The O2Tracer is a two-wire, loop-powered oxygen transmitter designed for simple, accurate, and economical measurement of oxygen levels at percentage concentrations. It is suitable for various industrial applications.

Specifications
  • Measuring Ranges: Configurable between 0-100% O2, with specific ranges like 0-5% and 0-25%.
  • Calibration: Can be calibrated using a calibration gas or air for percent range measurements.
  • Accuracy: +/- 2% FSD at constant temperature, +/- 5% FSD from 0 to 50°C.
  • Response Time: 90% FSD in less than 10 seconds at 25°C.
  • Operating Temperature: 0 - 50°C.
  • Pressure: 0.1 - 1 bar.
  • Signal Output: 4-20 mA/DC.
  • Alarm Value: Standard at 22.8 mA, optional at 3.4 mA.
  • Voltage: 10-35 VDC with reverse voltage protection up to 40VDC.
  • Load: Typical 470 Ohm, maximum 750 Ohm.
  • Display: 6-digit backlit display.
  • Sensor: Micro-Fuel Cell.
  • Housing: IP65 rated aluminum enclosure.
  • Size: 80 x 75 x 55 mm.
  • Connection: 6 mm or ¼” tubing.
  • Weight: 0.3 kg.

Features
  • Compact and inexpensive design.
  • Long operating life of the sensor.
  • Reverse voltage protection and temperature compensation.

Applications
The O2Tracer is suitable for use in air, oxygen, and nitrogen generators, gas manufacturing and filling stations, heat treating, and bright annealing processes.

Additional Notes
It is not recommended for measuring corrosive gases. The manufacturer reserves the right to revise specifications and product configurations without notice.
See more

Catalog excerpts

O₂ Tracer-1

O2Tracer 4 – 20 mA Loop Powered two wire Oxygen Transmitter Two Wire Process Oxygenanalyzer The Model O2Tracer is simple, accurate and very economic measurement of Oxygen at % levels. ☺ One User specific measuring range The unit will offer with different ranges between 0 – 100% of oxygen. ☺ Sensor The flow will be done trough the electronic enclosure A measuring of corrosive gases is not recommend with this measurement The O2Tracer uses a special fuel cell to measure the oxygen concentration. The sensor meets the industrial requirements for accuracy, sensitivity, easy to use and operating life. ☺ Calibration The calibration of the instrument for trace oxygen measurements in gas, should be done with a calibration gas. The concentration can be chosen freely within the measuring range. In the percent range the unit can calibrated with air. compact inexpensive Sensor with long operating life IP65 enclosure reverse voltage protection and temperature compensation Value analog Output in

 Open the catalog to page 1
O₂ Tracer-2

Measuring ranges all ranges between : Response time : Signal output : Alarm value : Typical • different sensors available • different Measuring ranges available • Air-, oxygen- and nitrogen generators • Gas manufacturers - filling stations • heat treating and bright annealing oxygen sensor reverse voltage protection up to 40VDC 6 Digit Display, backlight Micro-Fuel Cell, Remark; we have the right to revise specifications and product configurations without notice Tel.: 02804/ 9101-47 Fax: 02804/ 9101-48 e-mail: [email protected] www.prochem-online.com

 Open the catalog to page 2
*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.
About DirectIndustry
DirectIndustry is a marketplace that offers B2B players a complete experience, from product sourcing to purchase. DirectIndustry is an entity of VirtualExpo Group.