video corpo

EPP-Series

EPP-Series

EPP-Series

Product catalog summary
Design and Operation
The EPP-Series gauge-block calibration system utilizes an LM 20 laser-interferometric gauging probe with a measuring range of 20 mm and a resolution of 1 nm. It is designed for calibrating parallel-sided gauge blocks. The system's accuracy is supported by the German national bureau of standards (PTB), which found metric errors of less than 10 nm. The system requires only 15 reference-standard gauge blocks for calibrating a 122-piece set, reducing calibration costs and time.

Major Performance Features
  • Reduces the number of standard gauge blocks needed, cutting costs and speeding up measurements.
  • Calibrates non-standard sizes and materials.
  • Maintains high linearity and constant force across the measuring range.
  • Includes PEKAL software for temperature correction and data processing.
  • Optional temperature sensors available.

Technical Data
  • Calibration range: 0.5 mm to 100 mm.
  • Upper probe: SIOS Model LM 20, 20 mm range, 1 nm resolution, ≤ ± 10 nm uncertainty.
  • Lower probe: Inductive type, 0.6 N force.
  • Operating temperature: 20°C ± 0.5 K, with temperature resolution of ± 0.01 K.

Applications
The system is suitable for calibrating plane-parallel gauge blocks with rectangular cross-sections and complies with ISO 3650 standards.

Signal-Processing and Control Software
  • Controls probe motion and transmits data.
  • Compensates for errors and computes center sizes.
  • Outputs deviations and classifies gauge blocks by accuracy.
  • Generates custom test reports and certificates.

Contact Information
SIOS Meßtechnik GmbH, Am Vogelherd 46, D-98693 Ilmenau, Tel: +49-(0)3677-64470, Email: [email protected], URL: http://www.sios.de
See more

Catalog excerpts

EPP-Series-1

Gauge Block Calibration System

 Open the catalog to page 1
EPP-Series-2

Design and Operation The EPP-Series gauge-block calibration system for calibrating parallel-sided gauge blocks employs an LM 20 laser-interferometric gauging probe as its upper gauging probe. It has a measuring range of 20 mm and a length resolution of 1 nm. Investigations conducted by the German national bureau of standards (PTB) yielded metric errors of less than 10 nm for this gauging probe when used for calibrating gauge blocks. According to a PTB recommendation, the total number of reference-standard gauge blocks required for calibrating a 122-piece set of gauge blocks may thus be reduced...

 Open the catalog to page 2

All Piezosystem jena GmbH catalogs and technical brochures

  1. SL 04-Series

    2  Pages

  2. SL 03-Series

    2  Pages

  3. SL 02-Series

    2  Pages

  4. XY Stage

    4  Pages

  5. 24DV40

    3  Pages

  6. PosiCon 1000

    2  Pages

  7. HVP 1000/200

    3  Pages

  8. HVP 300/20

    3  Pages

  9. d-Drive

    2  Pages

  10. ENV

    4  Pages

  11. LCS - Series

    2  Pages

  12. SP-DS Series

    2  Pages

  13. SP-TR Series

    2  Pages

  14. MI-Series

    2  Pages

  15. NMM-1

    2  Pages

  16. LM-Series

    2  Pages

  17. Series PA

    4  Pages

  18. series N

    3  Pages

  19. Piezo-Actuators

    27  Pages

  20. Accessories

    2  Pages

  21. PSH25 OEM

    3  Pages

  22. nanoMIPOS 400

    2  Pages

  23. Mikroskopie

    4  Pages

  24. TRITOR 100

    2  Pages

  25. TRITOR 38

    2  Pages

Archived catalogs

  1. General Catalog

    98  Pages

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.