Cells, Calibration Standards and Accessories for Automatic Polarimeters
6Pages

{{requestButtons}}

Catalog excerpts

Cells, Calibration Standards and Accessories for Automatic Polarimeters - 1

Cells, Calibration Standards

Open the catalog to page 1
Cells, Calibration Standards and Accessories for Automatic Polarimeters - 2

Two Temperature Control Solutions 1. Patented TempTrol™ Technology Eliminates the Need for a Water Bath 2. Temperature Control With Water Bath Glass Stopper TempTrol heating and cooling transfer surface TempTrol cell with Temperature is mating heating selected via touch and cooling screen. Temperature transfer surface selection of 20°C shown above. Place the TempTrol cell in your TempTrol equipped Autopol® IV or V sample chamber to measure to within ±0.2°C of the USP, EP, JP or BP specified temperature (normally 20°C or 25°C ±0.5°C) Water Inlet NIST Thermometer Rudolph provides a...

Open the catalog to page 2
Cells, Calibration Standards and Accessories for Automatic Polarimeters - 3

Call us to select a cell that meets your unique application TYPE Type 40T TempTrol Polarimeter Sample Cells: These cells are constructed of stainless steel and Delrin® and come with Luer® fittings at either end of the cell for easy filling with a plastic syringe or pipette. The cells have removable metal collars that allow for easy cleaning and glass cell end replacement. Select Hastelloy® “H” when working with strong acids or bases. All cells come with Luer® lock and plug stoppers. *Select HP for Automation or High Pressure Applications. Type 14 Polarimeter Sample Cells: Pyrex glass with...

Open the catalog to page 3
Cells, Calibration Standards and Accessories for Automatic Polarimeters - 4

Cells That Meet The Most Demanding Applications Selection Autopol® 880 Dual Wavelength Saccharimeter With over 50 cells to choose from, Rudolph Research Analytical has a cell to meet every application. Both instruments shown with pour in sampling Temperature Probe HPLC Tubing HPLC Fitting Figure 3: Continuous Sampling Picture C: Pour In Sampling With Temperature Correction Picture D: Micro Volumes for HPLC Continuous Sampling Capability Autopol®s Accept Sample Cells from 10mm to 200mm Long Some polarimeters accept only special sample cells, with maximum lengths of 100mm. Autopol®s accept...

Open the catalog to page 4
Cells, Calibration Standards and Accessories for Automatic Polarimeters - 5

NIST Traceable Calibration Standards Picture E TempTrol Quartz Plate Picture F Standard Quartz Plate TempTrol Quartz Plates allow for temperature control and correction. Standard Quartz Plates allow for temperature correction only. Manufactured to NIST and ICUMSA Standards Rudolph Research Analytical Quartz Control Plates provide a permanent calibration standard for polarimeters and saccharimeters. Every Rudolph calibration standard is manufactured using extremely stable crystalline quartz which is manufactured and polished to specific planeness, parallelism, purity and axis standards....

Open the catalog to page 5
Cells, Calibration Standards and Accessories for Automatic Polarimeters - 6

Call Us To Select A Calibration Standard That Meets Your Application TempTrol™ Quartz Plates indicated by the letter “T” have the facility for automatic temperature control when purchased with an Autopol® IV or V with TempTrol TempTrol Quartz Control Plate A700T-1 Standard Quartz Control Plate A700-1 Has a rotation between +10 to +35 angular degrees (˚Arc) or +28 to +102 International Sugar Scale degrees (˚ISS)/(˚Z). The Quartz Control Plate consists of one right turning quartz plate. Standard Rotations at 589 nm: 10.5°, 11°, 11.5°, 12°, 12.5°, 13°, 15°, 16°, 20°, 21.5°, 23°, 34°, 34.5°...

Open the catalog to page 6