Overview: The chlori::lyser by s::can GmbH is a plug & measure device designed for monitoring free and total chlorine levels in drinking and pool water. It operates on an amperometric principle with a membrane-covered sensor, ensuring long-term stability and minimal maintenance.
Key Features:
  • Stable readings despite fluctuations in pH, temperature, and flow.
  • Low cross-sensitivity to surfactants.
  • Factory pre-calibrated for ease of use.
  • Measures temperature in addition to chlorine levels.
Technical Specifications:
  • Measuring Principle: Amperometric, potentiostatic 3-electrode system.
  • Resolution: 0.001 mg/l for E-520-1 and E-525-1; 0.01 mg/l for E-520-2 and E-525-2.
  • Response Time (T90): 2 minutes.
  • Power Supply: 9 to 30 VDC, with typical power consumption of 0.5 W.
  • Operating Temperature: 0 to 45°C; Operating Pressure: 0 to 3 bar.
  • Dimensions: 35 x 208 mm; Weight: 150 g.
  • Conformity: EMC EN 61326-1.
Installation and Operation:
  • Mounting: Flow cell installation with quick connect process connection.
  • Recommended Flow: 15 to 30 l/h in s::can flow cell.
  • Interfaces: Compatible with s::can terminals via sys plug (IP67) and RS485.
Recommended Accessories:
  • con::cube V3 and con::lyte for integration.
  • moni::tool Software for operation.
  • Flow cell for four s::can physical probes.
  • Connection cables and spare parts for electrolytes and membrane caps.
See more

Catalog excerpts

chlori::lyser-1

Messgeräte Sonstige Daten chlori::lyser chlori::lyser monitors free chlorine (Cl2 + HOCl + OCl-) or total chlorine (free chlorine + combined chlorine) ∙ s::can plug & measure ∙ measuring principle: amperometric (membrane covered) 24 ∙ ideal for drinking and pool 55,2 ∙ long term stable and lowest maintenance in operation ∙ replacement of membrane only once a year ∙ readings stable even at high fluctuations of pH, temperature and flow ∙ low cross sensitivity to many surfactants ∙ compensates fluctuations of pH in an unmatched way pH range from 4 to 9 FCl; pH range from 4 to 12 TCl ∙ factory precalibrated ∙ mounting and measurement in a flow cell ∙ operation via s::can terminals & s::can software ∙ additionally also measures temperature recommended accessories part number article name D-330-xxx con::cube V3 D-320-xxx con::lyte S-11-xx-moni moni::tool Software F-45-four flow cell for four s::can physical probes C-1-010-sensor 1 m connection cable for s::can physical and ISE probes E-525-1/2-KIT Total Chlorine electrolyte and membrane cap (spare parts) E-520-1/2-KIT Free Chlorine electrolyte and membrane cap (spare parts)

 Open the catalog to page 1
chlori::lyser-2

technical specification measuring principle measuring principle detail resolution amperometric potentiostatic 3-electrode system E-520-1 and E-525-1: 0.001 mg/l E-520-2 and E-525-2: 0.01 mg/l) automatic compensation instrument temperature automatic compensation cross pH sensitivities response time (T90) 2 min. integration via con::cube con::lyte con::nect power supply 9 ... 30 VDC power consumption (typical) 0.5 W power consumption (max.) 0.6 W interface to s::can terminals housing material weight (min.) dimensions (Ø x l) operating temperature operating pressure installation / mounting process...

 Open the catalog to page 2

All S::can catalogs and technical brochures

  1. con::cube V3

    2  Pages

  2. chlodi::lyser

    2  Pages

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.