ION MOBILITY SPECTROMETER (IMS) ANALYZER - 1

Advanced IMS Analysis with High Resolving Power Ion Mobility Spectrometer Engine Patented technology enables fast, accurate results with simple sample preparation. High Resolving Power (64-150) High Sensitivity (19pg DtBP) Robust Ion Gate IMS-MS Compatible Microprocessor Control Easily Integrates with MS Software Use with Multiple Sources, API techniques Customize to Various Sources: • • • • • DESI DART Electrospray Corona Discharge Radiation The PHOTONIS Ion Mobility Spectrometer Engine is a patented platform that can be customized for a variety of specific applications where samples need to be quickly identified. Its unique design allows the user to quickly and inexpensively perform a simple IMS analysis and receive fast, accurate results even at room temperature sampling. It can be designed to operate with any number of ionization sources, can be scaled or customized, or can operate as a stand-alone unit. PHOTONIS' IMS Engine is a complete integrated unit which includes reaction and drift tubes, a high-performance ion gate, heater and integrated gas controls. The new PHOTONIS IMS Engine features a range of innovative technologies which combine to provide significantly higher resolving power when compared to currently available commercial IMS instruments, producing values ranging from 64-150. This IMS unit also features a new, robust ion gate technology which improves resolution and simplifies serviceability. Key components of the IMS Engine are made from Resistive Glass to replace conventional lens assemblies. In this patented platform, the reaction and drift tubes are comprised of Resistive Glass, providing a one-piece construction that produces minimal divergence and uniform counter flow of drift gas without the need for additional containment. The new PHOTONIS IMS Engine provides a compact and reliable analysis platform with improved performance. Its high resolving power and robust photo-etched ion gate make it ideal for use in a variety of applications where high volume, accuracy and speed of identification is critical to applications such as those involving product safety, composition analysis and environmental control. PHOTONIS' ability to design custom prototypes and develop them into full-scale production is unsurpassed in the detector and sensor industry. Schedule your technology assessment with PHOTONIS to see how our patented IMS analysis engine can improve your sample analysis. 660 Main Street, Sturbridge, MA USA 01566

Open the catalog to page 1
ION MOBILITY SPECTROMETER (IMS) ANALYZER - 2

Ion Mobility Spectrometer Engine Specifications: Patented Technology Physical Characteristics Reaction and Drift Tube ID Drift Tube Length Electrical Characteristics AC Power: High Voltage Range The use of Resistive Glass technology in Ion Mobility Spectrometers creates highly uniform electric fields that guide analyte ions to the detector with minimal divergence, resulting in higher resolving power. Drift and reaction tubes are one-piece construction that enables the uniform counter-flow of gas without the need for additional containment. Resistive Glass tubes are only available from PHOTONIS...

Open the catalog to page 2

All PHOTONIS catalogs and technical brochures

  1. Cricket

    2 Pages

  2. Channeltron

    4 Pages

  3. CPBP

    2 Pages

  4. FOLG

    2 Pages

  5. FieldMaster

    4 Pages

  6. HPD

    4 Pages

  7. PF1054

    1 Pages

  8. PF1053

    1 Pages

  9. Onyx

    2 Pages

  10. XD-4 ?

    2 Pages

  11. Streak Tube

    16 Pages