analytics portable user interface kit PN 909-80005 MODBUS master with data logger Zero stability Response time Span drift Signal stability Signal strength Power Consumption Serial Output Optional Local Alarm Relays 10-30 VDC, Nominal...

 Open the catalog to page 2

All Safeguard Analytics catalogs and technical brochures

  1. SA-4ES

    1  Page

  2. SCX

    2  Pages

  3. SA-2ES

    1  Page

  4. SA-XES

    1  Page

  5. SA2

    1  Page

Archived catalogs

  1. SA-XDS

    2  Pages

  2. IR-5100L-CH4

    2  Pages

  3. SA-200

    2  Pages

  4. Model SA-XES

    2  Pages

  5. IR-5100D-CH4

    2  Pages

  6. SA-1N4

    2  Pages

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.