CLS-23 Capacitive Level Meter Riels® Instruments
5Pages

{{requestButtons}}

Catalog excerpts

CLS-23 Capacitive Level Meter Riels® Instruments - 1

CAPACITIVE LEVEL INDICATOR LEVEL CONTROLS Capacitive level sensors CLS – 23 Capacitive level sensors (switches) Riels CLS–23 are designed for limit level detection of electrically conductive and non-conductive fluids in vessels, reservoirs, sumps, pipes, tanks, etc.The sensitivity of the sensor can be easily set by placing magnetic pen on a sensitive spot.The connection is done by means of two wires directly into a circuit with a relay or to binary input of a control system. • For level detection of electrically conductive and non-conductive fluids • Compact miniature performance • Easy setting by means of magnetic pen • Direct mounting to vessels, tanks, sumps, tubes • High temperature performance available • 2-wire connection directly to relay circuit (current switch) • LED state indication Sensor variants All data and contents of this card are the exclusive property of Riels Instruments Uncoated short bar electrode, for sensing of electrically non-conductive liquids (mineral and plant oils, resins, etc.). Mounting in horizontal position. Electrode length 30 mm. Insulated (coated) short bar electrode, for non-aggressive electrically conductive liquid sensing (water, water solutions). The insulation is made from polypropylene (PP). Electrode length 30 mm. Insulated (coated) short bar electrode, for moderately aggressive electrically conductive liquid sensing (chemicals, water, moderately aggressive water solutions). Higher temperature resistance than variant "11". The insulation is made from FEP. Electrode length 30 mm. Partly insulated rod electrode, for level detection of conductive and non-conductive liquids, partially resistant to vapours (water) condensation in the sensing area. The insulation is made from FEP. Vertical mounting; horizontal mounting (from the side) is possible for shorter electrodes (up to 200 mm). Electrode length from 50 mm to 1 m. Fully insulated rod electrode, for universal use, for level detection of conductive liquids (water, water solutions). Resistant to vapours (water) condensation in the sensing area and partially resistant to medium spraying. The insulation is made from FEP. Vertical mounting; horizontal mounting (from the side) is possible for shorter electrodes (up to 200 mm). Electrode length from 50 mm to 1 m. Uncoated removable rod electrode, for level detection of conductive and non-conductive liquids. Vertical mounting; horizontal mounting (from the side) is possible for shorter electrodes (up to 200 mm). Electrode length from 50 mm to 1 m. Accessories Standard – included in the level sensors price Optional – for extra charge • Extra cables (over the standard length 2 m) • Non-rewirable connector plug M12 • Rewirable connector plug M12 • Steel welding flange ON –18x1,5 or ON–G3/8'' • Stainless steel welding flange NN –18x1,5 or NN–G3/8'' Safety, protections and compatibility The level sensor is equipped with a protection against electric shock on electrode, polarity, overvoltage and short-term current overload on the output. Electromagnetic compatibility is provided by conformity with standards EN 55011 / B, EN 61326-1, EN 61000-4-2 (8 kV), -4-3 (10 V/m), -4-4 (2 kV), -4-5 (1 kV) and -4-6 (10 V). Areas of applications Detection of various types of liquids - water, oils, coolants, water solutions, etc. Suitable for metal vessels, containers, tanks, sumps, tanks, pipes. Suitability for non-metallic containers (glass, plastic containers, etc.) please consult with the manufacturer.

Open the catalog to page 1
CLS-23 Capacitive Level Meter Riels® Instruments - 2

All data and contents of this card are the exclusive property of Riels Instruments Technical specifications Supply voltage 6 ... 30 V DC Supply current - OFF state max. 0.6 mA Remanent voltage - ON state max. 6 V Output time delay 0.1 s Ambient temperature range (around the housing with electronics) -20 ... +80°C Cable ("B" performance with cable gland) PVC 2x 0.34 mm

Open the catalog to page 2
CLS-23 Capacitive Level Meter Riels® Instruments - 3

Process connection Normal performance for non-explosive areas High temperature performance for non-explosive areas Metric thread Metric thread Pipe thread (BSP) Sealing thread Material performance Sensor part Housing Process coupling Electrode Electrode insulation Electrode insulation All variants All variants All variants CLS–23N–11 CLS–23_–12, 20, 21 Plastic material PP Stainless steel W.Nr. 1.4305 (AISI 303) Stainless steel W.Nr. 1.4305 (AISI 303) Plastic material PP Plastic material FEP Temperature and pressure resistance All data and contents of this card are the exclusive property of...

Open the catalog to page 3
CLS-23 Capacitive Level Meter Riels® Instruments - 4

Installation and recommendations Applies to: CLS–23_–20, 21, 30 For top mounting (vertical position) it is necessary to keep the distances from walls and from other sensors. Installation into the auxiliary gauging (calming) tube. We recommend to keep the tube diameter. All data and contents of this card are the exclusive property of Riels Instruments Applies to: All types variants Applies to: CLS–23_–20, 30 (only for electrically non-conductive liquids) CLS–23_–21 (for electrically conductive liquids) When installing the sensor into the pipe it is necessary to choose properly internal...

Open the catalog to page 4
CLS-23 Capacitive Level Meter Riels® Instruments - 5

Continuous light – Sensor is closed (switched ON) Dark – Sensor is open (switched OFF) Rapid flashing (period 0.2 sec.) * – Unrecognized upper and lower limits or setting mistake Slow flashing (period 0.8 sec.) – Short circuit at sensor output * Sensor for each flash of the LED switches its output on for approx. 3 ms. This period is sufficiently short to avoid unwanted switching of relay contacts. For binary inputs, we recommend to set the filter so as not to respond to pulses shorter than 3 ms. All data and contents of this card are the exclusive property of Riels Instruments Order code...

Open the catalog to page 5

All Riels Instruments catalogs and technical brochures

  1. miniELCOR

    4 Pages

  2. EL-USB-4

    5 Pages

  3. EE33

    6 Pages

  4. EE300Ex-HT

    6 Pages

  5. AVFM 5.0

    3 Pages

  6. EE 75

    5 Pages

  7. EE776

    6 Pages

  8. EE771/2

    6 Pages

  9. TRICOR

    15 Pages

  10. FFU

    7 Pages

  11. RIF600W

    4 Pages

  12. RIF180

    3 Pages

  13. MC 505

    1 Pages

  14. HM-R

    3 Pages

  15. HD408T

    2 Pages

  16. RIB300

    3 Pages

  17. 23/25 ED

    2 Pages

  18. RIB180

    4 Pages

  19. FLD48

    2 Pages

  20. CLM-36

    6 Pages

  21. DLM-35

    2 Pages

  22. DLS-27

    7 Pages

  23. CLS-23

    5 Pages

  24. LFP

    2 Pages

  25. GRLM70

    9 Pages

  26. RIL600

    5 Pages

  27. LITTLE EX

    1 Pages

  28. DEEPER

    1 Pages

  29. DCX22

    2 Pages

  30. MN03

    1 Pages

  31. LEO RECORD

    2 Pages

  32. EV220B

    13 Pages

  33. WWG

    1 Pages

  34. MATIC/V/MC

    2 Pages