O2Tracer

O2Tracer

O2Tracer

Product catalog summary
Overview
The O2Tracer is a two-wire, loop-powered oxygen transmitter designed for simple, accurate, and economical measurement of oxygen levels in various industrial applications. It offers user-specific measuring ranges between 0-100% oxygen.

Sensor and Calibration
The device uses a special fuel cell sensor that meets industrial standards for accuracy, sensitivity, and longevity. Calibration is performed using a calibration gas within the measuring range or with air for percent range measurements.

Key Features
  • Compact and inexpensive design
  • Long operating life of the sensor
  • IP65 enclosure for protection
  • Reverse voltage protection and temperature compensation

Specifications
  • Measuring Ranges: 0-100% O2, with specific ranges like 0-5% and 0-25%
  • Accuracy: +/- 2% FSD at constant temperature, +/- 5% FSD from 0 to 50°C
  • Response Time: Less than 10 seconds for 90% FSD at 25°C
  • Operating Temperature: 0 - 50°C
  • Pressure Range: 0.1 - 1 bar
  • Signal Output: 4-20 mA/DC
  • Alarm Value: Standard at 22.8 mA, optional at 3.4 mA
  • Voltage: 10-35 VDC with reverse voltage protection up to 40VDC
  • Load: Typical 470 Ohm, maximum 750 Ohm
  • Display: 6-digit backlit display
  • Housing: IP65 aluminum enclosure
  • Size: 80 x 75 x 55 mm
  • Weight: 0.3 kg

Applications
The O2Tracer is suitable for use in air, oxygen, and nitrogen generators, gas manufacturing and filling stations, heat treating, and bright annealing processes.

Additional Notes
The device is not recommended for measuring corrosive gases. Specifications and product configurations are subject to change without notice.
See more

Catalog excerpts

O2Tracer-1

O2Tracer 4 – 20 mA Loop Powered two wire Oxygen Transmitter Two Wire Process Oxygenanalyzer The Model O2Tracer is simple, accurate and very economic measurement of Oxygen at % levels. ☺ One User specific measuring range The unit will offer with different ranges between 0 – 100% of oxygen. ☺ Sensor The flow will be done trough the electronic enclosure A measuring of corrosive gases is not recommend with this measurement The O2Tracer uses a special fuel cell to measure the oxygen concentration. The sensor meets the industrial requirements for accuracy, sensitivity, easy to use and operating life. ☺ Calibration The calibration of the instrument for trace oxygen measurements in gas, should be done with a calibration gas. The concentration can be chosen freely within the measuring range. In the percent range the unit can calibrated with air. compact inexpensive Sensor with long operating life IP65 enclosure reverse voltage protection and temperature compensation Value analog Output in

 Open the catalog to page 1
O2Tracer-2

Measuring ranges all ranges between : Response time : Signal output : Alarm value : Typical • different sensors available • different Measuring ranges available • Air-, oxygen- and nitrogen generators • Gas manufacturers - filling stations • heat treating and bright annealing oxygen sensor reverse voltage protection up to 40VDC 6 Digit Display, backlight Micro-Fuel Cell, Remark; we have the right to revise specifications and product configurations without notice Tel.: 02804/ 9101-47 Fax: 02804/ 9101-48 e-mail: [email protected] www.prochem-online.com