Product data sheet DewMaster

Product data sheet DewMaster

Product data sheet DewMaster

Product catalog summary
DewMaster Highlights
  • Utilizes EdgeTech's proven chilled mirror technology with options for one, two, or three-stage sensors, either integral or remote. Comes with a NIST traceable calibration certificate.
  • Features a scrolling menu accessible via the front panel and RS232 interface.
  • Offers a depression range choice of 45, 65, or 95°C, depending on the cooling stage.
  • Includes Automatic Balance Control (ABC) for automatic correction of mirror contaminants and restandardization, enhancing unattended operation.
  • Dual alarm capability with user-programmable settings, available in latching or auto-reset modes.
  • Provides serial and analog outputs, including RS232 communication for remote programming and data output.
  • Supports various output formats (0-5 VDC, 4-20 mA, 0-10 VDC, or 0-20mA) that are user-scalable.
  • Includes a pressure transducer input for psychrometric variables.
Benefits
  • Measures dew point, absolute pressure, PPMv, and other psychrometric variables.
  • Wide measuring range from -75 to 100°C.
  • Automatic correction for mirror contaminants and certified measurements against NIST standards.
  • Standard features included without extra charges.
Applications
  • Suitable for use in annealing furnaces, process dryers, clean rooms, curing ovens, plant air, environmental chambers, lab calibration standards, emission testing, metrology, and continuous industrial processes.
Measurement Performance
  • Dew/Frost Point and RH Ranges: -40 to 60°C, -50 to 100°C, -75 to 100°C, 0 to 100%RH, 0 to 999,999 PPMv.
  • Accuracy: ± 0.2°C Dew/Frost Point, ± 0.5% of Full Scale Pressure.
  • Depression: 45°C, 65°C, 95°C with a slew rate of 1.7°C/second maximum above 0°C.
  • Remote Sensors: Up to 75 meters (250 feet).
  • Resolution: 0.1°C or °F, 0.01 psia, 0.1 PPMv < 1000 PPMv, 1 PPMv > 1000 PPMv.
Functional Specifications
  • Power Requirements: 90-230 VAC ±10%, 50-400Hz.
  • Operating Temperature: Control Unit: -10 to 60°C, DP Sensors: -40 to 100°C.
  • Auxiliary Coolant: Water or other, 2 liters/minute at 100 psia maximum.
  • Sample Pressure: 0 to 20 bar, Sample Flow Rate: 0.25 to 2.4 liters/minute.
  • Outputs: 4-20 mA, 0-5 VDC, RS232.
  • Alarms: Two form C, SPDT alarm relays, programmable from keyboard or RS232.
  • Display: 8 Line LCD with graphic data display and backlight.
Accessories / Options
  • Remote Mounting Kit with sensor cable up to 250ft.
  • Panel/Rack Mounting Kit, Sampling Module, Explosion-Proof Sensor Housing, NEMA-4 Wall Mount Enclosure, In-Line Particulate/Coalescing Filter.
See more

Catalog excerpts

Product data sheet DewMaster-1

EdgeTech field-proven chilled mirror technology. One, two or three stage sensors available - integral or remote from controller. NIST traceable calibration certificate. Scrolling menu via front panel & RS232. Depression range choice of either 45, 65 or 95°C . (one, two or three-stage cooling) Automatic Balance Control (ABC) o Automatically corrects for most mirror contaminants. o Automatically restandardizes the instrument. o You choose time and interval of restandardization, for best convenience. o Increase the length of unattended operation. Dual alarm capability o User programmable. o Can be set for latching or auto-reset mode. o Form C for greater versatility. Serial and analog outputs. RS232 communication with terminal, printer, or computer o Remote programming. o Data output such as dew point and alarm status. 0-5 VDC, 4-20 mA, 0-10 VDC, or 0-20mA o User scalable to drive peripheral devices. o Available simultaneously, one output per measurement parameter. o Scalable throughout the entire operating range of psychrometric variables. o Program via front panel or remote serial port. Pressure transducer input for psychrometric variable. Applications EdgeTech field proven technology Capable of measuring dew point, absolute pressure, PPMv and other psychrometric variables Primary chilled mirror measurement of water vapor content in gas Wide measuring range (-75 to 100°C) Automatic correction for mirror contaminants Certified measurements against NIST traceable standards Standard features, where others charge extra Analog & RS232 outputs Pro-Chem Analytik GmbH & Co.KG Neuer Bruchweg 1, 46509 Xanten Tel.: 02801/ 98510-70 Fax: 02801/ 98510-71 e-mail: [email protected] www.prochem-online.com Annealing Furnaces Process Dryers Clean Rooms Curing Ovens Plant Air Environmental Chambers Lab Calibration Standards Emission Testing Metrolo

 Open the catalog to page 1
Product data sheet DewMaster-2

Functional Power Requirements 90-230 VAC ±10%, 50-400Hz Auxiliary Coolant Water (or other) - 2 liters/minute (0.5 gallon/minute) at 100 psia maximum, to augment depression capability of sensor when necessary. Operating Temperature Control Unit: -10 to 60°C (15 to 140°F). DP Sensors: -40 to 100°C (-40 to 212°F). Sample Pressure 0 to 20 bar (0 to 300 psia), S1, S2 and S3. Depression Slew Rate 1.7°C (3°F)/Second Maximum, above 0°C Sample Flow Rate 0.25 to 2.4 liters/minute (0.5/5.0 SCFH). Remote Sensors Up to 75 meters (250 feet) Outputs (3 outputs available) 4-20 mA, 0-5 VDC, RS232. Resolution...

 Open the catalog to page 2
*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.
About DirectIndustry
DirectIndustry is a marketplace that offers B2B players a complete experience, from product sourcing to purchase. DirectIndustry is an entity of VirtualExpo Group.