Benefits
  • Utilizes EdgeTech's Optical Chilled Mirror (OCM) technology for high accuracy.
  • Portable and self-sufficient humidity calibration system.
  • Offers highest accuracy for Relative Humidity (RH) and Ambient Temperature (AT):
    • RH Accuracy: ±0.5%
    • RH Range: 5% to 95%
    • AT Accuracy: ±0.2°C
    • AT Range: 10°C to 50°C
  • Independent control of temperature and RH.
  • Automatic correction for mirror contaminants.
  • Certified measurements against NIST traceable standards.
  • Standard features included without extra charges.
Portable Relative Humidity Calibrator

The Model RH CAL is a microprocessor-based, programmable humidity calibration system suitable for both laboratory and field use. It provides NIST traceable humidity calibrations with high accuracy for RH and AT. The system allows independent control of temperature and humidity, making it adaptable to various calibration protocols. It employs EdgeTech’s OCM technique for traceability and feedback control, ensuring it is a standalone, portable system.

Unlike other systems, RH CAL is self-sufficient, requiring no compressed air or water connection, enhancing its portability. Features like the Automatic Balance Cycle (ABC) and integral ambient temperature probe ensure superior accuracy and fast response.

Specifications
  • Accuracy:
    • RH: ±0.5%
    • AT: ±0.2°C
    • Dew/Frost Point: ±0.2°C nominal
  • Temperature and RH Ranges:
    • RH: 5 to 95%
    • AT: 10 to 50°C
    • Dew/Frost Point: -40 to 60°C, D2 sensor
  • Slew Rate: 1.0°C (1.8°F) / second max., above 0°C
  • Dew/Frost Point and Temperature Sensors: 3-wire Platinum Resistance Thermometer (PRT), 100 ohms at 0°C, nominal
  • Repeatability: ±0.05°C
  • Hysteresis: Negligible
Measurement Performance

The RH CAL uses a unique control scheme with volumetric proportional control valves to maintain precise RH control. It modulates 'Dry' and 'Wet' valves to quickly generate and stabilize desired humidity levels. Continuous monitoring via the integral chilled mirror allows for instant reaction to changes in humidity, maintaining control stability.

  • Sample Flow Rate: 1 liter/minute (2.0 SCFH)
  • Operating Temperature: Control Unit: 0 to +50°C (+32 to +122°F)
  • Outputs: Analog (0-5VDC or 4-20mA) and RS232C serial port
  • Sensor Materials: Chromium, glass, epoxy, anodized aluminum
  • Power Requirements: 100 to 240 VAC, 50-60 Hz, 150 Watts Maximum
  • Display: Multi-Line Graphical LCD Display
  • Dimensions: 52.4W x 43.7D x 21.7H cm (20 5/8 W x 17 3/16 D x 8 9/16 H in)
  • Enclosure: Ultra high-impact structural copolymer carrying case
See more

Catalog excerpts

Product data sheet RH-Calibrator-1

The Model RH CAL is a microprocessor based, programmable humidity calibration system that is at home in the metrology lab or out in the field performing on site NIST traceable humidity calibrations. The system offers the highest accuracy available for both relative humidity and ambient temperature. Unlike other comparative systems RH CAL is entirely selfsufficient. It does not need compressed air or a water connection to operate, which allows this system to be truly portable. With the RH CAL, temperature and relative humidity are controlled independently; therefore you are not limited to performing calibrations at the surrounding ambient temperature, which may not be appropriate for your unique calibration protocol. Using EdgeTech's Optical Chilled Mirror (OCM) primary measurement technique for traceability and feedback control, RH CAL is a standalone, portable humidity calibration system • EdgeTech field proven Optical Chilled Mirror (OCM) technology • Completely self-sufficient and portable humidity calibration system • Highest accuracy available for both RH and temperature (AT): • Temperature and RH controlled independently • Automatic correction for mirror contaminants • Certified measurements against NIST traceable standards • Standard features, where others charge extra Continuing to RH CAL's ease of use are features such as maintenance reducing Automatic Balance Cycle (ABC), and integral ambient temperature probe and D2 chilled mirror sensor which are located in the sample chamber providing not only superior accuracy but the fastest response.

 Open the catalog to page 1
Product data sheet RH-Calibrator-2

Model RH CAL Specifications MEASUREMENT PERFORMANCE Temperature and RH Ranges RH: 5 to 95% AT: 10 to 50°C Dew/Frost Point: -40 to 60°C, D2 sensor Accuracy RH: ±0.5% AT: ±0.2°C Dew/Frost Point: ±0.2°C nominal Dew / Frost Point and Temperature Sensors 3-wire Platinum Resistance Thermometer (PRT) 100 ohms at 0°C, nominal FUNCTIONALITY Power Requirements 100 to 240 VAC, 50-60 Hz 150 Watts Maximum Sample Flow Rate 1 liter / minute (2.0 SCFH) Operating Temperature Control Unit: 0 to +50°C (+32 to +122°F) Outputs Analog (0-5VDC or 4-20mA) and RS232C serial port Depression 60°C (113°F), nominal, D2 sensor...

 Open the catalog to page 2