Group: Process Insights
Catalog excerpts

PROCESS ANALYZER PRODUCT DATASHEET ClearView® db Hydrogen Peroxide Vapor (HPV™) Analyzer System Measuring H2O2 and H2O Under Ambient or Vacuum Conditions • Real-time HPV sterilant gas monitoring • Proven technology works in vacuum and atmospheric • H2O2 and H2O concentration by direct optical NIR • Vital for resistometers and HPV cycle development • Independent reference for HPV load production • Optical RH and relative saturations
Open the catalog to page 1
Our HPV™ Analyzer System is a simple turnkey solution for the measurement of hydrogen peroxide and water (H2O2 and H2O) concentrations in vapor phase. These are both measured together because they are codependent. The analyzer operates in real time, which takes the guesswork out of determining the H2O2 and H2O concentrations during cycle development and throughout the actual sterilization cycle. The user gains continuous, accurate data for documentation and validation. Low Pressure Monitoring – Accurate, Reliable Results Unlike chemical sensors, the HPV Analyzer System is able to measure...
Open the catalog to page 2
Quick Installation and Start-up The complete system consists of an HPV Analyzer, one or two G-SST™ Vapor Probes, and a pair of fiber optic cables for each probe. The G-SST Vapor Probe mounts in the sterilizer chamber side wall, vapor transfer/recycle lines or optionally on the chamber inner roof with a fiber feed through sanitary cap. The HPV Analyzer System is pre-calibrated at the factory. No programming is required by the user. Start-up requires powering the HPV Analyzer System, connecting the probe or probes via the fiber optic cables, taking a ZERO reading in the dehumidified isolator...
Open the catalog to page 3
Fiber Cable • 00 µm Ultra Low-OH fiber with Stainless Steel SMA-905 connectors 5 • evlar® protected or armored cables available K G-SST Vapor Probe for HPV Analyzer System • perating temperature range: 5 °C to 90 °C O • 0 cm path length – double pass 25 cm 5 • uitable for vacuum (down to 1 x 10-5 Torr) and positive pressure (up to 1520 Torr, 2 atm) S • MA 905 fiber optic interface S • uilt-in validation filter (optional) B • robe may be mounted fully inside the isolator or through isolator wall using a 2 inch sanitary flange P The Smart Choice for Reliable HPV Measurement O ur HPV Analyzer...
Open the catalog to page 4
Specifications H2O2 Vapor Measurement Range: Enclosure Options Stainless Steel: Painted Carbon Steel: Sample Points (channel): Response Time: 1 second, minimum. User settable Relative to concentration at time of reference Operating Environment Ambient Temperature: Optimal Ambient Temperature Stability: < ±2 °C Relative Humidity:
Open the catalog to page 5
PROCESS ANALYZER GAIN REAL-TIME INSIGHT INTO YOUR PROCESS Process Insights manufactures and delivers premium sensors, monitors, detectors, analyzers, instrumentation, and software that are mission-critical to keep your operations, personnel, and the environment safe – every day across the globe. Get the most reliable, precision analytical technologies available on the market today. We will work to match your needs and budget, and provide the optimal, and most stable process analysis solution for your application. CENTERS OF EXCELLENCE | PROVIDING PROVEN SOLUTIONS Process Insights is...
Open the catalog to page 6All Process Insights catalogs and technical brochures
-
HALO Max QCL™
4 Pages
-
innovative water solutions
24 Pages
-
SERIES 2000
5 Pages
-
QuickTONultra
4 Pages
-
573 Dew Point Mirror™
8 Pages
-
T12
4 Pages
-
SERIES 3520™
4 Pages
-
SERIES 3500™ & 3510™
4 Pages
-
SERIES 3000™
4 Pages
-
SERIES 2520™
4 Pages
-
SERIES 2500™ & 2510™
4 Pages
-
ZRO 2000™
4 Pages
-
OXY-SEN™
4 Pages
-
SERIES 1300™
4 Pages
-
PCM 5000™
2 Pages
-
PCM 1000™
4 Pages
-
EVM™
4 Pages
-
Diamond 20™
4 Pages
-
Diamond MX™
4 Pages
-
Hydrocarbon SmartSystem
8 Pages
-
RefinIR™
4 Pages
-
RovIR™
4 Pages
-
Fiber Optic Cable
4 Pages
-
Turbidity Process Flow Cell
4 Pages
-
Teflon Process Flow Cell
4 Pages
-
Gas Process Flow Cell
4 Pages
-
Axial Process Flow Cell
4 Pages
-
G-SST™
4 Pages
-
RTD-SST
4 Pages
-
O-SST
4 Pages
-
SST
4 Pages
-
ClearView
4 Pages
-
LAB 508 UV-VIS
8 Pages
-
508 UV-VIS
8 Pages
-
LAB NIR-O
8 Pages
-
NIR-O™ PROCESS ANALYZER
8 Pages
-
QuickTOCtrace™
8 Pages
-
QuickTOCuv™
4 Pages
-
QuickCODlab™
8 Pages
-
ToxAlarm™
4 Pages
-
NitriTox™
4 Pages
-
QuickTOCeco™
4 Pages
-
QuickTOCairport™
6 Pages
-
RGA
6 Pages
-
MB
8 Pages
-
HRQ
6 Pages
-
APIMS
2 Pages
-
MAX-QMS
2 Pages
-
MAX300-EGA
8 Pages
-
PRODUCT DATASHEET MAX300-LG
4 Pages
-
MAX300-CAT
4 Pages
-
MAX300-BIO
8 Pages
-
MAX300-AIR
8 Pages
-
MAX300-RTG™ 2.0
8 Pages
-
MODEL ESS-SCVP™
8 Pages
-
HDT
8 Pages
-
XDT
8 Pages
-
XPDM
8 Pages
-
LPDT2
4 Pages
-
Spark H2 O in CO2
6 Pages
-
HALO KA H2O
3 Pages
-
HALO OK
8 Pages
-
HALO H2O in N2O
2 Pages
-
HALO H2
4 Pages
-
HALO 3 NH3
3 Pages
-
HALO 3 HCI
3 Pages
-
HALO 3 CH4
3 Pages
-
HALO 3 H2O
4 Pages
-
HALO 3 D2O/HDO
4 Pages
-
HALO 3 CO2
4 Pages
-
HALO 3 CO
2 Pages
-
HALO 3 CH2O
2 Pages
-
HALO KA Max
4 Pages
-
VeraSpec™ APIMS
2 Pages
-
973
8 Pages
-
573
8 Pages
-
473 Dew Point Hygrometer
8 Pages
-
Zircomat
8 Pages
-
9610
8 Pages
-
9800 Calorimeter
4 Pages
-
Flare Gas Analyzers
1 Pages
-
PRODUCT BROCHURE MAX300-LG
8 Pages
-
MAX300-RTG 2.0
8 Pages