Product catalog summary
Area of Use – Sewage Treatment Plants
Analysis of Dry Mass Content of Waste Water Sludge
The PMV 50 PLUS microwave moisture analyzer is a modern tool used in both laboratories and during the technological process for the rapid determination of dry residue in wastewater sludge. It offers speed and accuracy for testing samples obtained through various methods such as centrifugation or pressing in sewage treatment plants. The analysis duration is notably short, taking only a few minutes, which allows for efficient control over technological processes.
Main Features
  • Quick analysis without the need for costly and time-consuming calibration.
  • Reliable and accurate analysis, unaffected by sample consistency or composition.
  • Intuitive touch screen interface suitable for operators of all experience levels.
  • Volumetric drying for ultra-short drying duration.
  • Predefined drying programs.
  • Advanced process visualization and documentation with reports, graphs, and statistics.
Comparison of Methods
The determination of dry residue in wastewater sludge was conducted at the Municipal Waste Water Treatment Plant in Radom using both a reference method and the PMV 50 PLUS microwave radiation method. The reference method involves conventional heating at 105°C, while the PMV 50 PLUS uses microwave radiation interacting with water dipole particles in the sample. This can result in slight differences in water content results for the same sample.
Sample Analysis Results
Sample TypeReference MethodPMV 50 PLUSDifferenceAnalysis Duration (min)
Digested sludge (1)6.14% ± 0.04%6.40% ± 0.06%0.24%3 ÷ 4
Digested sludge (2)2.86% ± 0.07%3.02% ± 0.22%0.16%4.5
Excess sludge (3)4.02% ± 0.08%4.34% ± 0.07%0.32%∼ 4
Excess sludge (4)1.31% ± 0.05%1.50% ± 0.10%0.19%∼ 4
Sample Sources
  • (1) Flottweg centrifuge
  • (2) Closed digester chamber
  • (3) Buffer storage
  • (4) Open digester tank
Analysis Parameters for PMV 50 PLUS
  • Microwave power: 100%
  • Sample temperature: 90°C
  • Finish mode: Auto 2
  • Sample preparation: mixing
  • Method: Sample dried between two filters.
Summary
The PMV 50 PLUS microwave moisture analyzer is effective for determining the dry residue of wastewater sludge in sewage treatment plants.
Contact Information
Radwag Balances & Scales, RADWAG – Headquarters, Toruńska 5, 26-600 Radom, Tel.: +48 48 386 60 00, e-mail: [email protected], http://radwag.com
See more

Catalog excerpts

PMV 50 PLUS-1

AREA OF USE – SEWAGE TREATMENT PLANTS Analysis of Dry Mass Content of Waste Water Sludge PMV 50 PLUS Microwave Moisture Analyzer Method The PMV 50 PLUS microwave moisture analyzer is a modern solution used not only in laboratories, but also during the technological process for fast determination of dry residue in waste water sludge. PMV 50 PLUS ensures speed and accuracy when it comes to tests of various samples obtained either by centrifugation, or acquired from press or other technological instruments used in sewage treatment plants. Short analysis duration (barely few minutes) stand for economy and means possibility of control over technological instruments (feedback). Main Features  Quick analysis of any sample, free of costly and time-consuming calibration.  Reliability – easy-going, accurate analysis lacking any discrepancy due to differences in either consistency or composition of the sample.  Touch screen – intuitive, clear user interface, regardless of the experience level of the operator.  Volumetric drying - ultra short drying duration.  Defined drying programs.  Process visualisation and documenting advanced reports, grap

 Open the catalog to page 1
PMV 50 PLUS-2

Determination of dry residue in waste water sludge was carried out in the Municipal Waste Water Treatment Plant in Radom using a reference method and a microwave radiation method (PMV 50 PLUS). When it comes to the reference method, the sample is heated convectionally in temperature of 105oC, wheras in the case of PMV 50 PLUS moisture analyzer method there is an interaction of the microwave radiation with dipole particles (water) existing in the sample structure. In consequence there might be noted slight differences between the obtained water content results of the same sample. Reference method...

 Open the catalog to page 2
*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.