Overview
The PH10M-iQ PLUS is an advanced metrology head that enhances the traditional PH10M PLUS by incorporating inferred qualification, reducing calibration time and increasing measurement efficiency. This innovation aligns with Renishaw's commitment to continuous product development.

Key Features
  • Inferred Qualification: Allows operation in all orientations without requalification after an initial simple procedure.
  • Dual Functionality: Combines the capabilities of the PH10M PLUS with inferred qualification for touch-trigger probing routines.
  • Performance Optimization: Users can enhance accuracy by increasing qualification positions during probe qualification.

Specifications
  • Dimensions: Length 117 mm, Width 62 mm, Weight 645 g.
  • Angular Movement: A-axis 0° to 105° in 7.5° steps, B-axis -180° to 180° in 7.5° steps, totaling 720 positions.
  • Repeatability: 0.4 µm at 62 mm from the A-axis center.
  • Temperature Range: Operating 10°C to 40°C, Storage -10°C to 70°C.
  • Compatibility: Compatible with all PH10M PLUS probes; inferred qualification is exclusive to touch-trigger probes.

Machine Specifications
  • Horizontal Arm Machine: MPEe = ± (9 + L / 100) µm.
  • Bridge Machine: MPEe = ± (1 + L / 750) µm.

Additional Information
  • Warranty: 2 years.
  • Contact: Renishaw plc, New Mills, Wotton-under-Edge, Gloucestershire, UK.

Disclaimer
Renishaw has made efforts to ensure document accuracy but excludes liability for inaccuracies. Renishaw and the probe emblem are registered trademarks.
See more

Catalog excerpts

PH10M-iQ PLUS-1

Renishaw plc New Mills, Wotton-under-Edge, Gloucestershire GL12 8JR United Kingdom Less time calibrating, more time measuring PH10M-iQ PLUS offers all the benefits of a traditional PH10M PLUS but with the addition of inferred qualification. This further demonstrates Renishaw’s commitment to continuous product development. PH10M-iQ PLUS increases throughput by removing the need to qualify each head position that is used. This allows more time to be spent measuring. Following an initial head localisation, a simple qualification procedure of as few as three positions allows the user to operate PH10M-iQ PLUS in every orientation possible without requalifying. Machine type Mounting orientation Qualification positions PLTI - Typical positional span* Two heads in one PH10M-iQ PLUS provides all the functionality of Renishaw’s industry standard PH10M PLUS but with the addition of inferred qualification for touch-trigger probing routines. Where an application or specific angle demands ultimate touch-trigger accuracy then PH10M-iQ PLUS can be qualified in the same way as a standard PH10M PLUS. Fine tune your performance PH10M-iQ PLUS must perform a simple qualification procedure before inferred mode can be used, after which any head position can be used for measurement without having to requalify. Depending on your metrology requirements you can optimise the accuracy of your system by increasing the number of qualification positions during probe qualification. Horizontal arm machine specification Using a PAA3 300 mm extension bar, standard force TP20 module, 10 mm x 4 mm diameter stylus on a machine with the following specification; MPEe = ± (9 + L / 100) μm (L in mm) ISO 10360-2 (2009) Bridge machine specification Using a PAA1 adaptor, standard force TP20 module, 10 mm x 4 mm diameter stylus on a machine with the following specification; MPEe = ± (1 + L / 750) μm (L in mm) ISO 10360-2 (2009) *Positional span of sphere location (PLTI) evaluated as per ISO 10360-5 (2010) but covering all 720 head

 Open the catalog to page 1
PH10M-iQ PLUS-2

Compatibility Renishaw’s complete range of PH10M PLUS compatible probes are also compatible with PH10M-iQ PLUS. However, the added functionality of inferred qualification can only be used with touch-trigger probes. TP20, TP2-5W, TP200, TP6, TP6A and TP7M touch-trigger probes can be used in traditional PH10 mode or can utilise inferred qualification; whereas SP25M and SP600M scanning probes can only be used in traditional PH10 mode and cannot utilise inferred qualification. Specification Head Probe mount Renishaw Autojoint (multiwire) M8 threaded probes can be used with a PAA probe adaptor 0.4...

 Open the catalog to page 2

All RENISHAW catalogs and technical brochures

  1. QC20 ballbar

    16  Pages

  2. RTP20

    2  Pages

  3. RLMD01_09

    13  Pages

  4. RLBD01_04

    9  Pages

  5. RLCD03_03

    9  Pages

  6. HiLin™

    20  Pages

  7. SP80

    4  Pages

  8. SP600

    4  Pages

  9. Precision styli

    60  Pages

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.
About DirectIndustry
DirectIndustry is a marketplace that offers B2B players a complete experience, from product sourcing to purchase. DirectIndustry is an entity of VirtualExpo Group.