The OxyTrans LC OEM is a two-wire, loop-powered oxygen transmitter designed for accurate and economical measurement of oxygen at trace and percentage levels. It is suitable for various industrial applications.

Specifications
  • Measuring Ranges: 0-100 ppm, 0-1000 ppm, 0-10,000 ppm, and 0-25% O2.
  • Calibration: Can be calibrated with a calibration gas or air.
  • Accuracy: +/- 2% FSD at constant temperature, +/- 5% FSD from 0 to 50°C.
  • Resolution: 0.1 ppm for 0-100 ppm, 1 ppm for 0-1000 and 0-10,000 ppm, 0.1% for percentage range.
  • Response Time: 90% FSD at 25°C, varying from 10 to 45 seconds depending on the range.
  • Operating Temperature: 0 - 50°C.
  • Pressure: 0.1 - 1 bar.
  • Signal Output: 4-20 mA/DC.
  • Alarm Value: Standard at 3.8 mA, optional at 23 mA.
  • Voltage: 10-35 VDC with reverse voltage protection up to 40VDC.
  • Load: Typically 470 Ohm, maximum 750 Ohm.
  • Display: 6-digit alphanumeric with bar graph.
  • Housing: IP65 rated.
  • Size and Weight: 120 x 160 x 65 mm, 0.3 kg.

Features
  • Compact and inexpensive design.
  • Long operating life of the sensor.
  • IP65 enclosure for protection.
  • Reverse voltage protection and temperature compensation.
  • No cross-sensitivity to H2 gas.
  • Flow-through measuring cell made of synthetic material.

Applications
  • Nitrogen plants.
  • Gas manufacturers.
  • Metallurgical industries.
  • Chemical industry.
  • Heat treating and bright annealing.
See more

Catalog excerpts

Product data sheet OxyTrans LC OEM-1

OxyTrans LC OEM 4 – 20 mA Loop Powered two wire Oxygen Transmitter Two Wire Process Oxygenanalyzer The Model OxyTrans is simple, accurate and economic measurement of Oxygen at trace and % levels. ☺ One User specific measuring range The unit will offer with ranges between 0-100 ppm and 0 – 25% of oxygen. ☺ Sensor The OxyTrans uses a special fuel cell to measure the oxygen concentration. The sensor meets the industrial requirements for accuracy, sensitivity, easy to use and operating life. ☺ Calibration The calibration of the instrument for trace oxygen measurements in gas, should be done with a calibration gas. The concentration can be chosen freely within the measuring range. In the percent range the unit can calibrated with air. ☺ Features compact inexpensive Sensor with long operating life IP65 enclosure reverse voltage protection and temperature compensation no cross sensitivity to H2 gas ☺ Flow-through measuring cell in synthetic material ☺ The measuring cell is modular and is made of synthe

 Open the catalog to page 1
Product data sheet OxyTrans LC OEM-2

Response time Signal output Alarm value oxygen sensor : 10 - 35 VDC reverse voltage protection up to 40VDC : 6 Digits, alphanumeric with bar graph : Micro-Fuell Cell, ( uber alles ) Typical • different sensors available • 4 different Measuring ranges • synthetic material Cell block Applications are found in • Nitrogen plants • Metallurgical industries • Chemical industry • heat treating and bright annealing Neuer Bruchweg 1,46509 Xanten