Overview
Priorclave Limited offers high-capacity PowerDoor autoclaves designed for laboratories with limited space and the need to sterilize large items. The RSV range features a vertical door movement, eliminating the need for additional space for door swing.

Specifications
The autoclaves are available in two chamber sizes: 450 litres and 700 litres. Both models feature a burnished 316 stainless steel chamber, front loading format, and an operating range up to 140°C and 2.4 bar. The 700-litre model has a heater power of 42kW and weighs 1360kg, while the 450-litre model has a heater power of 32.5kW and weighs 1180kg.

Key Features
  • STAYPUT locking geometry ensures the door remains locked even if the drive fails.
  • Tactrol 2® microprocessor control allows for variable setting of process time and temperature.
  • Optional features include automatic timed free-steaming, water cooling jackets, pre-cycle vacuum, and post-cycle drying.
  • Epoxy coated panels with anti-bacterial treatment effective against MRSA.
  • Pressure vessel warranty of 10 years and insurance approved.

Support and Service
Priorclave provides comprehensive pre- and after-sales support, including planned maintenance, calibration, and performance qualification services. The company is ISO9000 certified and UKAS approved, ensuring safe and effective service with full traceability.

Compliance
All Priorclave autoclaves comply with international standards such as BS2464-1993, PC500:2012+A1 Cat.3 European PED 97/23/EC B1+D, and are CE marked for BS EN 61010-2-41.

Installation Requirements
The 700-litre model requires a minimum installation space of 2500 x 1540mm, while the 450-litre model requires 2050 x 1540mm. Both models operate on a three-phase power supply.
See more

Catalog excerpts

Rectangular Powerdoor 450 & 700L Priorclave-1

RSV EH450/700 datasheet High Capacity PowerDoor Autoclaves Where laboratory space is at a premium and there is a need to sterilise tall, extra bulk items the RSV range of POWERDOOR autoclaves may prove the perfect answer since the door moves vertically within the autoclave frame, eliminating need to allow space for conventional door swing. The RSV higher capacity autoclaves are available in two chamber sizes of 450 and 700 litres and can be supplied in standard format or with a fascia panel. Although easy access to chamber is gained using Push-Button controls, inherent safety features prevent door opening during sterilising cycles and at high temperatures and pressures. An advantage of Priorclave PowerDoor autoclaves is the fixed gasket onto which the door closes. When raised in line with the chamber opening, the door automatically moves forward directly onto the seal to create an air-tight lock. This action prevents undue wear on the fixed seal, extending service intervals and improving operational performance of these autoclaves. Key Features • Special STAYPUT locking geometry keeps the pressure door fully locked even in event of drive failure. The push-button controls manage the functionality of thermal and pressure locks to prevent opening at load temperature above 80°C and pressures above 0.2 bar. • Tactrol 2® microprocessor control with simple and fully variable setting of process time, temperature with graphic indication of cycle status. • Automatic timed free-steaming. • Water cooling jackets.1 • Pre-cycle vacuum in which multiple Vacuum stages are programmed with interspersed heating to achieve steam penetration. Incorporates Vacuum Assisted Cooling.1 • Pulsed freesteaming to maximise efficiency of Pre-cycle vacuum phase.1 • Post Cycle Drying.1 • Epoxy coated panels and frame members are treated with an anti-bacterial agent which is effective against all bacteria and fungi including MRSA. • Pressure vessel carries a 10 year warranty and is insurance approved. • Service and support direct from manufacturer (or appointed distributor). 1 Optional Priorclave Limited 129-131 Nathan Way, West Thamesmead Business Park London, SE28 0AB T: +44 (0)20 8316 6620 F: +44 (0)20 8855 0616 For Direct Steam Heated models substitute SH for EH. Internal Steam Generator versions available.

 Open the catalog to page 1
Rectangular Powerdoor 450 & 700L Priorclave-2

Installation Specification model: PS/RSV/EH450 Steam heated models have an extra 45mm working height. Our highly trained service technicians provide dedicated pre- and after-sales support including planned maintenance programmes to ensure you get the most from your Priorclave. Our service doesn’t stop there though, as we offer calibration and performance qualification services to ensure that your autoclave and laboratory meet all the required standards. We are happy to service and calibrate other makes of autoclave and other laboratory equipment, offering a complete package to the laboratory. Our...

 Open the catalog to page 2
*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.